At least 109 records · Page 6Linked to original sources

Pharmacy access to syringes among injecting drug users: follow-up findings from Hartford, Connecticut.

OBJECTIVE: To break the link between drug use and the human immunodeficiency virus (HIV), in 1992 the state of Connecticut rescinded a 14-year ban on pharmacy sales of syringes without a physician's prescription. In 1993, the Center for Disease Control and Prevention (CDC) evaluated the impact of the new legislation on access to syringes among injecting drug users (IDUs) and found an initial pattern of expanded access. However, it also found that some pharmacies, after negative experiences with IDU customers, reverted to requiring a prescription. This chapter reports findings from a four-year follow-up study of current IDU access to over-the-counter (OTC) pharmacy syringes in Hartford, Connecticut. METHODS: Through structured interviews, brief telephone interviews, and mailed surveys, data on nonprescription syringe sale practices were collected on 27 pharmacies, including 18 of the 21 pharmacies in Hartford and none from pharmacies in contiguous towns, during June and July 1997. Interview data on pharmacy syringe purchase from two sample of IDUs, a group of out-of-treatment injectors recruited through street outreach, and a sample of users of the Hartford Needle Exchange Program, also are reported. RESULTS: The study found that, while market trends as well as negative experiences have further limited pharmacy availability of nonprescription syringes, pharmacies remain an important source of sterile syringes for IDUs. However, the distribution of access in not even; in some areas of the city it is much easier to purchase nonprescription syringes than in other. All of the seven pharmacies located on the north end of Hartford reported that they had a policy of selling OTC syringes, whereas only six (54.5%) of the II pharmacies located on the south end have such a policy. Overt racial discrimination was not found to be a barrier to OTC access to syringes. CONCLUSIONS: To further decrease acquired immunodeficiency syndrome (AIDS) risk among IDUs, there is a need for public education to counter empirically unsupported stereotypes about IDUs that diminish their access to health care and AIDS prevention resources and services. In states or cities where pharmacy sale of nonprescription syringes is illegal, policy makers should examine the benefits of removing existing barriers to sterile syringe acquisition. In cases in which pharmacy sale of nonprescription syringes is legal, local health departments should implement educational programs to inform pharmacy staff and management about the critically important role low-cost (or cost-free), sterile syringe access can play in HIV prevention.

Centers for Disease Control and Prevention, U.S.↗

Impact of the change in Connecticut syringe prescription laws on pharmacy sales and pharmacy managers' practices.

We assessed the impact of the 1992 change in Connecticut syringe prescription laws on pharmacy sales and pharmacy managers' sales practices. A mail survey was conducted in 1994 of all current pharmacy managers in the five largest cities in Connecticut (Hartford, New Haven, Waterbury, Bridgeport, and Stamford) and a random sample of those practicing in all other areas. Of these, 89.3% of the pharmacies in the five largest cities and 85.1% in the other areas had ever sold syringes without a prescription since the July 1992 law went into effect. Most pharmacists identified safety issues as very important in their personal decision about the sale of syringes without a prescription. Although the purpose of the change in the prescription law was to provide expanded access to sterile syringes by injection drug users (IDUs), only 31.4% of the managers who were allowed to sell in all instances and 18.1% of those who sold at their discretion were very willing to sell syringes to IDUs. In the logistic regression model of pharmacies with a sell-in-all-instances policy, the perceived benefit of the sale of syringes on health and community well-being was the only influence independently associated with managers support for nonprescription sales. Overall, managers reported they did not know what other pharmacists thought (40.4%) or did (42.9%) regarding the sale of syringes. When pharmacists had discretion over syringe sales, managers' beliefs about what other Connecticut pharmacists thought and did about the nonprescription sale of syringes remained a significant influence on the degree of support for sales. Most pharmacies implemented and maintained policies permitting the sale of syringes without a prescription. Several issues, including risk of discarded contaminated syringes around pharmacies and in the community and reluctance to sell to IDUs, reduced pharmacists willingness to sell syringes. Efforts to incorporate pharmacists as active partners in HIV prevention in IDUs should promote the sale of syringes without a prescription to IDUs as acceptable public health practice.

Acquired Immunodeficiency Syndrome↗

A randomised controlled trial of the reciprocating syringe in arthrocentesis.

