Crohn Disease: mural attenuation and thickness at contrast-enhanced CT Enterography--correlation with endoscopic and histologic findings of inflammation.

PURPOSE: To determine retrospectively if quantitative measures of small-bowel mural attenuation and thickness at computed tomographic (CT) enterography correlate with endoscopic and histologic findings of small-bowel inflammation and to estimate the performance of these measures in predicting inflammatory Crohn disease. MATERIALS AND METHODS: The institutional review board approved this HIPAA-compliant retrospective study, which was conducted with patient informed consent. CT enterography data in 96 patients (31 male patients and 65 female patients) who underwent ileoscopy with or without biopsy were examined for CT signs of active Crohn disease. The most highly enhancing segment of terminal ileum and a normal-appearing ileal loop were identified. After it was confirmed that semiautomated software could accurately measure mural attenuation and thickness, the selected terminal ileal and normal-appearing (control) ileal loops were examined (20 automated measurements at each location) to quantify mural attenuation and wall thickness. Results were compared with endoscopy and histology reports by using logistic regression analysis and receiver operating characteristic curves. RESULTS: Quantitative measures of terminal ileal mural attenuation and wall thickness correlated significantly with active Crohn disease (P < .001). Small-bowel wall thickness was not a significant factor after attenuation was taken into account. A threshold attenuation value with a sensitivity of 90% (18 of 20) for definite Crohn disease (compared with a sensitivity of 80% [16 of 20] for radiologist assessment) was selected. In patients who underwent ileal biopsy, threshold attenuation had a sensitivity identical to that of ileoscopy (81% [26 of 32]; 95% confidence interval: 64%, 93%) in predicting histologic inflammation. CONCLUSION: Quantitative measures of mural attenuation and wall thickness at CT enterography correlate highly with ileoscopic and histologic findings of inflammatory Crohn disease. Quantitative measures of mural attenuation are sensitive markers of small bowel inflammation.

Adolescent↗

Age- and sex-specific differences in the factor of risk for vertebral fracture: a population-based study using QCT.

UNLABELLED: We used QCT scans obtained in 687 men and women, 21-97 years of age, to estimate the factor of risk for vertebral fracture, Phi(vert), defined as the ratio of spinal loading to vertebral strength. With age, vertebral strength declined and Phi(vert) increased significantly more in women than men. Age- and sex-specific differences in Phi(vert) closely resembled previously reported vertebral fracture incidence. INTRODUCTION: Despite the high prevalence of vertebral fractures, little is known about the interaction between spinal loading and vertebral fragility. MATERIALS AND METHODS: We assessed the ratio of spinal loading to vertebral strength (i.e., the factor of risk, Phi(vert)) in an age- and sex-stratified population-based sample of 700 women and men 21-97 years of age. We measured volumetric BMD (vBMD, mg/cm3) and cross-sectional area (CSA, cm2) of the midvertebral bodies of L1-L3 using QCT and computed vertebral compressive strength from these data using engineering beam theory. A biomechanical model of the trunk was used to estimate compressive forces applied to the L3 vertebral body during standing, bending forward, and bending forward while lifting 10 kg. The factor of risk for fracture, Phi(vert), was computed as the ratio of spinal compressive force to vertebral strength for each activity. RESULTS: Men had a higher vertebral strength at all ages, largely because of their greater CSA. Whereas both sexes exhibited a marked decline in vertebral compressive strength with age (p < 0.001), the decline was greater in women than men (-43% versus -31%, p = 0.008). Compressive forces on L3 were greater in men than women, because of their greater body weight and height. For both sexes, forces during bending and lifting were 8-fold higher than those experienced during upright standing. For all activities, Phi(vert) increased with age, but significantly more so in women than men (p < 0.001). For bending and lifting, Phi(vert-bending) exceeded 1.0 in 30% of women and 12% of men > or =50 years of age, values that are similar to the reported frequency of vertebral fracture. CONCLUSION: These findings illustrate potential mechanisms underlying vertebral fractures and provide strong rationale for further evaluation of this QCT-based biomechanical approach for assessment of fracture risk.

Adult↗

First performance evaluation of a dual-source CT (DSCT) system.

