Accuracy comparison of a 16 and 64 multidetector-row computed tomography scanner to image small high-density structures.
OBJECTIVES: The accuracy in imaging small high-density structures is compared for 16 and 64 multidetector-row computed tomography (MDCT) scanners. MATERIALS AND METHODS: Phantom experiments and different quantification methods are used to establish size measurement accuracy, object signal, and image noise, for both MDCT systems. RESULTS: At similar scanning doses, image noise is larger ( approximately 55%) for the 64 MDCT compared with the 16 MDCT, leading to lower signal-to-noise ratios ( approximately 28% for objects <2 mm). Object spread in the xy-plane is similar for both systems; while it is reduced along the z-axis (by approximately 0.18 mm) for the 64 MDCT. Measurement accuracy of the 64 MDCT is not significantly higher (P < 0.05) compared with the 16 MDCT when using a relative-threshold corresponding to 50% of the object maximum attenuation value. However, when using a fixed-threshold, interscanner and interprotocol measurement differences are statistically significant (eg, volume relative errors are reduced by approximately 17% on average for the 64 MDCT). CONCLUSIONS: Measurement accuracy of the 16 and 64 MDCT scanners is not significantly different when using a 50% relative threshold. However, image noise is significantly larger for the 64 MDCT. Compared with a fixed-threshold based method, the 50% relative-threshold strongly reduces interscanner and interprotocol measurement dependency and improves accuracy.