OBJECTIVE: To evaluate the outcomes of arthrocentesis with the new highly controllable, one handed reciprocating procedure syringe compared with a conventional syringe. METHODS: 100 arthrocentesis procedures were randomised between the reciprocating syringe and the conventional syringe. Outcome measures included patient pain, procedure duration, operator satisfaction, synovial fluid volume, cell counts, and complications. RESULTS: 50 arthrocentesis procedures with the conventional syringe resulted in a mean (SD) procedure time of 3.39 (1.88) minutes, a mean VAPS (patient pain) score of 5.35 (3.15), and a mean VASS (operator satisfaction) score of 4.88 (1.92); 30 of the 50 subjects experienced moderate to severe pain (VAPS score 5 or greater) during arthrocentesis. In contrast, the reciprocating syringe resulted in a reduced procedure time of 1.94 (1.14) minutes (p<0.001), a reduced VAPS (patient pain) score of 2.54 (1.60) (p<0.001), and an increased VASS (operator satisfaction) score of 8.91 (0.79) (p<0.001). Only five of the 50 of subjects experienced moderate to severe pain with the reciprocating syringe. Synovial cell counts were similar between the two syringes (p>0.05), but there was a trend toward greater volume (greater synovial fluid yield) and fewer red blood cells with the reciprocating syringe. CONCLUSIONS: Arthrocentesis with a conventional syringe results in moderate to severe pain in 60% of subjects. The reciprocating syringe prevents significant pain, reduces procedure time, and improves physician performance of arthrocentesis. The reciprocating syringe is superior to the conventional syringe in arthrocentesis.

Adult↗

Injection drug users report good access to pharmacy sale of syringes.

OBJECTIVE: To examine injection drug users (IDUs) opinions and behavior regarding purchase of sterile syringes from pharmacies. DESIGN: Focus groups. SETTING: Urban and rural sites in Colorado, Connecticut, Kentucky, and Missouri. PATIENTS OR OTHER PARTICIPANTS: Eight focus groups, with 4 to 15 IDU participants per group. INTERVENTIONS: Transcripts of focus group discussions were evaluated for common themes by the authors and through the use of the software program NUD*IST. MAIN OUTCOME MEASURES: Knowledge of human immunodeficiency virus (HIV), pharmacy use, barriers to access from pharmacies, high-risk and risk-reducing behavior, and rural/urban difference. RESULTS: Almost all participants knew the importance of using sterile syringes for disease prevention and reported buying syringes from pharmacies more than from any other source. Two IDUs believed pharmacists knew the syringes were being used for injecting drugs and perceived pharmacists' sales of syringes to be an attempt to contribute to HIV prevention. Most IDUs reported that sterile syringes were relativity easy to buy from pharmacies, but most also reported barriers to access, such as having to buy in packs of 50 or 100, being made to sign a book, having to make up a story about being diabetic, or having the feeling that the pharmacists were demeaning them. While the majority of IDUs reported properly cleaning or not sharing syringes and safely disposing of them, others reported inadequate cleaning of syringes and instances of sharing syringes or of improper disposal. There were few differences in IDUs' reported ability to buy syringes among states or between urban and rural sites, although the data suggest that IDUs could buy syringes more easily in the urban settings. CONCLUSION: For the most part, participants understood the need for sterile syringes in order to protect themselves from HIV, hepatitis B virus, and hepatitis C virus and saw pharmacies as the best source of sterile syringes. Although these data are not generalizable, they suggest that pharmacists can and do serve as HIV-prevention service providers in their communities.

Attitude of Health Personnel↗

Community syringe collection and disposal policies in 16 states.

OBJECTIVE: To review laws, regulations, and guidelines that affect the collection and disposal of hypodermic needles, syringes, and lancets used outside of professional health care settings (hereafter referred to as "community syringes"). DESIGN: Law and policy analysis. SETTING: Alabama, California, Florida, Georgia, Hawaii, Massachusetts, Michigan, Minnesota, New Jersey, New York, Ohio, Oregon, Rhode Island, South Carolina, Washington, and Wisconsin. INTERVENTION: Information on syringe collection and disposal in the community was gathered from federal and state records and state agency personnel. MAIN OUTCOME MEASURES: Legally permissible means of syringe collection and disposal available to persons in the community injecting medical treatments and injection drug users. RESULTS: Laws, regulations, or guidelines in 13 states allowed community syringes to be legally discarded in household trash; guidelines for in-trash disposal varied among the states. Only 6 states had laws or regulations that specifically addressed community syringe collection. In 10 states, infectious waste laws and regulations that apply to medical facilities such as clinics would also apply to community syringe collection sites. CONCLUSION: In the 16 states studied, laws, regulations, and guidelines relating to community syringe collection and disposal were somewhat inconsistent and confusing and presented potential barriers to safe disposal. States should consider amending laws, regulations, and guidelines to promote community syringe collection programs. A national effort is needed to achieve consistent community syringe collection and disposal laws and guidelines for all states. Pharmacists can aid in safe syringe disposal by counseling their patients about safe disposal, providing or selling sharps containers, and accepting used syringes for safe disposal. Pharmacists can join other interested groups in advocating clarification of disposal laws and regulations that favor community programs designed to keep syringes out of the trash so that they can be disposed of as infectious waste.

