Validation of a digital color Doppler flow measurement method for pulmonary regurgitant volumes and regurgitant fractions in an in vitro model and in a chronic animal model of postoperative repaired tetralogy of Fallot.

OBJECTIVES: The purpose of this study was to validate a digital color Doppler (DCD) automated cardiac flow measurement method for quantifying pulmonary regurgitation (PR) in an in vitro and a chronic animal model of the right ventricular outflow tract of postoperative tetralogy of Fallot (TOF). BACKGROUND: There has been no reliable ultrasound method that can accurately quantitate PR. METHODS: We developed an in vitro model of mild pulmonary stenosis and wide-open PR that mimics the patterns of flow seen in patients with postoperative TOF. Thirteen different forward and regurgitant stroke volumes (RSVs) across the noncircular shaped cross-sectional outflow tract flow area were estimated using the DCD method in two orthogonal planes. In six sheep with surgically created PR, 24 different hemodynamic states with PR strictly quantified by electromagnetic probes were also studied. RESULTS: The RSVs and regurgitant fractions (RFs) obtained by the DCD method using average values from two orthogonal planes correlated well with reference values (RSV: r = 0.99, mean difference = 0.02 +/- 0.39 ml/beat for in vitro model; r = 0.97, mean differences = 1.79 +/- 1.84 ml/beat for animal model, RF: r = 0.98, mean difference = -1.10 +/- 4.34% for in vitro model; r = 0.94, mean difference = 2.73 +/- 6.75% for animal model). However, the DCD method using a single plane had limited accuracy for estimating pulmonary RFs and RSVs. CONCLUSIONS: The DCD method using average values from two orthogonal planes provides accurate estimation of RSVs and RFs and should have clinical importance for serially quantifying PR in patients with postoperative TOF.

Animals↗

Effect of left ventricular wall mass on Doppler filling patterns in the developing normal human heart.

To assess gestational age- and growth-related changes in left ventricular (LV) size, LV wall, and LV transmitral flow velocity patterns, 2-dimensional (2-D) and Doppler echocardiographic studies were performed in 89 normal fetuses aged 16 to 38 weeks. Serial studies were designed in 7 fetuses. Variables measured from 4-chamber views were chamber areas and myocardial wall areas. From these measurements, area shortening fraction and ratio of myocardial wall area to end-diastolic chamber area were calculated. LV end-diastolic chamber area and myocardial wall area increased exponentially with advancing gestational age (r = 0.88 and 0.90, respectively, p < 0.001). Area shortening fraction showed no significant changes with gestational age. Ratio of myocardial wall area to LV end-diastolic chamber area decreased gradually with increasing gestational age (r = -0.77, p < 0.001). With increasing gestational age, mitral peak velocities of early diastole increased (r = 0.82, p < 0.01) with little change in peak velocity during atrial contraction. Multiple regression analysis showed that age-related increases in peak velocity of early diastole were related to advancing gestational age and also to decreases in ratio of myocardial wall area to LV end-diastolic chamber area. Low peak filling velocities during early diastole in younger fetuses may be related partly to relative increase in LV wall mass. The gestational age-related decreases in LV wall mass may be one of the important mechanisms of gestational age-related alterations in diastolic properties, especially relaxation processes.

Blood Flow Velocity↗

A digital 3-dimensional method for computing great artery flows: in vitro validation studies.

