A microcomputer based controller for neuromuscular block during surgery.

Abstract

A microcomputer based control system has been developed to automatically induce and maintain neuromuscular block during surgery. The system repeatedly adjusts the infusion rate of a muscle relaxant, succinylcholine, based on the evoked, rectified, and integrated electromyogram (EMG). The system was tested in 12 patients for a 30-min infusion period with a setpoint of 80% depression of the evoked, rectified, and integrated thenar EMG. The mean time to reach the setpoint for 10 of the patients was 5.5 (+/- 1.87 SD) min and the mean time for 95% recovery after infusion was stopped was 5.4 (+/- 0.83 SD) min. Average overshoot was 9.9% of the baseline (+/- 3.1% SD), and the average time within +/- 10% of the setpoint was 22.1 min (+/- 7.26 SD). The total dose of succinylcholine for these 10 patients ranged from 1.21 to 3.77 mg/kg with a mean of 1.92 mg/kg. The other two patients were relatively insensitive to the drug and the controller was unable to bring the response to the setpoint due to a ceiling placed on infusion rate by the control algorithm.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

G Ritchie, J P Ebert, T C Jannett, I Kissin, L C Sheppard. 1985. A microcomputer based controller for neuromuscular block during surgery.. https://doi.org/10.1007/bf02371246

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Cost-effective gel documentation using a web-cam.

In search for a cost effective gel documentation system applicable for different fields of molecular biology, we analyzed the capabilities of a cheap CCD-camera originally designed to capture images for transmission through the internet (web-cam) with regard to gel documentation. The camera was connected to a personal computer by universal serial bus (USB) and used for the documentation of DNA separated on agarose gels and stained by ethidium-bromide using the software provided with the camera. The web-cam provided digital images of sufficient quality for routine documentation and combined the low set-up costs of a Polaroid system with the low running costs of video capture systems, hence is ideal as a start-up system and as augmentation to existing equipment.

Computers↗