Search PubMedSearch

PubMed · 2254047

Automatic modules for extracorporeal circulation control.

Abstract

Maintenance of a constant level in the oxygenator during surgical extracorporeal circulation (ECC) is of paramount importance. We propose a device able to regulate the pump speed accordingly to the level in the oxygenator, and an automatic clamp to control venous return from the patient. The level regulator permits pump speed control and also stops the pump when a prefixed low level is reached (during ECC). Linearization of the level sensor has been made in order to optimize the PI regulator. Simulation results are presented and the regulator is evaluated in a clinical environment. The automatic clamp provides a good solution for reproducible and easy transitions in venous return. When coupled with the level regulator, the pump speed will increase automatically from zero to the correct equilibrium speed, with variable duration (from 30 seconds to 5 minutes) chosen by the operator. The same technique is used for stopping the ECC. These two devices are, in fact, modules, making up part of an evolutive device for ECC automation. The modules can exchange data with a supervising microcomputer.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

N Chauveau, W Van Meurs, R Barthelemy, J P Morucci. 1990. Automatic modules for extracorporeal circulation control.. https://pubmed.ncbi.nlm.nih.gov/2254047/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A robotics-based automated assay for inorganic and organic phosphates.

Phosphate analyses are fundamental to a broad range of biochemical applications involving inorganic phosphate and organic phosphoesters such as phospholipids, phosphorylated proteins, and nucleic acids. A practical automated method utilizing robotics is described in this report. Five colorimetric methods of phosphate analyses based on formation of a phosphomolybdate complex and compatible with the automated assay were tested, and the fundamental chemistry is discussed. The relative sensitivities are malachite green > crystal violet > quinaldine red > ascorbate reduction > antimony-modified ascorbate reduction, although only a fourfold improvement was observed in going from the modified ascorbate procedure to malachite green. Malachite green was selected to optimize the assay because this dye provided the highest sensitivity. However, where color stability and low blanks are more important than sensitivity, the ascorbate reduction and quinaldine red methods were found to be better choices than malachite green. Automation using a robotic liquid-handling system substantially reduces the labor required to process large arrays of samples. The result is a sensitive, nonradioactive assay of inorganic phosphate with high throughput. A digestion step in an acid-resistant 96-well plate was developed to extend the assay to phosphate esters. The robotic-based assay was demonstrated with inorganic phosphate and a common phospholipid, phosphatidylcholine.

Automation