Quantitative multi-detector emission computerized tomography using iterative attenuation compensation.

Abstract

An iterative procedure to correct for attenuation has been developed for a multidetector, single-photon emission tomographic scanner. The difference between measured and estimated data projections is used at each iteration to form an error image which is used, in turn, to correct the image. A damping factor that minimizes chi2 is applied after each iteration to speed convergence. Several phantoms of different size, with various concentration distributions, have been used to compare this method with a first-order multiplicative attenuation correction used previously with this scanner. The first-order correction is inadequate for most of the phantoms studied, whereas relative and absolute quantitative capability is demonstrated for the iterative attenuation correction. The reconstructed average number of counts per pixel is a linear function of activity concentration up to 5 muCi/ml for all regions of uniform activity whose size is greater than or equal to 5 cm. The importance of using an accurate attenuation distribution with this method is demonstrated with a torso-like phantom.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

S C Moore, J A Brunelle, C M Kirsch. 1982. Quantitative multi-detector emission computerized tomography using iterative attenuation compensation.. https://searchpubmed.org/record/pubmed%3A6980972/quantitative-multi-detector-emission-computerized-tomography-using-iterative-attenuation-compensatio

Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search PubMed by Indiciae: BibTeX · RIS.

KEEP EXPLORING

Related citations

Content uniformity and assay requirements in current regulations.

The acceptance of a tablet batch is based both on the content uniformity test and on the assay. It is shown that these two characteristics are not independent, and the acceptance criteria for them are not even consistent. For content uniformity range three methods of calculation are compared: the present European Pharmacopoeia method, a tolerance range method with improved k tolerance factor and a one-way random effects analysis of variance model. To resolve the inconsistency several options are discussed: applying the holistic content uniformity range alone; using content uniformity standard deviation and assay mean simultaneously or applying a criterion based on Taguchi's quadratic loss function.

Mathematics↗

Stable time filtering of strongly unstable spatially extended systems.

Many contemporary problems in science involve making predictions based on partial observation of extremely complicated spatially extended systems with many degrees of freedom and with physical instabilities on both large and small scale. Various new ensemble filtering strategies have been developed recently for these applications, and new mathematical issues arise. Because ensembles are extremely expensive to generate, one such issue is whether it is possible under appropriate circumstances to take long time steps in an explicit difference scheme and violate the classical Courant-Friedrichs-Lewy (CFL)-stability condition yet obtain stable accurate filtering by using the observations. These issues are explored here both through elementary mathematical theory, which provides simple guidelines, and the detailed study of a prototype model. The prototype model involves an unstable finite difference scheme for a convection-diffusion equation, and it is demonstrated below that appropriate observations can result in stable accurate filtering of this strongly unstable spatially extended system.

Mathematics↗