Coronary flow velocity monitoring after angioplasty associated with abrupt reocclusion.
Explore the source record for details and available documents.
Biomedical subjects
Publications and source records attributed to R G Bach.
Explore the source record for details and available documents.
The purpose of this study was to evaluate the incidence of ventricular ectopy and catheter movement during left ventriculography with a new 5F halo angiographic catheter that has a unique helical-tip design unlike the design of standard 5F and 6F pigtail catheters. The pigtail catheter is presently preferred for left ventriculography, although its use is associated with a high incidence of ventricular ectopy, which often limits precise interpretation of data. In this study, 155 patients (in 145 unpaired and 10 paired studies) underwent left ventriculography during diagnostic cardiac catheterization. In the unpaired group, the 5F Halo catheter was used in 63 studies and standard 5F and 6F pigtail catheters in 40 and 42 studies, respectively. An additional 10 patients had two consecutive left ventriculograms with 5F Halo and pigtail catheters. Ventriculograms were performed with the same technique in the 30-degree right anterior oblique projection. The left ventricle was divided into a basal zone, midzone, and apical zone. Catheter movement within the ventricle was scored as significant if there was at least one zone change. Ventricular ectopy was quantified by a simultaneous electrocardiographic recording during contrast injection. There were no significant differences in the left ventricular systolic or end-diastolic pressures, left ventricular score, or diagnostic quality of the ventriculograms between the 5F Halo catheter group and the 5F and 6F pigtail catheter groups. Mean ventricular ectopy with the 5F Halo catheter was significantly less (0.9 +/- 1.4 ventricular premature beats [VPBs]) than with the 5F pigtail catheter (2.3 +/- 2.5 VPBs, p < 0.001) or the 6F pigtail catheter (2.9 +/- 2.9 VPBs, p < 0.001).(ABSTRACT TRUNCATED AT 250 WORDS)
Explore the source record for details and available documents.
Continuous measurement of blood flow velocity during interventional procedures has the potential to provide an early warning of coronary flow instability, which can lead to abrupt closure or other adverse events before angiography. The magnitude and fluctuations of the average velocity over time (trend) was studied by using a 0.018-inch Doppler-tipped angioplasty guide wire in 32 patients after coronary angiography (n = 20), atherectomy (n = 2), urgent stent (n = 6), urgent vein graft thrombolysis (n = 4), or acute myocardial infarction (n = 2). The patients (mean age 60 +/- 11 years) had postprocedural in-laboratory flow monitoring for a mean of 19 +/- 11 (range 8 to 36) minutes. The coronary artery monitored was the left anterior descending in 13, circumflex in 6, right coronary artery in 9, and saphenous vein graft in 4. Seven patients had flow-related events during continuous flow velocity monitoring before serial angiographic study. These events included coronary vasospasm (abrupt flow acceleration), vasovagal flow cessation, cyclical flow variations resulting from accumulation of intraluminal thrombus, and rapid decline of flow velocity. The last two patterns were associated with abrupt vessel closure during angioplasty. Continuous flow velocity monitoring is easily incorporated into routine interventional procedures and provides an early indication of unstable flow and the potential for abrupt vessel closure and other adverse events.
BACKGROUND: The physiological assessment of angiographically intermediate-severity stenoses remains problematic. Functional measurements of poststenotic intracoronary Doppler coronary flow reserve can be performed in humans but have not been correlated with hyperemic myocardial perfusion imaging or angiographic data in this patient population. METHODS AND RESULTS: Thirty-three patients undergoing diagnostic quantitative coronary angiography (QCA) for assessment of intermediate-severity coronary artery disease (mean QCA percent diameter stenosis, 56 +/- 14%) were studied. Proximal and distal poststenotic Doppler coronary flow velocities were measured (left anterior descending coronary artery, 16; right coronary artery, 10; left circumflex artery, 7 patients) before and during peak maximal hyperemia with intracoronary adenosine (8 to 12 micrograms). Intravenous pharmacological stress (adenosine, 20 patients; dipyridamole, 13 patients) 99mTc-sestamibi tomographic perfusion imaging was performed within 1 week of coronary flow-velocity studies. kappa statistics were calculated to measure the strength of correlation among coronary flow velocities, perfusion imaging data, and QCA results. QCA stenosis severity (abnormal, > or = 50% diameter stenosis) and poststenotic Doppler coronary flow reserve (ratio of abnormal distal hyperemic to basal flow, < or = 2.0) were correctly correlated in 20 of 27 patients (74%; kappa = .48). QCA stenosis severity and 99mTc-sestamibi imaging (abnormal if one or more reversible myocardial segments were present in the poststenotic zone) were correlated in 28 of 33 patients (85%; kappa = .63). 99mTc-sestamibi imaging results agreed with the basal (nonhyperemic) proximal-to-distal velocity ratio (normal, < 1.7) in 15 of 31 patients (48%; kappa = .17). The strongest correlation occurred between hyperemic distal flow-velocity ratio measurements and 99mTc-sestamibi perfusion imaging results in 24 of 27 patients (89%; kappa = .78). All 14 patients with abnormal distal hyperemic flow-velocity values had corresponding reversible 99mTc-sestamibi tomographic defects. More reversibly hypoperfused segments were present in patients with abnormal poststenotic hyperemic flow-velocity ratios (abnormal, 2.4 +/- 0.7 segments; normal, 0.6 +/- 1.0 segments; P < .05). The number of poststenotic myocardial 99mTc-sestamibi perfusion defects was correlated with the QCA percent cross-sectional area reduction (P < .02) and with minimal luminal diameter (P < .05) of intermediate-severity coronary artery stenoses. CONCLUSIONS: Two technologically diverse functional measures of stenosis severity--Doppler-derived poststenotic hyperemic intracoronary flow reserve and vasodilator stress 99mTc-sestamibi myocardial perfusion imaging--are highly (89%) correlated. The physiological assessment of coronary stenoses of angiographically intermediate severity may be improved by the use of these techniques.