Blood-Borne Pathogens↗

Introducing auto-disable syringes to the national immunization programme in Madagascar.

OBJECTIVE: To evaluate the safety and coverage benefits of auto-disable (AD) syringes, weighed against the financial and logis- tical costs, and to create appropriate health policies in Madagascar. METHODS: Fifteen clinics in Madagascar, trained to use AD syringes, were randomized to implement an AD syringe only, mixed (AD syringes used only on non-routine immunization days), or sterilizable syringe only (control) programme. During a five-week period, data on administered vaccinations were collected, interviews were conducted, and observations were recorded. FINDINGS: The use of AD syringes improved coverage rates by significantly increasing the percentage of vaccines administered on non-routine immunization days (AD-only 4.3%, mixed 5.7%, control 1.1% (P<0.05)). AD-only clinics eliminated sterilization sessions for vaccinations, whereas mixed clinics reduced the number of sterilization sessions by 64%. AD syringes were five times more expensive than sterilizable syringes, which increased AD-only and mixed clinics' projected annual injection costs by 365% and 22%, respectively. However, introducing AD syringes for all vaccinations would only increase the national immunization budget by 2%. CONCLUSION: The use of AD syringes improved vaccination coverage rates by providing ready-to-use sterile syringes on non-routine immunization days and decreasing the number of sterilization sessions, thereby improving injection safety. The mixed programme was the most beneficial approach to phasing in AD syringes and diminishing logistical complications, and it had minimal costs. AD syringes, although more expensive, can feasibly be introduced into a developing country's immunization programme to improve vaccination safety and coverage.

Child, Preschool↗

The effect of syringe exchange use on high-risk injection drug users: a cohort study.

OBJECTIVE: To determine whether syringe exchange program use is associated with cessation of syringe sharing among high-risk injection drug users. DESIGN AND METHODS: Between 1992 and 1996, street-recruited injection drug users were interviewed and received HIV testing and counseling semi-annually, as part of a dynamic cohort study. We examined a cohort of 340 high-risk injection drug users for whom two observations, 6-months apart, were available and who reported syringe sharing at the first interview. Multivariate logistic regression analysis was performed to determine the relationship between syringe exchange program use and cessation of syringe sharing, while controlling for confounding factors. RESULTS: At follow-up interview, 60% (204 of 340) reported quitting syringe sharing. High-risk injection drug users who began using the syringe exchange program were more likely to quit sharing syringes [adjusted odds ratio (AOR), 2.68; 95% confidence interval (CI), 1.35-5.33], as were those who continued using the syringe exchange program (AOR,1.98; 95% CI, 1.05-3.75) in comparison with non-syringe exchange program users, while controlling for confounding factors. CONCLUSIONS: The initiation and continuation of syringe exchange program use among high-risk injection drug users is independently associated with cessation of syringe sharing. Syringe exchange program use can be an important component in reducing the spread of blood-borne infectious diseases among high-risk injection drug users.

Adult↗

Syringe laws and pharmacy regulations are structural constraints on HIV prevention in the US.

