Monaghan 225 ventilator use under hyperbaric conditions.

Abstract

The Monaghan 225 ventilator was tested to ambient pressures of 6 atmospheres absolute (ATA) in a hyperbaric chamber. The ventilator would function with delivered tidal volume which was independent of ambient pressure. Ventilatory rate declined in an exponential fashion. At 6 ATA, the ventilatory rate was 45 percent of the preset rate at 1 ATA. By decreasing the circuit resistance and increasing the inspiratory flow rate, the 6 ATA rate could be increased to 72 percent of the 1 ATA value. The maximum minute ventilation of the ventilator at 1 ATA was approximately 48 L/min; at 6 ATA, its maximum was 18 L/min. Synchronized intermittent mandatory ventilation, assist/control, and PEEP functions were satisfactory at 6 ATA. While using 100 percent O2 to power the ventilator at 2.82 ATA, the oxygen leakage was 57.7 L/min (converted to 1 ATA pressure, 20 degrees C), of which 33.7 L/min was successfully scavenged using simple techniques. A minor modification was made to the ventilator, allowing it to be driven by compressed air while maintaining complete flexibility in setting the FIo2. The ventilator has proven stable and reliable in clinical use at ambient pressures up to 6 ATA.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R E Moon, L V Bergquist, B Conklin, J N Miller. 1986. Monaghan 225 ventilator use under hyperbaric conditions.. https://doi.org/10.1378/chest.89.6.846

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Setup reproducibility in radiation therapy for lung cancer: a comparison between T-bar and expanded foam immobilization devices.

PURPOSE: Physiologic and non-physiologic tumor motion complicates the use of tight margins in three-dimensional (3D) conformal radiotherapy. Setup reproducibility is an important non-physiologic cause of tumor motion. The objective of this study is to evaluate and compare patient setup reproducibility using the reusable T-bar and the disposable expanded foam immobilization device (EFID) in radiation therapy for lung cancer. METHODS AND MATERIALS: Two hundred forty-four portal films were taken from 16 prospectively accrued patients treated for lung cancer. Patients were treated with either a pair of anterior and posterior parallel opposing fields (POF), or a combination of POF and a three-field isocentric technique. Each patient was treated in a supine position using either the T-bar setup or EFID. Six patients were treated in both devices over their treatment courses. Field placement analysis was used to evaluate 3D setup reproducibility, by comparing positions of bony landmarks relative to the radiation field edges in digitized simulator and portal images. Anterior-posterior, lateral, and longitudinal displacements, as well as field rotations along coronal and sagittal planes were measured. Statistical analyses of variance were applied to the deviations among portal films of all patients and the subgroup treated with both immobilization methods. RESULTS: For the T-bar immobilization device, standard deviations of the setup reproducibility were 5.1, 3.7, and 5.1 mm in the anterior-posterior, lateral, and longitudinal dimensions, respectively. Rotations in the coronal plane and the sagittal plane were 0.9 degrees and 1.0 degrees, respectively. For the EFID, corresponding standard deviations of set up reproducibility were 3.6 mm, 5.3 mm, 5.4 mm, 0.7 degrees and 1.4 degrees, respectively. There was no statistically significant difference (p = 0.22) in the 3D setup reproducibility between T-bar and EFID. Subgroup analysis for the patients who were treated with both immobilization devices did not reveal a difference either. There was no consistent systematic error from simulator to treatment unit identified for either immobilization device. CONCLUSION: Although the optimal immobilization technique and patient positioning for thoracic radiotherapy have yet to be determined, this study indicates that T-bar is comparable with EFID in its setup reproducibility. In view of the inherent advantages of T-bar, it has become a standard immobilization device at our institution. The observed range of displacements in field positioning with either immobilization device implies that one cm (two standard deviations [SD] of setup error) will be a more appropriate margin to allow for setup variability in radiation therapy for lung cancer.

Equipment Design