Peak-clipping in induction loop systems.

Abstract

The decision to fit an induction loop system in many buildings is often dictated by cost, in particular that of the amplifier. In order to accommodate the peaks of speech the amplifier must be 4 times as powerful as that required for average speech. This often makes a profound difference to the cost and tips the balance against a loop being installed. Experience with two loops fitted for an audience of hard of hearing people where the peaks were clipped without any apparent loss of discrimination initiated this study. The findings showed that although high fidelity is to be preferred, peak clipping in loop systems can be tolerated without serious loss of intelligibility.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

E Trinder. 1984. Peak-clipping in induction loop systems.. https://doi.org/10.3109/03005368409078934

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A broadband phased-array system for direct phosphorus and sodium metabolic MRI on a clinical scanner.

Despite their proven gains in signal-to-noise ratio and field-of-view for routine clinical MRI, phased-array detection systems are currently unavailable for nuclei other than protons (1H). A broadband phased-array system was designed and built to convert the 1H transmitter signal to the non-1H frequency for excitation and to convert non-1H phased-array MRI signals to the 1H frequency for presentation to the narrowband 1H receivers of a clinical whole-body 1.5 T MRI system. With this system, the scanner operates at the 1H frequency, whereas phased-array MRI occurs at the frequency of the other nucleus. Pulse sequences were developed for direct phased-array sodium (23Na) and phosphorus (31P) MRI of high-energy phosphates using chemical selective imaging, thereby avoiding the complex processing and reconstruction required for phased-array magnetic resonance spectroscopy data. Flexible 4-channel 31P and 23Na phased-arrays were built and the entire system tested in phantom and human studies. The array produced a signal-to-noise ratio improvement of 20% relative to the best-positioned single coil, but gains of 300-400% were realized in many voxels located outside the effective field-of-view of the single coil. Cardiac phosphorus and sodium MRI were obtained in 6-13 min with 16 and 0.5 mL resolution, respectively. Lower resolution human cardiac 23Na MRI were obtained in as little as 4 sec. The system provides a practical approach to realizing the advantages of phased-arrays for nuclei other than 1H, and imaging metabolites directly.

Amplifiers, Electronic↗

Noise characteristic design of CMOS source follower and voltage amplifier for active semiconductor micro-electrodes for neural signal recording.

A noise performance design method for the pre-amplifiers of an active neural probe is given. The on-chip circuitry of the active neural probe consists of CMOS devices that show high-/low-frequency noise, so that the device noise can become dominant. Analysis of the signal-to-device-noise ratio (SDNR) for the CMOS source follower buffer and two-stage differential voltage amplifier is given. Closed-form expressions for the output noise power are derived and exploited to tailor the parameters that are controllable during circuit design. The output SDNR is calculated considering the real extracellular action potentials, the electrode-electrolyte interface and the noise spectrum of CMOS devices from typical foundries. It is shown that the output device noise power can be much higher than the output signal power if the devices at the input stage of the pre-amplifier are made as small as given fabrication technology permits. Quantitative information of the circuit parameters to achieve an SDNR higher than 5 for neural spikes with 60 microV amplitude are provided for both pre-amplifier types.

Amplifiers, Electronic↗

Human studies of a piezoelectric transducer and a microphone for a totally implantable electronic hearing device.

OBJECTIVE: For the surgical treatment of patients with moderate and severe sensorineural hearing loss, the authors have developed a totally implantable hearing device, the totally integrated cochlea amplifier (TICA). To evaluate the effectiveness of transducer and microphone of this device, three separate human studies were conducted. STUDY DESIGN: The first study using transducer prototypes involved self experiments in investigators with normal hearing. The second study used the transducer prototypes in patients with hearing loss, and the third study involved the temporary implantation of the final transducer prototype and microphone in patients undergoing otologic surgery. PATIENTS: In routine middle ear surgery, transducer prototypes were coupled to the ossicular chain of 28 patients. In addition to the transducer, in 5 patients the microphone was placed beneath the skin of the auditory canal, allowing the skin to cover the microphone membrane completely. RESULTS: The piezoelectric transducer reached an equivalent sound pressure level of 145 dB SPL < or =10 kHz. The dynamics for music reached 32 dB, which was identical with the results of the preoperative investigations using high-fidelity headsets (33 dB). The low nonlinear distortions of <0.1% and the frequency range of 10 kHz are reflected in the positive evaluation of the sound quality by 84% of the patients involved. When phonetically balanced speech material and music were presented under free field conditions at a sound level of 65 dB SPL, understanding of the phonetically balanced speech material was 100%. Most patients judged the presentations of music as clear and undistorted with all broadband components. CONCLUSIONS: Data in humans on the performance of the two main components of the TICA implant, the transducer and the microphone, are reported.

Amplifiers, Electronic↗