We present a performance evaluation of a recently introduced dual-source computed tomography (DSCT) system equipped with two X-ray tubes and two corresponding detectors, mounted onto the rotating gantry with an angular offset of 90 degrees . We introduce the system concept and derive its consequences and potential benefits for electrocardiograph [corrected] (ECG)-controlled cardiac CT and for general radiology applications. We evaluate both temporal and spatial resolution by means of phantom scans. We present first patient scans to illustrate the performance of DSCT for ECG-gated cardiac imaging, and we demonstrate first results using a dual-energy acquisition mode. Using ECG-gated single-segment reconstruction, the DSCT system provides 83 ms temporal resolution independent of the patient's heart rate for coronary CT angiography (CTA) and evaluation of basic functional parameters. With dual-segment reconstruction, the mean temporal resolution is 60 ms (minimum temporal resolution 42 ms) for advanced functional evaluation. The z-flying focal spot technique implemented in the evaluated DSCT system allows 0.4 mm cylinders to be resolved at all heart rates. First clinical experience shows a considerably increased robustness for the imaging of patients with high heart rates. As a potential application of the dual-energy acquisition mode, the automatic separation of bones and iodine-filled vessels is demonstrated.

Coronary Angiography↗

Population-based analysis of the relationship of whole bone strength indices and fall-related loads to age- and sex-specific patterns of hip and wrist fractures.

UNLABELLED: In an age- and sex-stratified population sample (n = 700), we estimated fall-related loads and bone strength indices at the UDR and FN. These load/strength ratios more closely simulated patterns of wrist and hip fractures occurring in the same population than did measurement of vBMD. INTRODUCTION: Areal BMD measurements, although associated with fracture risk, incompletely explain patterns of fragility fractures. Moreover, population-based assessments relating applied loads and whole bone strength to fracture patterns have not been made. MATERIALS AND METHODS: Using QCT, we assessed volumetric BMD (vBMD), cross-sectional geometry, and axial (EA) and flexural (EI) rigidities (indices of bone's resistance to compressive and bending loads, respectively) at the ultradistal radius (UDR) and femoral neck (FN) and estimated the loads applied to the wrist and hip during a fall. We used fall load (FL)/bone strength ratios to estimate fracture risk. RESULTS: vBMD in young adults was similar between sexes. Decreases in vBMD over life were also similar (30% and 28%) at UDR but were somewhat greater (46% and 34%) at FN in women versus men, respectively. In young adults, FL/strength ratios at UDR were 32-51% lower (better) in men than in women and increased (worsened) over life less in men (+4% to +22%) than in women (+20% to +33%). In young adults, FL/strength ratios at FN were only marginally better in men than in women but worsened less over life in men (+22% to +36%) than in women (+40% to +62%). CONCLUSIONS: The 6:1 female preponderance and the virtual immunity of men for age-related increases in wrist fractures are largely explained by the more favorable FL/strength ratios at UDR in young adult men (because of larger bone size and more favorable geometry) versus women and to maintaining this advantage over life. The 2-fold lower incidence of hip fractures in men versus women is largely explained by age-related increases (worsening) of FL/bone strength ratios that are only one-half of the increases in women. The moderate increases in these ratios with aging are insufficient to explain the >4-fold increase in hip fracture incidence after age 75 in both sexes, suggesting contributions of other factors, especially the well-documented increased frequency of injurious falls among the elderly.

Absorptiometry, Photon↗

Three-dimensional CT virtual endoscopy in the detection of simulated tumors in a novel phantom bladder and ureter model.

BACKGROUND AND PURPOSE: Cystoscopy and ureteroscopy have limitations in the evaluation for urothelial tumors, and both are invasive. We studied the utility of three-dimensional (3D) CT virtual endoscopy in phantom models. MATERIALS AND METHODS: A phantom pelvis was constructed of Plexiglas, porcine pelvic bones, and processed animal fat and scanned at various table speeds in a four detector-row CT machine for ability to detect "tumors" of Solidwater plastic polymer. Images were reconstructed at slice thicknesses of 2.5 to 5.0 mm and reconstructed in 3D for evaluation by two radiologists with no knowledge of the scanning parameters or tumor location. Similar studies were performed with a ureter model. RESULTS: With 5-mm slices, the sensitivity for bladder tumors ranged from 67% for 2-mm tumors to 100% for 4-mm tumors, with 12 false-positive findings. The overall sensitivity was 86% with 3.75-mm slices with one false positive, and with 2.5-mm slices, the sensitivity was 93%, again with one false positive. For the ureteral tumors, the overall sensitivities and numbers of false positives were 88.9% and eight with 5.0-mm collimation, 88.9% and four with 3.75-mm collimation, and 100% and three with 2.5-mm collimation. The effective radiation dose for all studies was equivalent to that of a standard abdomen/pelvis scan. CONCLUSIONS: Although virtual endoscopy traditionally has had difficulty detecting tumors <5 mm, the multidetector-row CT protocols used in this study could detect most lesions smaller than this. The scan also depicts the other tissues of the pelvis, which is valuable for staging. The 3D images were produced using data from the CT urogram parameters standard at our institution.

