A simple method for evaluating data from an interlaboratory study.

Abstract

Large-scale laboratory- and method-performance studies involving more than about 30 laboratories may be evaluated by calculating the HORRAT ratio for each test sample (HORRAT = [experimentally found among-laboratories relative standard deviation] divided by [relative standard deviation calculated from the Horwitz formula]). The chemical analytical method is deemed acceptable per se if HORRAT approximately 1.0 (+/- 0.5). If HORRAT is > or approximately 2.0, the most extreme values are removed successively until an "acceptable" ratio is obtained. The laboratories responsible for the extreme values that are removed should examine their technique and procedures. If > or approximately 15% of the values have to be removed, the instructions and the methods should be examined. This suggested computation procedure is simple and does not require statistical outlier tables. Proposed action limits may be adjusted according to experience. Data supporting U.S. Environmental Protection Agency method 245.1 for mercury in waters (manual cold-vapor atomic absorption spectrometry), supplemented by subsequent laboratory-performance data, were reexamined in this manner. Method-performance parameters (means and among-laboratories relative standard deviations) were comparable with results from the original statistical analysis that used a robust biweight procedure for outlier removal. The precision of the current controlled performance is better by a factor of 4 than that of estimates resulting from the original method-performance study, at the expense of rejecting more experimental values as outliers.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

W Horwitz, P Britton, S J Chirtel. A simple method for evaluating data from an interlaboratory study.. https://pubmed.ncbi.nlm.nih.gov/9850588/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Determination of iron in solutions with the ferric-xylenol orange complex.

Formation of a colored complex between ferric iron and xylenol orange has been used in a variety of applications, but there is disagreement about the number of complexes that exist, their stoichiometry, pH dependence, solvent effects, and other parameters. In this study we investigated the use of the complex to measure micromolar concentrations of iron in aqueous and alcohol solutions. The results show that the optimum wavelength for the measurement is 560 nm, that only a 1:1 ferric:xylenol orange complex forms, and that its molar absorption coefficient is affected by the solvent, the source of the xylenol orange, and the pH. Under the recommended conditions, formation of the complex is completed in less than 5 min and it is stable in several solvents. Its molar absorption coefficient in 25 mM H(2)SO(4) is 20,100 or 14,500 M(-1) cm(-1), depending on the source of the dye. A recommended protocol for the use of the assay is given.

Chemistry Techniques, Analytical

Vapor recognition with small arrays of polymer-coated microsensors. A comprehensive analysis.

A comprehensive analysis of vapor recognition as a function of the number of sensors in a vapor-sensor array is presented. Responses to 16 organic vapors collected from six polymer-coated surface acoustic wave (SAW) sensors were used in Monte Carlo simulations coupled with pattern recognition analyses to derive statistical estimates of vapor recognition rates as a function of the number of sensors in the array (< or = 6), the polymer sensor coatings employed, and the number and concentration of vapors being analyzed. Results indicate that as few as two sensors can recognize individual vapors from a set of 16 possibilities with < 6% average recognition error, as long as the vapor concentrations are > 5 x LOD for the array. At lower concentrations, a minimum of three sensors is required, but arrays of 3-6 sensors provide comparable results. Analyses also revealed that individual-vapor recognition hinges more on the similarity of the vapor response patterns than on the total number of possible vapors considered. Vapor mixtures were also analyzed for specific 2-, 3-, 4-, 5-, and 6-vapor subsets where all possible combinations of vapors within each subset were considered simultaneously. Excellent recognition rates were obtainable for mixtures of up to four vapors using the same number of sensors as vapors in the subset. Lower recognition rates were generally observed for mixtures that included structurally homologous vapors. Acceptable recognition rates could not be obtained for the 5- and 6-vapor subsets examined, due, apparently, to the large number of vapor combinations considered (i.e., 31 and 63, respectively). Importantly, increasing the number of sensors in the array did not improve performance significantly for any of the mixture analyses, suggesting that for SAW sensors and other sensors whose responses rely on equilibrium vapor-polymer partitioning, large arrays are not necessary for accurate vapor recognition and quantification.

Chemistry Techniques, Analytical

Fine chemical manipulations of microscopic liquid samples. 1. Droplet loading with chemicals.

Physically and chemically stable microscopic aqueous droplets of nano-, pico-, and femtoliter volumes were made and kept under heptane. Such droplets can contain chemicals of interest in pico-, femto-, and attomole amounts or less. Their fine chemical manipulation was achieved by a diffusional microburet (DMB), which consists of a pulled glass capillary whose microscopic tip is filled with a tiny diffusion membrane of agar or polyacrylamide gel. Once this tip is moved into a target droplet by a fine micromanipulator under a microscope, diffusional reagent delivery from the DMB body begins, driven by the concentration gradient within the tip. This system was tested in this work by delivering an inert fluorescent dye, Lucifer Yellow CH, into buffer droplets while fluorescence intensity in each droplet was recorded. An exponential decay of delivery rate was observed corresponding to an exponential saturation process for the accumulation of the delivered chemical inside the target droplet.

Chemistry Techniques, Analytical