Abstract

It is commonly accepted that TrueFISP (balanced FFE, FIESTA) belongs to the class of gradient-echo (GRE) sequences. GRE sequences are sensitive to dephasing effects of the transverse magnetization between the excitation pulse and echo acquisition, and phase coherence is only established directly after and before excitation pulses. However, an analysis of the phase evolution of transverse magnetization in a TrueFISP experiment shows very close similarities to the echo formation of a spin-echo (SE) experiment. If dephasing between excitation pulses is below +/-pi, TrueFISP exhibits a nearly complete refocusing of transverse magnetization at TE = TR/2. Only signals acquired before and after TR/2 show an additional T*2 sensitivity.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Klaus Scheffler, Jürgen Hennig. 2003. Is TrueFISP a gradient-echo or a spin-echo sequence?. https://doi.org/10.1002/mrm.10351

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Ultra-high-field 7T MRI reveals neural abnormalities of attention networks in relation to cognitive impairment in hypertension.

Hypertension is a significant risk factor for cognitive impairment (CI), yet the corresponding neural network abnormalities remain underexplored. In this study, we examined the associations among global and domain-specific cognitive dysfunction, neuroimaging measures, and blood pressure in a subgroup of hypertensive patients with CI (N = 41) from a randomized controlled trial who underwent ultra-high-field 7 T MRI. Structural atrophy related to CI was localized to regions overlapping the attention networks. Both whole-brain and within-network dysfunction of the attention networks were associated with worse global cognitive performance. Notably, hyperconnectivity within key attention network hubs, including the right anterior insula and posterior intraparietal sulcus, was associated with declined processing speed in hypertensive patients, mediating the association between pulse pressure and processing speed. These findings provide new insights into the neural pathophysiology of hypertension-related CI and suggest potential network-based targets for intervention.

Magnetic Resonance Imaging↗

Detecting stiff masses using strain-encoded (SENC) imaging.

A method is proposed for detecting stiff masses using strain-encoded (SENC) magnetic resonance imaging (MRI). An object of interest is compressed to produce local strain distribution that depends on local elasticity, where intensities correlate with the local through-imaging-plane strain component. Because the strain is lower inside a stiff mass than in the surrounding soft tissue, an intensity contrast in the resulting images would enable direct detection of the mass without postprocessing. The technique was validated by a phantom experiment in which a gel phantom with a stiff region was used. The advantages of the proposed method include short imaging time and uncomplicated postprocessing. However, in its current form the technique does not measure elasticity.

Magnetic Resonance Imaging↗

An analytical solution for the SSFP signal in MRI.

Among previous analyses of the steady-state free-precession (SSFP) signal in rapid MRI, one treatment resulted in equations that require the evaluation of infinite binomial series. Here, an analytical solution is derived by a transformation into the power series expansion of the derivative of the inverse sine function, which is essentially a root. The treatment is extended to include higher-order signals. The results demonstrate the identity of the vastly different equations for the SSFP signals reported so far. Applications consist of the derivation of closed expressions for the signal in echo-shifted MRI and a corresponding analysis of TrueFISP sequences.

Magnetic Resonance Imaging↗