Amount of air infused to patient increases as fluid flow rates decrease when using the Hotline HL-90 fluid warmer.

Abstract

OBJECTIVE: The intraoperative use of fluid warming devices has been recommended to avoid perioperative hypothermia and related adverse outcomes. To evaluate whether these devices might introduce risks of their own, we measured the volume of air escaping from a warmed intravenous solution that might be delivered to a patient. METHODS: In an operating room maintained at 19-19.5 degrees C, we tested an HL-90 Hotline fluid warmer with the L-70 fluid-warming set. One liter of lactated Ringer's solution was infused at flow rates of 150, 300, 500 and 3400 ml/h. The air that formed within the L-70 tubing during infusion was collected in a bubble trap placed at the end of the L-70 tubing. The volume of air in the bubble trap was measured. Twelve separate measurements were obtained at each flow rate. One additional study (n = 8) was performed using the L-10 Gas Vent to determine whether this equipment might reduce the volume of air infused when fluid flow rate was 300 mL/h. The volume of air collected at each flow rate was compared using ANOVA. RESULTS: Volume of air increased significantly from 1.0 +/- 0.2 mL to 2.9 +/- 0.4 ml as flow rate decreased from 3400 ml/h to 150 ml/h (p < 0.0001). The L-10 gas eliminator was ineffective in reducing the amount of air infused. CONCLUSIONS: We conclude that the use of the Hotline fluid warmer can result in infusion of air into the patient, introducing possible risk of air embolism.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

S Woon, P Talke. 1999. Amount of air infused to patient increases as fluid flow rates decrease when using the Hotline HL-90 fluid warmer.. https://doi.org/10.1023/a%3A1009953614884

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Ultrasound protocols used to detect vascular gas emboli in divers: a systematic review.

INTRODUCTION: Venous gas emboli (VGE) detected via ultrasound can be used as a surrogate marker for decompression stress. While Doppler ultrasound is the historical gold standard, two-dimensional (2D) ultrasonography offers advantages for on-site monitoring, including a wider field of view and reduced dependence on noise-free environments. This systematic review evaluates 2D ultrasonography protocols used in decompression research since the 2015 International Meeting on Ultrasound for Diving Research, identifying methodological similarities, differences, and adherence to consensus recommendations. METHODS: A search of PubMed and Scopus identified studies using 2D ultrasound to detect VGE in divers. Inclusion criteria were: (1) use of 2D ultrasound, (2) detection of VGE or monitoring of decompression stress, (3) inclusion of a diver cohort, and (4) publication after 2015. Data extraction focused on VGE scoring systems, ultrasound hardware, measurement protocols, and operator experience. Risk of bias was assessed using ROBINS-I-V2, and compliance with the 2015 consensus recommendations was evaluated. RESULTS: Twenty studies were included. The Eftedal-Brubakk scale was most commonly used (n = 15), with cardiac ultrasound as the primary imaging modality; one study assessed a peripheral vessel. Common shortcomings included post-dive measurements lasting less than two hours, underreporting of operator experience and hardware specifications, limited individual-level data, and inappropriate use of parametric statistics for ordinal bubble grade data. No study fully complied with all consensus recommendations. CONCLUSIONS: This review demonstrates that, although two-dimensional ultrasound is widely used for post-dive VGE assessment, methodological heterogeneity with multiple shortcomings remain. Furthermore, nearly all studies restricted imaging to the heart, thus leaving peripheral vessel assessment largely unexplored.

Embolism, Air↗

Infant death due to air embolism from peripheral venous infusion.

An otherwise healthy male infant was brought to the hospital because the mother suspected superficial infection at the operative site 5 days after an inguinal hernia repair. He was admitted to the pediatric unit overnight to be evaluated by his surgeon the next morning. When a venous infusion of maintenance fluids was started, the patient immediately went into cardio-respiratory arrest and was pronounced dead after resuscitation efforts failed. Subsequently, air collections were found in both venous and arterial circulations, including the splenoportal system. Detailed review of the clinical presentation and course, laboratory results, radiological, and pathological findings, along with a review of pertinent literature provides an explanation for the death by air embolism. Apparent inconsistent findings both radiographically and at autopsy are resolved. The mechanism of distribution of air to both systemic and splenoportal circulation is discussed. We believe this to be only the eighth case reported in English-language literature of infantile death from peripheral venous infusion. In all age groups, we find only six other cases in the English-language literature of gas found concomitantly in both the systemic and portal venous systems.

Embolism, Air↗