PubMed · 9494243
Computerized mechanical validation of custom-made stems.
Abstract
The new community laws regarding medical devices require that custom-made devices be prescribed by the surgeon, who assumes responsibility for the device along with the manufacturer. It is the purpose of this study to indicate a method that allows use to check whether custom-made prostheses in particular conform to the requirements established by European laws; these are prostheses produced in a single model and that have rather short development and realization times. The pre-clinical method of validation proposed, based on a calculation with finite elements, was first verified by in vitro measurements and then used to evaluate 15 cases selected among the most critical ones in terms of size and/or geometrical irregularity from a group of 100 stems that had already been implanted. The method proposed was confirmed to be accurate and reliable; in 3 out of 15 stems analyzed we found values for mechanical resistance that were less than minimum values required by law.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
M Viceconti, L Bellingeri, M Baleani, A Toni. Computerized mechanical validation of custom-made stems.. https://searchpubmed.org/record/pubmed%3A9494243/computerized-mechanical-validation-of-custom-made-stems
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search PubMed by Indiciae: BibTeX · RIS.