BACKGROUND: Conventional 2-dimensional Doppler large vessels are prone to inaccuracy. Three-dimensional (3D) volume imaging provides the opportunity to make cross-sectional flow calculations through digital spatiotemporal integration of flow velocity, area, and profile. METHODS: A new digital 3D color Doppler reconstruction method was used to generate radially acquired flow data sets. Raw scanline data with digital velocity assignments, obtained by scanning parallel to flow, were transferred from a specially programmed but otherwise conventional ultrasonographic system, which controlled a multiplane transesophageal probe, to a computer workstation via an Ethernet link for assimilation into color 3D data sets. This configuration was used to study 20 pulsatile laminar flows (stroke volumes 30 to 70 mL and peak flow rates 65 to 205 mL/s) in a curved tube model with an oval cross-sectional geometry. After generation of the color 3D data set, flow velocity values from cross sections perpendicular to the tubes were analyzed to determine flow rate and stroke volume. RESULTS: The flows from 3D digital velocity profiles showed close correlation with peak instantaneous flow rates (r = 0.99, y = 1.01x-0.9, standard error of estimate 4.1 mL/s). When interpreted with pulsed wave Doppler data obtained through the cardiac cycle, they also allowed computation of stroke volume (r = 0.98, y = 1.44x-2.5, standard error of estimate 3.8 mL). CONCLUSION: The ability to compute laminar flows from 3D digital data sets obtained parallel to the direction of flow and without the need for geometric assumptions represents an important opportunity for and advantage of 3D color Doppler echocardiography.

Coronary Vessels↗

Comparison of ventricular volume and mass measurements from B- and C-scan images with the use of real-time 3-dimensional echocardiography: studies in an in vitro model.

BACKGROUND: Real-time 3-dimensional (3D) echocardiography avoids geometric assumptions in volume analysis and permits immediate visualization in any plane without the need for cardiac or respiratory gating or computation time. This study compared the accuracy of volume and mass assessments between standard long-axis (B-scan) and short-axis (C-scan) views in a simplified but quantifiable left ventricular phantom. METHODS AND RESULTS: The model comprised an inner balloon within an outer balloon separated by ultrasonographic gel. First, to mimic different chamber volumes, 12 volumes (40 to 180 mL) of water within the inner balloon were scanned with a real-time 3D system. Second, 10 volumes (80 to 170 mL) of gel were inserted between the balloons to mimic varying cardiac mass, and the gel volume space (mass) was calculated by subtracting the inner from the outer balloon volume. "Chamber" and "mass" measurements for both B and C scans correlated closely with the actual values (r = 0.99). However, chamber volumes from C scans were consistently less than B-scan values (mean difference from reference for C scans: -5.2 +/- 1.2 mL, P <.0001; for the 2 orthogonal B scans: 0.03 +/- 1.4 mL and -0.9 +/- 1.5 mL, respectively, P = NS). Similarly, for gel volume measurements, B-scan results were closer to actual mass volumes (mean difference 0. 3 +/- 2.5 and 1.7 +/- 2.9 mL) than those of C scans, which tended to underestimate (-4.5 +/- 2.5 mL, P <.0001). CONCLUSION: Our study suggests that real-time 3D echocardiography should provide an accurate means of determining chamber volumes and cardiac mass. However, measurements performed from B-scan views may be closer to the actual values than those from C-scan views, presumably since they are less highly influenced by distortions related to lateral resolution.

Cardiac Volume↗

Clinical use of Real-time three-dimensional echocardiography in pediatric cardiology.

Real-time three-dimensional echocardiography is a technique that allows three-dimensional imaging without any geometrical assumptions or need for reconstruction from two-dimensional images. This imaging technique should be particularly useful in the pediatric population, for whom a complex cardiac anatomy must be defined. We review the potential usefulness of real-time three-dimensional imaging in the pediatric population and the potential limitations associated with this technique.

Cardiology↗

New echocardiographic windows for quantitative determination of aortic regurgitation volume using color Doppler flow convergence and vena contracta.

Color Doppler images of aortic regurgitation (AR) flow acceleration, flow convergence (FC), and the vena contracta (VC) have been reported to be useful for evaluating severity of AR. However, clinical application of these methods has been limited because of the difficulty in clearly imaging the FC and VC. This study aimed to explore new windows for imaging the FC and VC to evaluate AR volumes in patients and to validate this in animals with chronic AR. Forty patients with AR and 17 hemodynamic states in 4 sheep with strictly quantified AR volumes were evaluated. A Toshiba SSH 380A with a 3.75-MHz transducer was used to image the FC and VC. After routine echo Doppler imaging, patients were repositioned in the right lateral decubitus position, and the FC and VC were imaged from high right parasternal windows. In only 15 of the 40 patients was it possible to image clearly and measure accurately the FC and VC from conventional (left decubitus) apical or parasternal views. In contrast, 31 of 40 patients had clearly imaged FC regions and VCs using the new windows. In patients, AR volumes derived from the FC and VC methods combined with continuous velocity agreed well with each other (r = 0.97, mean difference = -7.9 ml +/- 9.9 ml/beat). In chronic animal model studies, AR volumes derived from both the VC and the FC agreed well with the electromagnetically derived AR volumes (r = 0.92, mean difference = -1.3 +/- 4.0 ml/beat). By imaging from high right parasternal windows in the right decubitus position, complementary use of the FC and VC methods can provide clinically valuable information about AR volumes.