The effect of balloon compliance on the safety and outcome of percutaneous transluminal coronary angioplasty (PTCA) is controversial. It has been proposed that PTCA balloons constructed from compliant polymers contribute to an increased risk of angiographic coronary dissection and potentially to adverse clinical results. To determine the effect of balloon material compliance on PTCA outcome, 1,076 PTCA procedures using balloons differing in compliance characteristics (polyethylene teraphthalate [noncompliant], polyethylene [intermediately compliant] or polyolefincopolymer [compliant]) were retrospectively analyzed. Baseline clinical, angiographic and procedural variables were similar among the 3 balloon material groups. In the noncompliant, intermediately compliant and compliant groups, the occurrence rates of intimal tears (10, 14 and 10%, respectively; p = NS for all comparisons) and coronary dissection (7, 9 and 8%, respectively; all p = NS) were also equivalent. The rates of acute occlusion, myocardial infarction, emergency bypass surgery and death were low and similar, and likewise unaffected by balloon material. The results provide evidence that the occurrence of adverse outcomes after PTCA is unrelated to balloon material compliance and support the clinical safety of balloons constructed of compliant or noncompliant polymers when used for appropriate coronary interventions.
Explore the source record for details and available documents.
Determination of the clinical and hemodynamic significance of coronary stenoses is often difficult and inexact. Angiography and coronary vasodilator reserve have been shown to be imperfect tools to determine the physiologic significance of coronary stenoses. Spectral flow velocity data, both proximal and distal to coronary stenoses, using an 0.018-in intracoronary Doppler-tipped angioplasty guidewire, were compared to translesional pressure gradients and angiography during cardiac catheterization. Patients were divided into 2 groups based on resting translesional gradients: Group 1 had gradients < 20 mm Hg and group 2 had gradients > or = 20 mm Hg. Proximal average peak velocity, diastolic velocity integral, and total velocity integral were statistically significantly lower in Group 1. The distal average peak velocity, and diastolic and total velocity integrals were all significantly (p < 0.01) decreased in patients with gradients > 20 mm Hg (group 2). The ratio of proximal-to-distal total flow velocity integral was also higher in group 2 patients (2.3 +/- 0.9) compared with group 1 (1.1 +/- 0.2; p < 0.001). There was a strong correlation between translesional pressure gradients and the ratios of the proximal-to-distal total flow velocity integrals (r = 0.8, p < 0.001) with a weaker relationship between quantitative angiography and pressure gradients (r = 0.6, p < 0.001). Angiography was a poor predictor of translesional gradients in angiographically intermediate stenoses (range 50-70%; r = 0.2, p = NS), while the flow velocity ratios continued to have a strong correlation (r = 0.8, p < 0.0001).(ABSTRACT TRUNCATED AT 250 WORDS)
Quantitation of coronary collateral flow in patients has been limited to angiographic techniques, which are subject to well-known methodologic limitations. The use of a Doppler-tipped angioplasty guidewire permits measurement of both antegrade and retrograde flow distal to totally or subtotally occluded vessels that may be supplied with acutely recruitable or angiographically mature collateral conduits. Using coronary flow velocity as an indicator of collateral flow, retrograde flow velocity was quantitated in 17 patients. Mean collateral flow velocity was approximately 30% of normal postangioplasty antegrade flow velocity. The phasic pattern of collateral flow was highly variable, but the retrograde diastolic and systolic flow velocity integrals were 20% and 40% (respectively) of post-procedure antegrade flow velocity. Preliminary studies with pharmacologic stimulation of the contralateral supply artery suggests that collateral flow is not increased by intracoronary nitroglycerin (200 micrograms) or adenosine (12 micrograms), but may be markedly augmented during mechanical stimulation of balloon occlusion. These data represent the first in a series of quantitative observations on control of the coronary collateral circulation in humans. Future investigations using the Doppler Flowire (Cardiometrics) will enhance understanding of factors modulating ischemia through collateral supply.
