Abstract
Repeated samplings and measurements in the monitoring of patients to look for changes are common clinical problems. The "reference change value", calculated as zp x [2 x (CVI2 + CVA2)](1/2), where zp is the z-statistic and CVI and CVA are within-subject and analytical coefficients of variation, respectively, has been used to detect whether a measured difference between measurements is statistically significant. However, a reference change value only detects the probability of false-positives (type I error), and for this reason, a model to calculate the risk of missing significant changes in serial results from individuals (probability of false-negatives) is investigated in this work by means of power functions. Therefore, when an analyte is being monitored in a patient, power functions estimate the probability of detecting a defined real change by measuring the difference. Thus, when a measured difference is the same as the calculated reference change value, then it will be detected in only 50% of situations.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Natàlia Iglesias Canadell, Per Hyltoft Petersen, Esther Jensen, Carmen Ricós, Per E Jørgensen. 2004. Reference change values and power functions.. https://doi.org/10.1515/cclm.2004.073
Cite the original work for its findings. Save a collection to share your selection of sources.