Animals↗

State-of-the-art imaging with 64-channel multidetector CT angiography.

During the past decade, computed tomography angiography (CTA) has become a standard noninvasive imaging method for the depiction of vascular anatomy and pathology. The quality and speed of CTA examinations have increased dramatically as CT technology has evolved from single-channel spiral CT systems (circa 1990) to multichannel (4-, 8-, 10-, and 16-slice) spiral CT systems (circa 1998 through 2001). In 2004, 64-slice multichannel CT technology became available in the United States, with the first installation occurring at the Mayo Clinic College of Medicine. We describe our initial experience using this state-of-the art technology, focusing on examinations that illustrate the increased diagnostic resolution of vascular anatomy and pathology, and the effectiveness of 2-and 3-dimensional postprocessing using a 64-channel CT system. Future applications of 64-channel technology for preoperative and postoperative vascular surgery and endovascular techniques are also described.

Angiography↗

Population-based study of age and sex differences in bone volumetric density, size, geometry, and structure at different skeletal sites.

UNLABELLED: In a population-based, cross-sectional study, we assessed age- and sex-specific changes in bone structure by QCT. Over life, the cross-sectional area of the vertebrae and proximal femur increased by approximately 15% in both sexes, whereas vBMD at these sites decreased by 39-55% and 34-46%, respectively, with greater decreases in women than in men. INTRODUCTION: The changes in bone structure and density with aging that lead to fragility fractures are still unclear. MATERIALS AND METHODS: In an age- and sex-stratified population sample of 373 women and 323 men (age, 20-97 years), we assessed bone geometry and volumetric BMD (vBMD) by QCT at the lumbar spine, femoral neck, distal radius, and distal tibia. RESULTS: In young adulthood, men had 35-42% larger bone areas than women (p < 0.001), consistent with their larger body size. Bone area increased equally over life in both sexes by approximately 15% (p < 0.001) at central sites and by approximately 16% and slightly more in men at peripheral sites. Decreases in trabecular vBMD began before midlife and continued throughout life (p < 0.001), whereas cortical vBMD decreases began in midlife. Average decreases in trabecular vBMD were greater in women (-55%) than in men (-46%, p < 0.001) at central sites, but were similar (-24% and -26%, respectively) at peripheral sites. With aging, cortical area decreased slightly, and the cortex was displaced outwardly by periosteal and endocortical bone remodeling. Cortical vBMD decreased over life more in women ( approximately 25%) than in men (approximately 18%, p < 0.001), consistent with menopausal-induced increases in bone turnover and bone porosity. CONCLUSIONS: Age-related changes in bone are complex. Some are beneficial to bone strength, such as periosteal apposition with outward cortical displacement. Others are deleterious, such as increased subendocortical resorption, increased cortical porosity, and, especially, large decreases in trabecular vBMD that may be the most important cause of increased skeletal fragility in the elderly. Our findings further suggest that the greater age-related decreases in trabecular and cortical vBMD and perhaps also their smaller bone size may explain, in large part, why fragility fractures are more common in elderly women than in elderly men.

Adult↗

The phantom portion of the American College of Radiology (ACR) computed tomography (CT) accreditation program: practical tips, artifact examples, and pitfalls to avoid.

The ACR CT accreditation program, begun in 2002, requires the submission of approximately 20 images, several completed data sheets and printouts of three Excel worksheets. The procedure manual is very detailed, yet participants unfamiliar with the program or having minimal CT experience have needed to redo aspects of their submission, or in some cases do not receive accreditation, due to mistakes made by the physicist. This review of the phantom portion of the ACR CT accreditation program supplements the ACR provided instructions with additional photos of phantom setup, region-of-interest (ROI), and image placement on the film sheets, and examples of completed portions of actual (but anonymous) submissions. Common mistakes, as well as uncommon but interesting images, are shown and explanations are given as to what could have been done to avoid the problem. Additionally, a review of CT dose measurement techniques and calculations will enable the physicist to better assist sites where typical exam doses are above the ACR reference values.

Accreditation↗

CT urography.