Adolescent↗

Three-dimensional reconstruction of the color Doppler-imaged vena contracta for quantifying aortic regurgitation: studies in a chronic animal model.

BACKGROUND: The purpose of this study was to investigate the use of 3-dimensional (3D) reconstruction of color Doppler flow maps to image and extract the vena contracta cross-sectional area to determine the severity of aortic regurgitation (AR) in an animal model. Evaluation of the vena contracta with 2-dimensional imaging systems may not be sufficiently robust to fully characterize this region, which may be asymmetrically shaped. METHODS AND RESULTS: In 6 sheep with surgically induced chronic AR, 18 hemodynamically different states were studied. Instantaneous regurgitant flow rates were obtained by aortic and pulmonary electromagnetic flowmeters (EMFs) as reference standards, and aortic regurgitant effective orifice areas (EOAs) were determined from EMF regurgitant flow rates divided by continuous-wave (CW) Doppler velocities. Composite video data for color Doppler imaging of the aortic regurgitant flows were transferred into a TomTec computer after computer-controlled 180 degrees rotational acquisition. After the 3D data transverse to the flow jet were sectioned, the smallest proximal jet cross section was identified for direct measurement of the vena contracta area. Peak regurgitant flow rates and regurgitant stroke volumes were calculated as the product of these areas and the CW Doppler peak velocities and velocity-time integrals, respectively. There was an excellent correlation between the 3D-derived vena contracta areas and reference EOAs (r=0.99, SEE=0.01 cm2) and between 3D and reference peak regurgitant flow rates and regurgitant stroke volumes (r=0.99, difference=0.11 L/min; r=0.99, difference=1.5 mL/beat, respectively). CONCLUSIONS: 3D-based determination of the vena contracta cross-sectional area can provide accurate quantification of the severity of AR.

Animals↗

Validation of the accuracy of both right and left ventricular outflow volume determinations and semiautomated calculation of shunt volumes through atrial septal defects by digital color Doppler flow mapping in a chronic animal model.

OBJECTIVES: The aim of the present study was to quantitate shunt flow volumes through atrial septal defects (ASDs) in a chronic animal model with surgically created ASDs using a new semiautomated color Doppler flow calculation method (ACM). BACKGROUND: Because pulsed Doppler is cumbersome and often inappropriate for color flow computation, new methods such as ACM are of interest. METHODS: In this study, 13 to 25 weeks after ASDs were surgically created in eight sheep, a total of 24 hemodynamic states were studied at a separate open chest experimental session. Electromagnetic (EM) flow probes and meters were used to provide reference flow volumes as the pulmonary and aortic flow volumes (Qp and Qs) and shunt flow volumes (Qp minus Qs). Epicardial echocardiographic studies were performed to image the left and right ventricular outflow tract (LVOT and RVOT) forward flow signals. The ACM method digitally integrated spatial and temporal color flow velocity data to provide stroke volumes. RESULTS Left ventricular outflow tract and RVOT flow volumes obtained by the ACM method agreed well with those obtained by the EM method (r = 0.96, mean difference = 0.78 +/- 1.7 ml for LVOT and r = 0.97, mean difference = -0.35 +/- 3.6 ml for RVOT). As a result, shunt flow volumes and Qp/Qs by the ACM method agreed well with those obtained by the EM method (r = 0.96, mean difference = -1.1 +/- 3.6 ml/beat for shunt volumes and r = 0.95, mean difference = -0.11 +/- 0.22 for Qp/Qs). CONCLUSIONS: This animal study, using strictly quantified shunt flow volumes, demonstrated that the ACM method can provide Qp/Qs and shunt measurements semiautomatically and noninvasively.

