Glycemic and safety outcomes of the insulin-only bionic pancreas in older adults and individuals with impaired awareness of Hypoglycemia: a post hoc analysis of a randomized pivotal trial.
AIMS: Evaluate the efficacy and safety of iLet Bionic Pancreas (BP) in older adults and individuals with impaired awareness of hypoglycemia (IAH). METHODS: This post hoc analysis used individual participant-level data from the Insulin-Only Bionic Pancreas Pivotal Trial (n = 440; NCT04200313). Eligible participants (n = 96) with type 1 diabetes, aged ≥ 60 years and/or had IAH (Clarke score ≥ 4), were randomized to BP with aspart/lispro (BP-Asp/Lis; n = 45), BP with fast-acting aspart configuration (BP-Fiasp; n = 31), or standard care (SC; n = 20) for 13 weeks. RESULTS: Compared with SC, time-in-range (70-180 mg/dL) significantly increased by 7.49 % (95 % CI: 2.61 to 12.38; ∼1.8 h/day) with BP-Asp/Lis and by 8.28 % (95 % CI: 3.15 to 13.41; ∼2.0 h/day) with BP-Fiasp, driven by reduced hyperglycemia. No significant differences were observed in hypoglycemia exposure. Severe hypoglycemia occurred in four participants (four events) on BP-Asp/Lis and one participant (two events) on SC. One diabetic ketoacidosis event occurred on BP-Fiasp due to an infusion set failure. CONCLUSIONS: In high-risk, clinically vulnerable populations, the BP system significantly improved glycemic control while maintaining safety parity with respect to hypoglycemia risk, providing a resilient therapeutic alternative for vulnerable cohorts.