With the recent introduction of multi-detector row helical computed tomography (CT), the radiologic evaluation of patients with urologic disease has changed rapidly. Two major approaches to CT urography have been developed. The first approach combines axial CT with timed excretory urography (EU) performed by using conventional radiography, digital radiography, or CT scanned projection radiography (SPR). This approach produces traditional projection urograms, and the timed imaging technique is familiar to radiologists and clinicians. Additional excretory phase CT can be performed when the EU findings are positive or indeterminate. Improved CT SPR processing technology produces radiographlike images, thus eliminating patient transportation between the CT and urography suites or the necessity for a CT suite with a ceiling-mounted x-ray tube and a modified CT tabletop for performance of EU. The second approach to CT urography combines axial CT with thin-section excretory phase CT. The near-isotropic volume data set enables creation of high-resolution two- and three-dimensional reformatted images. However, the increased amount of radiation and the time required for data manipulation are concerns. Further studies evaluating large numbers of patients with various urothelial abnormalities will be necessary to determine the optimal CT urography technique for clinical practice.

Equipment Design↗

Patient dose in cardiac computed tomography.

BACKGROUND: Computed tomographic (CT) imaging of the heart is becoming more widely available, and the quality of and interest in these examinations continues to increase. Yet, many physicians who perform or order these exams may be unfamiliar with the radiation doses that are delivered to the patient during specific cardiac CT examinations. OBJECTIVE: The purpose of this article is to describe several fundamental radiation dose quantities that are used to either measure or report the dose associated with a particular CT examination. In addition, radiation dose estimates are given for coronary artery calcium imaging and coronary angiography performed using either electron beam CT (EBCT) or multi-detector row CT (MDCT) systems.

Coronary Angiography↗

Physics and dosimetry in computed tomography.

CT data acquisition and image reconstruction techniques are closely related to image quality and patient radiation dose. There is little question that technological developments currently underway will change the nature of both, and result in improved quality and diagnostic value of cardiovascular CT images while hopefully minimizing radiation dose to the patient.

Algorithms↗

Quality assurance for computed-tomography simulators and the computed-tomography-simulation process: report of the AAPM Radiation Therapy Committee Task Group No. 66.

This document presents recommendations of the American Association of Physicists in Medicine (AAPM) for quality assurance of computed-tomography- (CT) simulators and CT-simulation process. This report was prepared by Task Group No. 66 of the AAPM Radiation Therapy Committee. It was approved by the Radiation Therapy Committee and by the AAPM Science Council.

Computer Simulation↗

Computer-aided detection of polyps in a colon phantom: effect of scan orientation, polyp size, collimation, and dose.

PURPOSE: To determine the importance of polyp size, orientation to the scan plane, collimation, scanner type (single or multislice helical), and radiation dose on computed tomography (CT) colonography computer-aided detection. MATERIALS AND METHODS: Eight tissue-equivalent simulated polyps were placed into the interior of an air-filled acrylic tube placed within a water-filled box. Their sizes, expressed by diameter and height in millimeters, were 10 x 10, 10 x 7, 10 x 5, 10 x 3, 7 x 7, 7 x 5, 7 x 3, and 5 x 5. Detection of the polyps was performed by applying our prototype automated polyp detector software to 48 CT colonography data sets of the phantom acquired with different CT scanner settings. RESULTS: We detected at least six of the eight polyps in 47 of 48 experiments. The two most frequently undetected polyps (7 x 7 and 5 x 5) had extreme eccentricity (their height was twice the radius of the base) and were most commonly missed for 90 degrees tube orientation, 5-mm collimation, and high table speed. False-positive detections occurred in only 5 of 48 experiments. CONCLUSION: Clinically significant 10-mm polyps can be detected with 100% sensitivity in all orientations, doses, collimations, and modes that we examined.

Artifacts↗

CT dose reduction and dose management tools: overview of available options.

In the past decade, the tremendous advances in computed tomography (CT) technology and applications have increased the clinical utilization of CT, creating concerns about individual and population doses of ionizing radiation. Scanner manufacturers have subsequently implemented several options to appropriately manage or reduce the radiation dose from CT. Modulation of the x-ray tube current during scanning is one effective method of managing the dose. However, the distinctions between the various tube current modulation products are not clear from the product names or descriptions. Depending on the scanner model, the tube current may be modulated according to patient attenuation or a sinusoidal-type function. The modulation may be fully preprogrammed, implemented in near-real time by using a feedback mechanism, or achieved with both preprogramming and a feedback loop. The dose modulation may occur angularly around the patient, along the long axis of the patient, or both. Finally, the system may allow use of one of several algorithms to automatically adjust the current to achieve the desired image quality. Modulation both angularly around the patient and along the z-axis is optimal, but the tube current must be appropriately adapted to patient size for diagnostic image quality to be achieved.

Algorithms↗