Animals↗

Accuracy of Doppler methods for estimating peak-to-peak and peak instantaneous gradients across coarctation of the aorta: An In vitro study.

Although data exist that address the attempt to correlate noninvasive Doppler-derived pressure gradients with invasive catheter pressure gradients in patients with coarctation of the aorta, few data exist about stiffness of the proximal descending aorta (precoarctation) and its relation to these pressure measurements. In this study, an in vitro flow model of a simulated neonatal aorta with a coarctation was developed. Three proximal descending aortas of different stiffnesses were used. The stiffness index of the proximal descending aorta was calculated as beta = ln [systolic pressure/diastolic pressure/(systolic diameter - diastolic diameter)]. We evaluated pressure gradients obtained by continuous wave Doppler and standard catheter methods and looked at acceleration of flow velocity determined by pulsed wave Doppler in the 3 precoarctation segments of differing stiffnesses. Pressures in the proximal descending aorta (precoarctation) increased with increasing stiffness, ranging from 105 mm Hg (soft) to greater than 300 mm Hg (stiff). Continuous wave Doppler instantaneous pressure gradients overestimated the catheter instantaneous pressure gradients substantially (mean 41% +/- 19%). The stiffer the precoarctation segment, the more the degree of overestimation: soft, 0% to 63% (= 3.47); medium, 13% to 54% (beta = 4.42); and stiff, 43% to 66% (beta = 5.91). Inclusion of the precoarctation velocity [V1] component in the Bernoulli equation did not significantly improve the correlation or the agreement. An additional observation was that pullback catheter peak-to-peak gradients were higher than simultaneous peak-to-peak gradients. In the stiff aorta, this difference could be greater than 22 mm Hg (>19%). Acceleration of flow velocity toward the coarctation was evident by pulsed wave Doppler interrogation. Increasing the stiffness of the precoarctation segment also increased the degree of acceleration within this proximal segment: soft, 0.4 to 0.8 m/s; medium, 0.5 to 1. 4 m/s; and stiff, 0.7 to 1.5 m/s. These data suggest that increasing stiffness of the proximal descending aorta can alter the continuous wave detected Doppler gradient and although the gradient itself has increased, it may not predict accurately the true severity of the localized, most severely obstructed segment.

Aorta↗

Evaluation of descending aortic flow volumes and effective orifice area through aortic coarctation by spatiotemporal integration of color Doppler data: An in vitro study.

Flow volumes in an in vitro model of the aorta with 3 different degrees of stiffness (stiff, moderately stiff, and compliant) proximal to a coarctation were calculated by using a digital color Doppler echocardiography flow calculation method that semiautomatically integrates spatial and temporal color flow velocity data. These flow volumes were compared with those obtained by the conventional pulsed Doppler method with reference to ultrasonic flowmeter. Flow volumes determined by the automated method agreed well with those obtained by ultrasonic flowmeter, even in this compliant aorta model with vessel size changing with pulsation, whereas the pulsed Doppler method overestimated the reference data, especially for more compliant descending aortic segments. The combination of flow data with continuous wave Doppler allows definition of effective orifice area for coarctation.

Aorta↗

Flow convergence flow rates from 3-dimensional reconstruction of color Doppler flow maps for computing transvalvular regurgitant flows without geometric assumptions: An in vitro quantitative flow study.

