Search PubMed⌕ Search

SEARCH · Search PubMed

Results for “ATR”

Search indexed PubMed citations on genomics, clinical trials, systematic reviews and public health. Explore titles, authors and supplied subject terms, then open the PubMed record.

Quote a phrase for an exact phrase match. Source license links do not imply unrestricted reuse.

At least 415 records · Page 23Linked to original sources

Study on a new index of fetal respiratory function: a novel optical analysis of the amniotic fluid by the vacuum-dried infrared attenuated total reflection (ATR) method.

For the purpose of detecting fetal pulmonary maturity with real-time convenience but reliable accuracy, infrared spectrophotometry of amniotic fluid lipids was introduced using our newly developed optical device. The peak areas at the wave number of 1740 cm-1 on the infrared spectra (called "PA1740") were measured in 29 specimens sampled from 29 pregnant women, and compared with the Lecithin/Sphingomyelin (L/S) ratio of the same samples. Statistical analysis revealed that the PA1740 was almost as reliable as the L/S ratio, and these indices illustrated pulmonary immaturity in cases with a PA1740 less than 0.9 and an L/S ratio less than 2.0. PA1740 is absolutely independent of the effect on the concentration of lecithin in the amniotic fluid because of a combination of the vacuum-drying process and the absorbance correction, and can be completed in less than 20 minutes from sampling of the amniotic fluid by a very simple manipulation. In conclusion, the new index, PA1740, will be a more practical method for prediction of the fetal pulmonary maturity than methods such as the well known L/S ratio.

Amniotic Fluid↗