Search PubMed⌕ Search

PubMed · 3307036

A solid-phase antiglobulin test.

Abstract

A solid-phase test (SPT), a machine-readable adaptation of the indirect anti-human globulin test (IAHG), is described. The SPT level of sensitivity was comparable to that of the manual IAHG test for the detection of IgG antibodies and was compared with manual pretransfusion compatibility tests. The average difference from the mean of duplicate SPT results was 5.94 percent of the optical density value. The data also demonstrated a linear and overlapping relationship between the scores of manual IAHG tests and SPT optical density values. The accuracy of the SPT declined slightly as the concentration of red cell samples increased; the optimum concentration of immunoglobulin used in the preparation of the solid phase was about 275 mg per l. The 97.5th percentile of negative tests occurred at the same optical density (0.13) when serums from either healthy donors or hospitalized patients were tested. Compatibility tests completed on 1304 blood samples from donors and 437 serum samples from hospitalized patients showed that 1.38 percent of results were SPT-positive and manual IAHG-negative, and 5.37 percent were SPT-negative and manual IAHG-positive. The lowest possible score in the manual test was allocated to 97.14 percent of the latter group, indicating good correlation between the two techniques.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R Douglas, J V Schneider, D Wilkie, P A Harden. A solid-phase antiglobulin test.. https://doi.org/10.1046/j.1537-2995.1987.27587320527.x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

International reference reagents: antihuman globulin. An ISBT/ICSH joint working party report. International Society of Blood Transfusion. International Committee for Standardization in Haematology.

An international working party has conducted a study designed to select a suitable reference reagent for antihuman globulin, to replace those first made available in 1987. The chosen preparation contains levels of anti-IgG and anti-C3 (anti-C3c and anti-C3d) potency that are considered suitable to serve for reference when evaluating either polyspecific antihuman globulin reagents or those containing their separate monospecific components. The reference material is available in 2-ml freeze-dried aliquots from seven assigned distribution centres.

Coombs Test↗

Evaluation and implementation of the gel test for indirect antiglobulin testing in a community hospital laboratory.

BACKGROUND: The gel test, developed by Lapierre in 1984, was designed to standardize antiglobulin testing while improving sensitivity and specificity of the method. PRINCIPLE: Anti-human serum immunoglobulin G (IgG) mixed with Sephadex G100 (gel phase) in a microtube traps red cell-IgG agglutination complexes during migration through the gel in a centrifugation step. Agglutination complexes are visibly detectable at various levels in the microtube as an inverse function of antibody coated on red cells. Unsensitized red cells form a cell pellet at the base of the microtube. OBJECTIVE: To determine if indirect anti-human globulin testing could be standardized and simplified by replacing the tube test with the gel test without compromising quality or increasing costs. SETTING: A medium-sized community hospital. RESULTS: In a blinded retrospective study, we used patient sera (n = 40), which included 10 positive specimens containing 18 known antibodies. Sixteen antibodies were detected and identified with the tube method (1 anti-D and 1 anti-C not detected). By the gel test, 18 antibodies were detected and identified. All negative samples showed 100% concordance. Favorable results were obtained in a nonblinded prospective correlation study (n = 121). Our technologists found the gel test easier to read and more reproducible and reliable than the tube method; they also found increased sensitivity for detecting weakly reacting antibodies. We successfully introduced the gel test into our laboratory as the standard method for indirect antiglobulin testing. Following implementation, improved personnel management was achieved. CONCLUSIONS: The gel test is a reliable and advantageous method and is appropriate for routine use for detection and identification of alloantibodies in a community hospital transfusion service laboratory.

Coombs Test↗