Translesional pressure gradients are currently not performed routinely for most angioplasty procedures because of the technical limitations. With the use of an 0.18 inch Doppler flow velocity guidewire, coronary flow velocity, both proximal and distal to a lesion, can be easily assessed. In branching arteries, significant lesions are characterized by a ratio of proximal to distal flow velocity of > 1.7, loss of the normal phasic diastolic predominant flow velocity pattern, and/or loss of distal hyperemia. We describe the use of coronary flow velocity in assisting important decision making in a young patient with a recent myocardial infarction and multiple coronary lesions. The decision for angioplasty of the hemodynamically significant stenosis was confirmed by translesional flow velocity measurements. A rational approach to coronary intervention in patients with multiple stenoses of intermediate severity appears to be facilitated by direct measurement of translesional flow dynamics.
A wire perforation of a totally occluded right coronary artery was treated successfully with prolonged balloon inflation. This non-surgical management is preferred if physiologic and anatomic conditions are suitable.
In the course of studying the effects of coronary angioplasty on branch vessel flow using two Doppler flow velocity guidewires, we quantitated simultaneous blood flow responses proximal and distal to a stenosis. The alterations of flow documented a horizontal epicardial steal induced during dipyridamole hyperemia, hyperemic flow reversal by intravenous aminophylline, and subsequent normalization of distal hyperemia after endoluminal enlargement by successful angioplasty. The quantitative physiology of the patient described here confirms one postulated mechanism of abnormal myocardial perfusion stress scintigraphy. Continuous dual flowire spectral coronary flow determinations appear to be a valuable method in verifying postulated mechanisms of various pharmacologic and mechanical stimuli influencing coronary blood flow in patients with atherosclerotic coronary artery disease.
Ostial lesions of saphenous vein grafts can be difficult to assess by angiography. A physiologic approach to the selection and recanalization of ostial lesions in saphenous vein grafts may be advantageous to overcome limitations of angiography. To assist in identifying favorable physiology and facilitate procedural decision making, the use of coronary flow velocity measurements with a Doppler-tipped 0.018-inch angioplasty flowire in three particularly difficult patients with varying types of saphenous vein graft aorto-ostial narrowings is reported.
Explore the source record for details and available documents.
Explore the source record for details and available documents.
Explore the source record for details and available documents.
Explore the source record for details and available documents.
OBJECTIVES: The purpose of this study was to examine the relation among the angiographic severity of coronary artery lesions, coronary flow velocity and translesional pressure gradients. BACKGROUND: Determination of the clinical and hemodynamic significance of coronary artery stenoses is often difficult and inexact. Angiography has been shown to be an imperfect tool for determining the physiologic significance of eccentric or irregular coronary lumen narrowing. METHODS: Using a 0.018-in. (0.046 cm) intracoronary Doppler-tipped angioplasty guide wire, spectral flow velocity data both proximal and distal to coronary stenoses were compared with translesional pressure gradient measurements and angiographic data obtained during cardiac catheterization in 101 patients. There were 17 patients with normal angiographic findings and 84 with coronary artery disease, with lesions ranging from 28% to 99% diameter narrowing. Patients with coronary disease were assigned to two groups on the basis of translesional gradients at rest. Group A (n = 56) had gradients < 20 mm Hg, and Group B (n = 28) had gradients > or = 20 mm Hg. RESULTS: Proximal average peak velocity, diastolic velocity integral and total velocity integral were slightly but statistically lower in Group A; however, the distal average peak velocity and diastolic and total velocity integrals were all markedly (all p < 0.01) decreased in patients with gradients > or = 20 mm Hg (Group B). In addition, the ratio of proximal to distal total flow velocity integral was higher in Group B (2.4 +/- 1.0) than in group A (1.1 +/- 0.3, p < 0.001). There was a strong correlation between translesional pressure gradients and the ratios of the proximal to distal total flow velocity integrals (r = 0.8, p < 0.001), with a weaker relation between quantitative angiography and pressure gradients (r = 0.6, p < 0.001). In angiographically intermediate stenoses (range 50% to 70%), angiography was a poor predictor of translesional gradients (r = 0.2, p = NS), whereas the flow velocity ratios continued to have a strong correlation (r = 0.8, p < 0.001). Only two patients with a proximal/distal total flow velocity ratio < 1.7 had a translesional gradient > 30 mm Hg. Both patients had a very proximal lesion in a nonbranching right coronary artery. CONCLUSIONS: These data demonstrate that in branching human coronary arteries, a close relation exists between translesional hemodynamics and distal coronary flow velocity. Translesional coronary flow velocity is a new and easily applicable method for determining the hemodynamic significance of coronary artery stenoses that is superior to angiography and can be applied at the time of diagnostic catheterization. These data will provide a rational approach to making decisions on the use of coronary interventional techniques when angiographic findings are questionable.