Microprocessor-controlled hemodynamics: a step towards improved efficiency and safety.

Abstract

Manual titration of sodium nitroprusside (SNP) is widely used for treatment of hypertension following cardiac surgery. This study compared conventional manual control with control by a research prototype of an automatic infusion module based on a proportional plus integral plus derivative (PID) negative feedback loop. Two groups of coronary artery bypass patients requiring SNP for postoperative hypertension were studied prospectively. In the first group, hypertension was controlled by manual adjustment of the SNP infusion rate, and in the second, the infusion rate was controlled automatically. The actual and desired mean arterial pressures (MAP) over consecutive ten-second epochs were recorded during the period of infusion. The MAP was maintained within 10% of the desired MAP 45.8% of the time in the manual group, compared with 90.0% in the automatic group, and the mean percent error in the automatic group was significantly less than in the manual group (P less than 0.01). It is concluded that adoption of such systems will result in improved patient safety and may facilitate more effective distribution of nursing staff within intensive care units.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

B E Keogh, J Jacobs, D Royston, K M Taylor. 1989. Microprocessor-controlled hemodynamics: a step towards improved efficiency and safety.. https://doi.org/10.1016/0888-6296(89)90003-3

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A user-friendly PC-based data acquisition and analysis system for respirometric investigations.

We have developed an easy-to-use computer-based system for recording, displaying, storing and analyzing signals generated by Clark-type oxygen electrodes. A user-friendly interface of Windows-based program BioMed significantly increases the productivity of investigations. It allows to process, control, present and archive the experimental data in real time. A 12-bit analog-to-digital-converter, analog and digital filters, a possibility to zoom the obtained respiratory curves and calculation of the respiration rates by a linear regression method increase the resolution of the estimated oxygen consumption rates. The new system enables to register even small changes, such as 3-5 ngatoms O/min, in respiration rates of biological objects -- enzymes, mitochondria and permeabilized muscle fibers. The system has been developed and is regularly used for the respirometric investigations at the Laboratory of Biochemistry, Institute for Biomedical Research, Kaunas University of Medicine.

Analog-Digital Conversion↗

Elimination of depth degeneracy in optical frequency-domain imaging through polarization-based optical demodulation.

A novel optical frequency-domain imaging system is demonstrated that employs a passive optical demodulation circuit and a chirped digital acquisition clock derived from a voltage-controlled oscillator. The demodulation circuit allows the separation of signals from positive and negative depths to better than 50 dB, thereby eliminating depth degeneracy and doubling the imaging depth range. Our system design is compatible with dual-balanced and polarization-diverse detection, important techniques in the practical biomedical application of optical frequency-domain imaging.

Analog-Digital Conversion↗