Pacemaker leads.

Abstract

Present day pacemaker leads are far superior in every respect to those of the past. Modification of fixation characteristics has reduced displacement rates to 1% or less in most centers. Fracture of multifilar leads is a rarity. Biodegradation of polyurethane insulation appears to be an isolated problem specific to individual lead models and may be related to physical stresses incurred during manufacture or lead insertion. Recent evidence has incriminated an interaction of polyurethane with silver which arises from the drawn braised strand conductor substrate of those leads in which this problem has been noted. This may explain why the problem has been restricted to specific lead models of one manufacturer to date. Lack of uniformity of lead terminal size between manufacturers and even within the same manufacturer's product line continues to baffle this observer. Although past problems of lead displacement have been markedly reduced, the difficulty of removing chronic leads which have become septic appears to have worsened. Modification of existing leads to ensure that the interface between electrode tip and proximal shaft is unidiametric is essential.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J W Harthorne. 1984. Pacemaker leads.. https://doi.org/10.1016/0167-5273(84)90321-8

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

The proarrhythmic potential of implantable cardioverter-defibrillators.

The implantable cardioverter-defibrillator (ICD) is remarkably effective in preventing sudden cardiac death in high-risk patients, but it also has the capacity to provoke or worsen cardiac arrhythmias. Tachyarrhythmias or bradyarrhythmias may result from the delivery of antitachycardia or antibradycardia therapies by tiered-therapy defibrillators. This proarrhythmia, although rarely fatal, increases the morbidity associated with defibrillator therapy. Proarrhythmia is related as much to suboptimal programming as to technical limitations of the device. The proarrhythmic potential of ICD therapy can be minimized by tailoring the "electrical prescription" according to characteristics of the clinical arrhythmia and individual ICD idiosyncrasies.

Cardiac Pacing, Artificial