OBJECTIVE: To review the legal and regulatory barriers that restrict pharmacy sales of syringes to injection drug users (IDUs) and to discuss how reducing these barriers can facilitate access to sterile syringes for IDUs and improve HIV prevention. BACKGROUND: IDUs' access to sterile syringes from community pharmacies in the United States is limited by state laws and regulations governing syringe sales. Restricted availability of sterile syringes from pharmacies is a structural barrier that greatly impedes HIV prevention for IDUs, who often share and reuse syringes because they cannot obtain and possess sterile syringes. These high-risk behaviors contribute to the transmission of HIV and other blood-borne pathogens among IDUs, their sexual partners, and their children. STATE EXPERIENCES: In Connecticut, because of high HIV prevalence among IDUs, restrictive syringe laws were changed. After the legal changes in Connecticut, both pharmacy sales of syringes in areas of high drug use and purchases of syringes in pharmacies (reported by IDUs) increased, while syringe sharing (reported by IDUs) decreased. Maine and Minnesota have made similar changes in laws. CONCLUSIONS: Increasing access to sterile syringes through pharmacies requires the repeal or modification of legal barriers. Pharmacy sale of syringes to IDUs is an inexpensive HIV prevention intervention with the potential to substantially reduce HIV transmission. Further studies are needed to document how changes to legal barriers can influence HIV prevention for IDUs.

Connecticut↗

Community-based programs for safe disposal of used needles and syringes.

OBJECTIVES: To review issues related to discarded syringes in the community and to describe community-based programs for the safe disposal of used needles and syringes. METHODS: We used the medical literature and chain referral to identify community-based syringe disposal programs other than syringe exchange programs (SEPs). We held a workshop in June 1996 involving staff from disposal programs; manufacturers of syringes, sharps containers, and other disposal devices; solid waste companies; public health staff; and researchers. RESULTS: Fifteen programs for the safe disposal of syringes were identified in the United States, Canada, and Australia. Of these, 12 primarily served persons with diabetes who use insulin, and 3 primarily served injection drug users (IDUs). The programs used three major strategies: puncture-resistant containers discarded in trash, community drop boxes, and sharps containers turned in for biohazard disposal at community sites, hospitals, or pharmacies. Participants in the workshop described key points in developing syringe disposal programs. Programs should involve pharmacists, physicians, waste disposal companies, public health departments, hospitals, diabetes educators, persons with diabetes who use insulin, and IDUs. For IDUs, criminal penalties for possession of syringes are a substantial deterrent to participation in community efforts to safely dispose of used syringes. The multiple and sometimes conflicting local, state, and federal laws and regulations concerning medical waste hinder development of multistate or national approaches to the safe disposal of syringes. More information is needed on community-based syringe disposal programs. CONCLUSION: Communities in the United States, Canada, and Australia have developed different approaches to achieve safe disposal of used syringes.

Australia↗

Preparation of 99mTc-MAG3: radiochemical purity is affected by the residence time of sodium chloride injection in a three-part syringe.

BACKGROUND: Routine technetium-99m mercaptoacetyltriglycine (99mTc-MAG3) radiochemical purity measurements have revealed occasional unacceptably low values. Preliminary investigations suggested a causal link with the residence time of sodium chloride injection in the syringe used to reconstitute the MAG3 kit. OBJECTIVES: To investigate the cause of this phenomenon, determine how it can be avoided and establish whether it occurs with other 99mTc radiopharmaceuticals. METHODS: 99mTc-MAG3 was prepared by drawing sodium chloride injection into a lubricated, three-part, 10 ml Plastipak syringe and using it to reconstitute a MAG3 kit immediately or after a 15 min incubation period. The radiochemical purity was measured by high-performance liquid chromatography. The experiment was repeated using lubricant-free, two-part, Norm-Ject syringes and lubricated, two-part, Monoject syringes (15 min incubation only). To investigate the influence of Plastipak's rubber components on the radiochemical purity, samples were prepared using sodium chloride injection that had been incubated with lubricated or lubricant-free rubber plunger ends. Similar experiments were performed to determine the effect of Plastipak on 99mTc-exametazime, 99mTc-sestamibi and 99mTc-tetrofosmin. RESULTS: The radiochemical purities of 99mTc-MAG3 prepared with sodium chloride injection incubated for 0 and 15 min in Plastipak syringes were 96.4+/-0.5% and 89.4+/-5.5%, respectively. The difference was significant (P<0.05, n=10). With Norm-Ject syringes, the radiochemical purities were 96.5+/-0.5% and 96.6+/-0.5%, respectively. The difference was not significant (P>0.05, n=10). With Monoject syringes, the radiochemical purity was 96.6+/-0.4% (n=10). 99mTc-MAG3 prepared using sodium chloride injection treated with lubricated and unlubricated syringe rubber plunger ends had radiochemical purities of 85.3+/-6.6% and 82.1+/-6.5% (n=5), respectively. The radiochemical purities of other 99mTc radiopharmaceuticals prepared using sodium chloride injection incubated for 0 or 15 min in Plastipak syringes were as follows: 99mTc-exametazime, 95.3+/-0.6% and 94.5+/-1.8%; 99mTc-sestamibi, 98.0+/-0.6% and 97.7+/-0.6%; 99mTc-tetrofosmin, 96.5+/-0.2% and 97.0+/-0.4%. None of the differences was significant (P>0.05, n=5). CONCLUSIONS: A lipophilic impurity, originating from the rubber plunger of a three-part Plastipak syringe, is formed in 99mTc-MAG3 when the sodium chloride injection used to reconstitute the kit is in the syringe for a prolonged time. The effect is eliminated by using a two-part syringe or by injecting the sodium chloride injection into the kit immediately. The phenomenon does not occur with 99mTc-exametazime, 99mTc-sestamibi or 99mTc-tetrofosmin.

