Dynamic stereotactic radiosurgery.

Abstract

Two radiosurgical procedures using a stereotactic frame and a linear accelerator X ray beam with a circular field diameter between 0.5 and 3 cm are presented. One technique is based on a single plane rotation (single plane radiosurgery) whereas the other uses simultaneous and continuous motions of both the gantry (approximately 360 degrees) and couch (approximately 180 degrees) during the radiosurgical procedure (dynamic radiosurgery). The dose, typically a few thousand cGy, is prescribed to the 90% isodose line which just covers the target volume. The dose fall-off outside the spherical target volume is considerably sharper for the dynamic rotation than for the single plane rotation, and is comparable to the dose fall-off obtained with the two presently known dedicated radiosurgical techniques: one based on focused cobalt beams and the other on proton beams. The dose fall-off in the dynamic radiosurgery discussed here is also comparable to that of previously described linear accelerator based multiple converging are techniques, making the dynamic radiosurgery an attractive alternative to presently known radiosurgical procedures. The radiation beam parameters are discussed and the stereotactic frame described. The dose distributions for both radiosurgical techniques are calculated in a single plane and then corrected for the attenuation effects in the stereotactic frame (approximately 2%) and for the effects of the dynamic rotation (approximately 2%). The skin doses are 0.7% and 2%, and the lens doses, if the beam passes through the eyes, are 2.5% and 3.5% for the dynamic rotation and single plane rotation, respectively. The scatter and leakage dose for the radiosurgical procedures is typically 0.2% to the patient's thyroid, 0.06% to the breast, and 0.02% to the gonads.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

E B Podgorsak, A Olivier, M Pla, P Y Lefebvre, J Hazel. 1988. Dynamic stereotactic radiosurgery.. https://doi.org/10.1016/0360-3016(88)90059-4

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A penalized likelihood approach to magnetic resonance image reconstruction.

Currently, images acquired via magnetic resonance imaging (MRI) and functional magnetic resonance imaging (fMRI) technology are reconstructed using the discrete inverse Fourier transform. While computationally convenient, this approach is not able to filter out noise. This is a serious limitation because the amount of noise in MRI and fMRI can be substantial. In this paper, we propose an alternative approach to reconstruction, based on penalized likelihood methodology. In particular, we focus on non-linear shrinkage estimators and show that this approach achieves a great reduction in integrated mean squared error (IMSE) of the estimated image with respect to the currently used estimator. This approach is extremely fast and easy to implement computationally. In addition, it can be combined with various alternative approaches to MR image reconstruction and can be easily adapted to other, non-MRI contexts, in which the observed data and the quantities of interest are related via a linear transform.

Brain Diseases↗