Automatic implantable defibrillator discharge resulting from routine pacemaker programming.

Abstract

A middle-aged gentleman with an idiopathic dilated cardiomyopathy, drug-refractory sustained ventricular arrhythmia, and high-degree AV block was managed with an automatic implantable cardioverter-defibrillator (AICD) and a Cordis Multicor II VVI pacemaker. During a routine follow-up visit, the pacemaker threshold was determined. Seven seconds after reprogramming the Cordis pacemaker to the "stat" VVI mode, the AICD discharged. The time to discharge after reprogramming the pacemaker equalled the previously determined AICD charge time. No other rhythm disturbance was documented and the rate with double counting of pacemaker stimulus and QRS complex was less than the triggering rate for the AICD. In conclusion, the AICD can be triggered by pacemaker programming signals unrelated to subsequent pacemaker function.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

C Gottlieb, J M Miller, M E Rosenthal, F E Marchlinski. 1988. Automatic implantable defibrillator discharge resulting from routine pacemaker programming.. https://doi.org/10.1111/j.1540-8159.1988.tb05013.x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A biventricular ICD system with biventricular defibrillation.

We describe the case of a 59-year-old gentleman with severe dilated cardiomyopathy requiring implantation of a dual-chamber biventricular implantable cardioverter-defibrillator (ICD). High defibrillation thresholds (DFT) were encountered at implant with an inadequate defibrillation safety margin. Testing of all possible shock vectors/polarities with and without the SVC coil and optimization of the distal RV coil position all proved inadequate. A satisfactory defibrillation safety margin was achieved following placement of a second lead in the coronary sinus to enable biventricular defibrillation. This case highlights an additional strategy for combating high DFTs and is an option even in dual-chamber biventricular ICD systems.

Electric Countershock↗