Drug Contamination↗

Brand and size matter when choosing a syringe to relieve pressure in a tracheal tube cuff.

We studied the use of an inline syringe as a pressure relief valve for tracheal tube cuffs during exposure to nitrous oxide to see if the technique works. Bench testing was done to determine the stick and slip characteristics of syringes of different brands and sizes. Cuffs were inflated with 20 mL of air, producing a cuff pressure of 100-120 mm Hg. Then the plunger of the syringe was allowed to passively rebound to a steady pressure at which the plunger stopped ("stick pressure"). After several minutes, pressure in the syringe was forcibly increased with a second syringe until the plunger started moving again ("slip pressure"). Stick pressure varied from 18 to 82 mm Hg depending on the brand and size of syringe used. Slip pressures exceeded stick pressures by 20-120 mm Hg. Cuff pressure increased in a linear fashion during nitrous oxide exposure, and no syringe demonstrated automatic pressure reduction. We conclude that a syringe attached to the pilot balloon connector can be used to control tracheal tube cuff pressure during nitrous oxide anesthesia. However, not all syringes are suitable for this purpose: large syringes are better than small syringes, and the Terumo brand is more suitable than BD or Monoject. The system does not work automatically, and intermittent compression of the syringe plunger to overcome static friction is required to avoid overdistension.

Elasticity↗

Evaluation of the stability of (99m)Tc-ECD and stabilized (99m)Tc-HMPAO stored in syringes.

OBJECTIVE: To determine and compare the stability of (99m)Tc-ECD and stabilized (99m)Tc-HMPAO when stored in syringes over an 8-h period. METHODS: (99m)Tc-ECD and stabilized (99m)Tc-HMPAO were prepared according to the manufacturers' protocols, with the following exception: eluate less than 60 min old was used to prepare (99m)Tc-HMPAO rather than the recommended 30 min. Once prepared, 185 MBq (5 mCi) of both products were drawn into 5-mL syringes and allowed to sit at room temperature. At 2, 4, 6, and 8 h after preparation, the radiochemical purity (RCP) of the contents of the syringes was determined and compared to the RCP of the products in vials. Retention of activity of each product in syringes was also evaluated by measuring activity remaining in each syringe (and filter, in the case of (99m)Tc-HMPAO) after expressing its contents. RESULTS: The RCP of stabilized (99m)Tc-HMPAO stored in syringes decreased from a mean of 87.7% at 2 h to 74.0% at 8 h after preparation. In contrast, (99m)Tc-ECD retained an RCP of greater than 94% throughout the time tested. The impurity that appeared to increase over time with (99m)Tc-HMPAO was found to be sodium pertechnetate. Total retention of activity remaining in the syringe and filter ranged from 11.6% at 2 h to 9.5% at 8 h for (99m)Tc-HMPAO; the syringe itself retained less than 5% of the total activity at all time periods. (99m)Tc-ECD exhibited 6.2% to 11.3% retention of activity in the syringe. The sorption of sodium pertechnetate to the syringe for the same time period was less than 1%. CONCLUSIONS: (99m)Tc-ECD is a more stable product than stabilized (99m)Tc-HMPAO in a syringe. Both products demonstrate retention of radioactivity in the syringe. Some of this retention may denote sorption of the products to plastic.

Cysteine↗

New York City pharmacists' attitudes toward sale of needles/syringes to injection drug users before implementation of law expanding syringe access.