OBJECTIVE: This study was designed to develop and test a 3-dimensional method for direct measurement of flow convergence (FC) region surface area and for quantitating regurgitant flows with an in vitro flow system. BACKGROUND: Quantitative methods for characterizing regurgitant flow events such as flow convergence with 2-dimensional color flow Doppler imaging systems have yielded variable results and may not be accurate enough to characterize those more complex spatial events. METHOD: Four differently shaped regurgitant orifices were studied: 3 flat orifices (circular, rectangular, triangular) and a nonflat one mimicking mitral valve prolapse (all 4 orifice areas = 0.24 cm(2)) in a pulsatile flow model at 8 to 9 different regurgitant flow rates (10 to 50 mL/beat). An ultrasonic flow probe and meter were connected to the flow model to provide reference flow data. Video composite data from the color Doppler flow images of the FC were reconstructed after computer-controlled 180 degrees rotational acquisition was performed. FC surface area (S cm(2)) was calculated directly without any geometric assumptions by measuring parallel sliced flow convergence arc lengths through the FC volume and multiplying each by the slice thickness (2.5 to 3.2 mm) over 5 to 8 slices and then adding them together. Peak regurgitant flow rate (milliliters per second) was calculated as the product of 3-dimensional determined S (cm(2)) multiplied by the aliasing velocity (centimeters per second) used for color Doppler imaging. RESULTS: For all of the 4 shaped orifices, there was an excellent relationship between actual peak flow rates and 3-dimensional FC-calculated flow rates with the direct measurement of the surface area of FC (r = 0.99, mean difference = -7.2 to -0.81 mL/s, % difference = -5% to 0%), whereas a hemielliptic method implemented with 3 axial measurements of the flow convergence zone from 2-dimensional planes underestimated actual flow rate by mean difference = -39.8 to -18.2 mL/s, % difference = -32% to -17% for any given orifice. CONCLUSIONS: Three-dimensional reconstruction of flow based on 2-dimensional color Doppler may add quantitative spatial information, especially for complex flow events. Direct measurement of 3-dimensional flow convergence surface areas may improve accuracy for estimation of the severity of valvular regurgitation.

Blood Flow Velocity↗

Quantitative assessment of chronic aortic regurgitation with 3-dimensional echocardiographic reconstruction: comparison with electromagnetic flowmeter measurements.

Two-dimensional echocardiography and color Doppler are useful in the qualitative assessment of aortic regurgitation. However, color Doppler planar methods are not accurate in quantifying regurgitant flow, in part because of the complex geometry of aortic regurgitant flow events. Three-dimensional echocardiographic reconstruction is a new technique that provides dynamic 3-dimensional images of intracardiac color flow jets. We sought to determine whether the measurement of aortic regurgitant jet volume by 3-dimensional echocardiography correlated with the true regurgitant volume, measured by electromagnetic flowmeter in vivo, to accurately reflect the severity of aortic regurgitation. We performed volume-rendered 3-dimensional echocardiography in 6 sheep with surgically induced chronic eccentric aortic regurgitation. We obtained a total of 22 aortic regurgitation states by altering loading conditions. Instantaneous regurgitant flow rates were obtained by aortic and pulmonary electromagnetic flowmeters. The maximum aortic regurgitant jet volume by 3-dimensional echocardiography and the maximum jet area by 2-dimensional echocardiography were measured and compared with electromagnetic flowmeter data. By electromagnetic flowmeter, aortic regurgitant flow rate varied from 0.14 to 3.1 L/min (mean 1. 25 +/- 0.78); aortic regurgitant stroke volume varied from 1 to 34 mL/beat (mean 12 +/- 8), and regurgitant fraction varied from 3% to 42% (mean 25% +/- 12%). The maximum jet volume by 3-dimensional echocardiography correlated very well with the aortic regurgitant stroke volume (r = 0.92; P <.0001), with the mean regurgitant flow rate (r = 0.87; P <.0001), and with the regurgitant fraction (r = 0. 87; P <.0001) derived from electromagnetic flowmeter. Both intraobserver and interobserver variability on the measurement of the jet volume by 3-dimensional echocardiography were excellent (r = 0.98; P <.0001 and r = 0.90; P <.001, respectively). The maximum jet area by 2-dimensional echocardiography did not correlate with the aortic regurgitant stroke volume (r = 0.41; P = not significant) and related poorly with the regurgitant fraction (r = 0.52; P <.05) by electromagnetic flowmeter. Dynamic 3-dimensional echocardiography can allow better determination of the geometry of the aortic regurgitant jet and may assist of quantifying the severity of aortic regurgitation.