PubMed · 9844690
CT bronchoscopy: optimization of imaging parameters.
Abstract
The authors evaluated the relative importance of the following scanning parameters at computed tomographic bronchoscopy in an anesthetized adult sheep's thorax: section thickness (2, 4, 8 mm), pitch (1.0, 1.5, 2.0), milliampere setting (100, 175, 250 mA), and overlap of reconstructed sections (0%, 25%, 50%, 75%). Five blinded readers ranked the images twice in comparison with photographs of the mounted specimen. Differences in image quality were significant (P < .001) with section thickness of 2 mm and a pitch of 1.0. The milliampere setting had only a minor effect on image quality, and a 50% overlap of reconstructed sections was best.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
K D Hopper, T A Iyriboz, R P Mahraj, S W Wise, C J Kasales, T R TenHave, R P Wilson, J S Weaver. 1998. CT bronchoscopy: optimization of imaging parameters.. https://doi.org/10.1148/radiology.209.3.9844690. https://searchpubmed.org/record/pubmed%3A9844690/ct-bronchoscopy-optimization-of-imaging-parameters
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search PubMed by Indiciae: BibTeX · RIS.