A DNA microarray with 12,000 features was integrated with a microfluidic cartridge to automate the fluidic handling steps required to carry out a gene expression study of the human leukemia cell line (K562). The fully integrated microfluidic device consists of microfluidic pumps/mixers, fluid channels, reagent chambers, and a DNA microarray silicon chip. Microarray hybridization and subsequent fluidic handling and reactions (including a number of washing and labeling steps) were performed in this fully automated and miniature device before fluorescent image scanning of the microarray chip. Electrochemical micropumps were integrated into the cartridge to provide pumping of liquid solutions. The device was completely self-contained: no external pressure sources, fluid storage, mechanical pumps, mixers, or valves were necessary for fluid manipulation, thus eliminating possible sample contamination and simplifying device operation. Fluidic experiments were performed to study the on-chip washing efficiency and uniformity. A single-color transcriptional analysis of K562 cells with a series of calibration controls (spiked-in controls) to characterize this new platform with regard to sensitivity, specificity, and dynamic range was performed. The device detected sample RNAs with a concentration as low as 0.375 pM. Experiment also showed that the performance of the integrated microfluidic device is comparable with the conventional hybridization chambers with manual operations, indicating that the on-chip fluidic handling (washing and reaction) is highly efficient and can be automated with no loss of performance. The device provides a cost-effective solution to eliminate labor-intensive and time-consuming fluidic handling steps in genomic analysis.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Robin Hui Liu, Tai Nguyen, Kevin Schwarzkopf, H Sho Fuji, Alla Petrova, Tony Siuda, Kia Peyvan, Michael Bizak, David Danley, Andy McShea. 2006-03-15. Fully integrated miniature device for automated gene expression DNA microarray processing.. https://doi.org/10.1021/ac0518553

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Propagation of fluorescent viruses in growing plaques.

To study virus propagation, we have developed a method by which the propagation of the Lambda bacteriophage can be observed and quantified. This is done by creating a fusion protein of the capsid protein gpD and the enhanced yellow fluorescent protein (EYFP). We show that this fusion allows capsid formation and that the modified viruses propagate on a surface covered with host bacteria thus forming fluorescent plaques. The intensity of fluorescence in a growing plaque determines the distribution of phages. This provides a new tool to study the propagation of infection at the microscopic level.

Bacteriophage lambda↗

Combining optical tweezers and scanning probe microscopy to study DNA-protein interactions.

We present the first results obtained with a new instrument designed and built to study DNA-protein interactions at the single molecule level. This microscope combines optical tweezers with scanning probe microscopy and allows us to locate DNA-binding proteins on a single suspended DNA molecule. A single DNA molecule is stretched taut using the optical tweezers, while a probe is scanned along the molecule. Interaction forces between the probe and the sample are measured with the optical tweezers. The instrument thus enables us to correlate mechanical and functional properties of bound proteins with the tension within the DNA molecule. The typical friction force between a micropipette used as probe and a naked DNA molecule was found to be <1 pN. A 16 micro m DNA molecule with approximately 10-15 digoxygenin (DIG) molecules located over a 90 nm range in the middle of the DNA was used as a model system. By scanning with an antidigoxygenin (alpha-DIG) antibody-coated pipette we were able to localize these sites by exploiting the high binding affinity between this antibody-antigen pair. The estimated experimental resolution assuming an infinitesimally thin and rigid probe and a single alpha-DIG/DIG bond was 15 nm.

Bacteriophage lambda↗

A rapid DNA digestion system.

This paper presents a novel microfluidic DNA digestion system incorporating a high performance micro-mixer. Through the appropriate control of fixed and periodic switching DC electric fields, electrokinetic forces are established to mix the DNA and restriction enzyme samples and to drive them through the reaction column of the device. The experimental and numerical results show that a mixing performance of 98% can be achieved within a mixing channel of length 1.6 mm when a 150 V/cm driving voltage and a 5 Hz switching frequency are applied. The relationship between the mixing performance, switching frequency, and main applied electric field is derived. It is found that the optimal switching frequency depends upon the magnitude of the main applied electric field. The successful digestion of lambda-DNA using Eco RI restriction enzyme is demonstrated. The DNA-enzyme reaction is completed within 15 min in the proposed microfluidic system, compared to 50 min in a conventional large-scale system. Hence, the current device provides a valuable tool for rapid lambda-DNA digestion, while its mixer system delivers a simple yet effective solution for mixing problems in the micro-total-analysis-systems field.

Bacteriophage lambda↗