In May 2000, New York State passed legislation permitting the sale, purchase, and possession of up to 10 needles and syringes (hereafter "syringes") without a prescription, intended to reduce blood-borne pathogen transmission among injection drug users (IDUs). To obtain baseline data on pharmacists' attitudes and practices related to human immunodeficiency virus (HIV) prevention and IDUs, a telephone survey was administered to 130 pharmacists systematically selected in New York City. Less than half of pharmacists were aware of the new law; 49.6% were willing to or supported providing nonprescription sales of syringes to IDUs. Pharmacists in support tended to be less likely to consider customer appearance "very important." Managing and supervising pharmacists were more likely than staff pharmacists to support syringe sales to IDUs. Managing and supervising pharmacists who stocked packs of 10 syringes and personal sharps disposal containers, pharmacists who supported syringe exchange in the pharmacy, and pharmacists who were willing to sell syringes to diabetics without a prescription were more likely to support syringe sales to IDUs. Syringe disposal was a prominent concern among all pharmacists. Those not in support of syringe sales to IDUs tended to be more likely to believe the practice would increase drug use. These data suggest the need for initiatives to address concerns about syringe disposal and tailored continuing education classes for pharmacists on HIV and viral hepatitis prevention among IDUs.

Attitude of Health Personnel↗

Stability of dopamine hydrochloride 0.5 mg/mL in polypropylene syringes.

OBJECTIVE: To determine the stability of dopamine hydrochloride diluted to 0.5 mg/mL in isotonic glucose infusion and stored in 50 mL polypropylene syringes from two different suppliers. METHODS: Polypropylene syringes containing dopamine hydrochloride 0.5 mg/mL were stored protected from light at 25 degrees C/60% relative humidity for 1 week, at 4 degrees C/ambient humidity for 3 months and at -20 degrees C/ambient humidity for 6 months. Samples were periodically withdrawn and the concentration of dopamine hydrochloride determined using a stability-indicating high-performance liquid chromatography method. The appearance of the solutions with respect to colour, clarity and precipitation was studied and the pH of the solutions measured. RESULTS: The concentration of dopamine hydrochloride in the solutions remained at least 95% of the initial concentration after storage at 25 degrees C for 1 week and at 4 degrees C for 3 months. Variation in dopamine concentration was observed in syringe samples, stored frozen, from two suppliers. The mean concentration was >/=95% for product stored in syringes from Codan and >/=90% of initial concentration when stored in syringes from Braun. In some of the syringes from Braun, the concentration was lower than 90%. The samples in all syringes had to be homogenized after thawing to give a uniform solution. There were no other evident changes in the physical appearance or pH of the solutions during the study. CONCLUSION: Dopamine hydrochloride 0.5 mg/mL in isotonic glucose solution is stable when protected from light for 1 week at 25 degrees C/60% relative humidity and for 3 months at 4 degrees C/ambient humidity in Codan and Braun syringes. The cause of content variability observed in solutions stored frozen in Braun syringes needs further investigation. Braun syringes should not be used if the product is to be frozen. The stability of the frozen product, in Codan syringes is 6 months.

Cardiotonic Agents↗

Over a decade of syringe exchange: results from 1997 UK survey.

AIMS: To describe syringe exchange provision in the United Kingdom. DESIGN: Two-phase cross-sectional survey: phase I, establishing a sampling frame of syringe exchange coordinators (n=420); phase II, surveying the coordinators seeking data on the number of syringe exchange outlets, visits and syringes distributed during April 1997 (68% response rate). SETTING: United Kingdom. FINDINGS: In 1997, nearly all Health Authorities in the United Kingdom (96%) operated some form of syringe exchange service, except Northern Ireland. In April 1997, 1 707 000 syringes were reported as being distributed. Assuming that non-responders coordinated the median number of outlets and distributed the median number of syringes as responders, we estimate that 27 million syringes were distributed annually from over 2000 outlets in the United Kingdom. The number distributed in Scotland was 3-4 times less than in England when measured as a number per adult (15-44), drug user in treatment, or estimated injecting drug user. CONCLUSIONS: Overall, there has been a 6.5-fold increase in syringe distribution in England since 1991. The number of syringes distributed in the United Kingdom may be higher than the United States. However, there appears to be unequal distribution of syringes within the United Kingdom, which may be associated with higher levels of HCV among injectors in Scotland compared to England.

Adolescent↗

Fluid delivery from infusion-pump syringes.

