PubMed · 16832467
Three-dimensional optic axis determination using variable-incidence-angle polarization-optical coherence tomography.
Abstract
Polarization optical coherence tomography (PSOCT) is a powerful technique to nondestructively map the retardance and fast-axis orientation of birefringent biological tissues. Previous studies have concentrated on the case where the optic axis lies on the plane of the surface. We describe a method to determine the polar angle of the optic axis of a uniaxial birefringent tissue by making PSOCT measurements with a number of incident illumination directions. The method is validated on equine flexor tendon, yielding a variability of 4% for the true birefringence and 3% for the polar angle. We use the method to map the polar angle of fibers in the transitional region of equine cartilage.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Nadezhda Ugryumova, Sergei V Gangnus, Stephen J Matcher. 2006-08-01. Three-dimensional optic axis determination using variable-incidence-angle polarization-optical coherence tomography.. https://doi.org/10.1364/ol.31.002305