The LightCycler: a microvolume multisample fluorimeter with rapid temperature control.

Abstract

Experimental and commercial microvolume fluorimeters with rapid temperature control are described. Fluorescence optics adopted from flow cytometry were used to interrogate 1-10-microL samples in glass capillaries. Homogeneous temperature control and rapid change of sample temperatures (10 degrees C/s) were obtained by a circulating air vortex. A prototype 2-color, 32-sample version was constructed with a xenon arc for excitation, separate excitation and emission paths, and photomultiplier tubes for detection. The commercial LightCycler, a 3-color, 24-sample instrument, uses a blue light-emitting diode for excitation, paraxial epi-illumination through the capillary tip and photodiodes for detection. Applications include analyte quantification and nucleic acid melting curves with fluorescent dyes, enzyme assays with fluorescent substrates and techniques that use fluorescence resonance energy transfer. Microvolume capability allows analysis of very small or expensive samples. As an example of one application, rapid cycle DNA amplification was continuously monitored by three different fluorescence techniques, Which included using the double-stranded DNA dye SYBR Green I, a dual-labeled 5'-exonuclease hydrolysis probe, and adjacent fluorescein and Cy5z-labeled hybridization probes. Complete amplification and analysis requires only 10-15 min.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

C T Wittwer, K M Ririe, R V Andrew, D A David, R A Gundry, U J Balis. 1997. The LightCycler: a microvolume multisample fluorimeter with rapid temperature control.. https://doi.org/10.2144/97221pf02

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Design, synthesis and biological evaluation of hydroxybenzothiazole-linked benzothiazole/benzoxazole conjugates as potent dual α-amylase and α-glucosidase inhibitors.

The current study focuses on the synthesis and evaluation of novel Hydroxybenzothiazole-Linked Benzothiazole/Benzoxazole Conjugates to target Diabetes Mellitus (DM) by inhibiting α-amylase and α-glucosidase. Spectroscopic methods, including 1H and 13C NMR spectroscopy, were employed to confirm the structures of newly synthesized conjugates. The findings of in-vitro analysis displayed that the synthesized derivatives inhibited α-amylase and α-glucosidase enzymes with IC50 values ranging from 3.65 ± 0.20 μM to 32.15 ± 3.20 μM on α-amylase and 5.92 ± 0.80 μM to 35.60 ± 3.40 μM on α-glucosidase, in contrast to the reference drug Acarbose (α-amylase IC50 = 8.25 ± 0.80 μM; α-glucosidase IC50 = 10.75 ± 1.10 μM). Among the series 9a-9f and 10a-10f, analogs 10 f, 10b, 9b, and 9e displayed superior anti-diabetic activity compared to the reference drug Acarbose. The inhibitory activity of these conjugates can be attributed to their favorable and stable interactions with critical amino acid residues of targeted enzymes, as revealed through molecular docking analysis. ADMET predictions and drug-likeness evaluations showed favorable pharmacokinetic features, while DFT investigations revealed electronic insights related to bioactivity. Experimental outcomes and in silico support display that these potent Hydroxybenzothiazole-Linked Benzothiazole/Benzoxazole Conjugates were comparable to an existing diabetic mellitus inhibitor while conserving an acceptable safety profile, specifying potential for further therapeutic development and optimization against diabetic Mellitus.

Benzothiazoles↗

Valvular heart disease in Parkinson's disease patients treated with dopamine agonists: a reader-blinded monocenter echocardiography study.

Fibrotic valvular heart disease (VHD) has been reported in association with ergot dopamine agonists (DAs), but the current database is insufficient regarding clinical relevance and comparison to data on non-ergot DAs. We evaluated the effects of four DAs (pergolide, cabergoline, ropinirole, pramipexole) on morphology and function of heart valves in patients with Parkinson's disease (PD) to determine the frequency and clinical relevance of DA-induced VHD. A total of 85 patients treated with ergot or non-ergot DAs and 38 age-matched controls were evaluated by transthoracic echocardiography. Valvular pathology was assessed by established criteria of valvular regurgitation and a VHD scoring system. Both grading systems revealed increased frequency of VHD in ergot DA patients compared to both non-ergot DA patients and controls with 22% of ergot DA patients having moderate VHD versus 3% of non-ergot DA patients and none of controls (P = 0.001). We did not find correlations of echocardiographic findings with duration/cumulative dose of treatment, age, or vascular risk factors. Our data suggest that ergot DAs are associated with higher prevalence of VHD compared to non-ergot DAs and controls. Standard echocardiography seems sufficient to detect VHD in PD patients treated with DAs.

Benzothiazoles↗

Detection and quantification of hepatitis B virus DNA by SYBR green real-time polymerase chain reaction.

A single-round real-time polymerase chain reaction (PCR) assay based on SYBR green dye technology for the detection and quantification of hepatitis B virus (HBV) DNA in serum was evaluated and compared with a qualitative nested PCR and the Cobas Amplicor HBV Monitor assay (Roche Molecular Diagnostics, Milan, Italy). The performance of the real-time PCR assay was evaluated in a routine clinical laboratory setting with a total of 212 clinical specimens. The sensitivity of the real-time PCR corresponded to 31 IU/ml (70 copies/ml), and comparison with the qualitative nested PCR showed significant concordance for 94% of samples. The linear curve over 7 log units, spanning 10(3)-10(9) IU/ml (2.28 x 10(3) to 2.28 x 10(9) copies/ml), was observed in the quantitative determination. The interexperimental variability coefficient of the assay ranged from 0.22 to 0.39 and the intraexperimental variability coefficient from 0.24 to 0.41. By excluding values outside of the dynamic ranges of both tests, the HBV Monitor and the real-time PCR gave an agreement within +/-1 log unit for 90% of samples, while those for the remaining 10% were found to be above 1 log unit but less than 1.5 log units. When the results inside and outside the dynamic range of the HBV Monitor were examined, 90% of the results were in agreement. In conclusion, the real-time PCR based on SYBR green technology proved suitable for routine diagnostic purposes, showing good sensitivity, high specificity, high reproducibility, and good linearity over a broad dynamic range of quantification.

Benzothiazoles↗