Fluid-delivery rates of five small-volume infusion-pump syringes were compared. The study consisted of a comparison of the infusion-pump syringes in their respective infusion pumps (1) set for continuous delivery at 1 mL/hr, (2) set for continuous delivery at 3 mL/hr, and (3) set to deliver 1-mL bolus volumes during continuous delivery at 4 mL/hr. The Life-care prefilled 30-mL syringe (Abbott), the DBL 30-mL syringe no. 770205 (DBL Inc.), and the Pump-Jet 30-mL syringe no. 1931 (International Medication Systems) were tested in the Lifecare PCA Plus II infusion pump no. 4100 (Abbott). The 30-mL Pump-Jet syringe no. 1911 (International Medication Systems) and the DBL 30-mL syringe no. 709700 (DBL Inc.) were tested in the Stratofuse PCA infusion pump (Baxter). The infusion pumps were set to deliver fluid continuously at 1 mL/hr for 30 hours, and the solutions were collected separately and weighed. The procedure was repeated with the infusion rate set at 3 mL/hr for 10 hours. For the third part of the study, each syringe was tested to deliver 1-mL boluses with 0, 5, 15, and 25 mL removed from the syringe. The solutions were collected and weighed before and after each bolus was delivered. The volume of solution collected was calculated by using the specific gravity of the solution. The syringes delivered significantly different volumes during the first hour of infusion at both the 1- and 3-mL/hr rates. Differences also existed across time for most of the syringes. Bolus volumes varied greatly after infusion of 0 or 5 mL of fluid but were acceptable for the remainder of the infusions.(ABSTRACT TRUNCATED AT 250 WORDS)

Humans↗

Fluctuations in syringe-pump infusions: association with blood pressure variations in infants.

Flow continuity of two brands of syringe pumps and four brands of syringes was studied as a possible cause of hemodynamic fluctuations observed in neonates. Cyclical fluctuations were observed in the blood pressure of 14 neonates receiving dopamine infusions by syringe pump at flow rates from 0.2 to 1 mL/hr. Atom 235 and IVAC 770 pumps and various sizes of Terumo, Becton Dickinson, Omnifix, and IVAC syringes were evaluated. Flow continuity was assessed by using a gravimetric technique. The force needed to initiate and maintain syringe plunger motion was also measured. Noncontinuous flow was encountered most commonly with Terumo syringes, which delivered boluses at regular intervals at flow rates up to 5 mL/hr. The interval was dependent on flow rate and was similar to the time between the blood pressure fluctuations observed clinically. The syringe plunger force exhibited regular fluctuations indicative of the plunger sticking, and simultaneous measurement of flow established a direct temporal relationship with boluses. The other syringes tested did not exhibit such fluctuations. No differences were found between the two syringe pumps. Syringe plunger sticking, resulting in intermittent boluses and potential blood pressure fluctuations, may occur at low flow rates and with certain syringe brands. This appeared to be the cause of hemodynamic fluctuations in neonates receiving dopamine infusions.

Blood Pressure↗

Effects of sample storage time, temperature and syringe type on blood gas tensions in samples with high oxygen partial pressures.

BACKGROUND: Although plastic arterial sampling syringes are now commonly used, the effects of sample storage time and temperature on blood gas tensions are poorly described for samples with a high oxygen partial pressure (PaO2) taken with these high density polypropylene syringes. METHODS: Two ml samples of tonometered whole blood (PaO2 86.7 kPa, PaCO2 4.27 kPa) were placed in glass syringes and in three brands of plastic blood gas syringes. The syringes were placed either at room temperature or in iced water and blood gas analysis was performed at baseline and after 5, 10, 20, 40, 60, 90, and 120 minutes. RESULTS: In the first 10 minutes measured PaO2 in plastic syringes at room temperature fell by an average of 1.21 kPa/min; placing the sample on ice reduced the rate of PaO2 decline to 0.19 kPa/min. The rate of fall of PaO2 in glass at room temperature was 0.49 kPa/min. The changes in PaCO2 were less dramatic and at room temperature averaged increases of 0.47 kPa for plastic syringes and 0.71 kPa for glass syringes over the entire two hour period. These changes in gas tension for plastic syringes would lead to an overestimation of pulmonary shunt measured by the 100% oxygen technique of 0.6% for each minute left at room temperature before analysis. CONCLUSIONS: Glass syringes are superior to plastic syringes in preserving samples with a high PaO2, and prompt and adequate cooling of such samples is essential for accurate blood gas analysis.

Blood Gas Analysis↗