PubMed · 160513
Evaluation of electrode array material for neural prostheses.
Abstract
Matrix support materials for brain surface electrodes used in neuroprosthetic applications were evaluated after chronic subdural implantation over the parietal cortex of the cat. Four types of array fabricated with Silastic, Dacron mesh, or platinum wire annuli were implanted for periods ranging from 5 weeks to 1 year. We evaluated the arrays by access resistance measurements and gross and histological observations of the tissue beneath both nonstimulated and stimulated electrodes. A porous type matrix constructed of Dacron mesh proved to be the superior design because of its minimal compression of the cortical surface, facility of handling during implantation and autopsy, and satisfactory electrical characteristics provided by a good electrode-brain interface. (Neurosurgery, 5: 681--686, 1979).
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
L A Bullara, W F Agnew, T G Yuen, S Jacques, R H Pudenz. 1979. Evaluation of electrode array material for neural prostheses.. https://doi.org/10.1227/00006123-197912000-00006. https://searchpubmed.org/record/pubmed%3A160513/evaluation-of-electrode-array-material-for-neural-prostheses
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search PubMed by Indiciae: BibTeX · RIS.