Search PubMed⌕ Search

SEARCH · Search PubMed

Results for “Elbow Prosthesis”

Search indexed PubMed citations on genomics, clinical trials, systematic reviews and public health. Explore titles, authors and supplied subject terms, then open the PubMed record.

Quote a phrase for an exact phrase match. Source license links do not imply unrestricted reuse.

At least 397 records · Page 22Linked to original sources

Primary replacement of the fractured radial head with a metal prosthesis.

Compression testing of cadaver specimens showed that excision of the radial head allowed proximal radial displacement. The insertion of a metallic radial head restored normal mechanics, while a silicone rubber implant did not. We reviewed 31 of 36 comminuted fractures of the radial head, 21 associated with dislocation or ulnar fracture, which had been treated by primary replacement with a Vitallium prosthesis. At a mean follow-up of 4.5 years, there was reliable restoration of stability and prevention of proximal radial migration. There had been no dislocations or prosthetic failures, but two implants had been removed for loosening. The prosthesis is recommended for use as a spacer to stabilise the elbow after severe injuries while the soft tissues heal.

Adult↗

Single-stage revision of peri-prosthetic infection following total elbow replacement.

This study reviews the predisposing features, the clinical, and laboratory findings at the time of diagnosis and the results of single-stage revision of prosthetic replacement of the elbow for infection. Deep infection occurred in six of 305 (1.9%) primary total elbow replacements. The mean follow-up after revision was 6.8 years (6 months to 16 years) and the mean age at the time of revision was 62.7 years (56 to 74). All six cases with infection had rheumatoid arthritis and had received steroid therapy. The infective organism was Staphylococcus aureus. Four of the six elbows had a developed radiolucency around one component or the other. Successful single-stage exchange arthroplasty was carried out with antibiotic-loaded cement in five of the six cases. In one, the revision prosthesis had to be removed following recurrence of the infection. The functional result was good in three elbows, fair in one, poor in one and fair in the resection arthroplasty.

Aged↗

The elbow joint and its total arthroplasty. Part I. A state-of-the-art review.

Various pathologies, notably rheumatoid arthritis and/or trauma, are all known to cause severe pain and/or reduction of functional motion at the elbow joint. Implantation of a total arthroplasty, in such cases, has emerged as the first choice for management. For acceptable clinical performance, such an arthroplasty should be designed fully cognizant of various germane aspects of the anatomy, kinesiology and biomechanics of the normal joint. These aspects, together with a critical appraisal of the different classes of arthroplasty designs and a discussion of areas for further research (aimed at improving the in vivo performance of the prosthesis), are the subjects of the present review.

Elbow Joint↗

Late infection of joint prostheses.

Late infection of an orthopaedic prosthesis can cause serious illness in elderly patients. Four cases are described to show the importance of recognizing this condition.

Aged↗

Comminuted radial head fractures treated with a modular metallic radial head arthroplasty. Study of outcomes.

BACKGROUND: Comminuted fractures of the radial head are challenging to treat with open reduction and internal fixation. Radial head arthroplasty is an alternative treatment with results that compare favorably with those reported after open reduction and internal fixation of similar fractures. The purpose of this study was to evaluate the two-year outcomes and the rate of recovery of a closely followed cohort of patients in whom an unreconstructible radial head fracture had been treated with a modular metallic prosthesis. METHODS: Twenty-six patients (seventeen female and nine male; mean age, fifty-four years) with an unreconstructible comminuted radial head fracture and associated elbow injuries were treated with a modular metallic radial head arthroplasty. Patients who had presented more than four weeks following the injury or had had the radial head arthroplasty as a second-stage or salvage procedure were excluded. Of the twenty-six patients, twenty-two had an associated elbow dislocation, and thirteen of them also had an associated fracture of the coronoid process. Patients were prospectively followed at three, six, twelve, and twenty-four months. Self-reported limb function, general health, range of motion, and isometric strength were assessed by an independent observer. RESULTS: Following treatment of the injury, significant decreases in self-reported and measured impairments were noted over time, with the majority of the recovery occurring by six months and little further recovery noted between six and twenty-four months. There were slight-to-moderate deficits in the range of motion and strength compared with the values on the contralateral, unaffected side. Patient satisfaction was high at three months and remained high at two years. All elbow joints remained stable, no implant required revision, and there was no evidence of overstuffing of the joint. Mild osteoarthritis was seen in five (19%) of the twenty-six patients. CONCLUSIONS: An arthroplasty with a modular metallic radial head is a safe and effective option for the treatment of unreconstructible radial head fractures associated with other elbow injuries. Recovery primarily occurs by six months, with minimal additional improvements over the next eighteen months.

Adult↗

Radial head prosthesis with a mobile head.

Radial head replacement is a useful method in the treatment of comminuted radial head fractures. Because such injuries are a common clinical problem and many complications have been reported after radial head replacements, a new radial head prosthesis is proposed. This new design, based on the shape-dimensional identification of the radial head, consists of two parts. The head, made of ultra-high molecular weight polyethylene, is articulated with a Vitallium stem. A series of functional quality and strength tests were conducted on this new prosthetic design. The implant was also examined via the finite element method. General preclinical investigations of clinical cases show that this prosthesis is a very promising design.

Arthroplasty, Replacement↗

[Unreconstructible radial head fracture: resection, implant of Swanson or prosthesis? Retrospective comparative study].

Treatment of unreconstructible comminuted fractures of the radial head remains a therapeutic challenge. Thirty two patients sustained unreconstructible radial head fracture between 1969 and 1999 and have been treated by resection (16 patients), by Swanson implant (8 patients), or by Judet prosthesis (8 patients). The three groups of patients were reviewed clinically and radiologically by two surgeons not involved in their treatment. Functional outcomes of the elbow (Morrey scoring-SOO scoring system, instability, cubitus valgus) and involvement of the wrist (pain, grasp, RUD instability) were evaluated with a mean follow up of 15 years (6-27.7). Elbow or wrist arthritis, ulnar variance, and evolution of Swanson implants were evaluated on standard radiographs. The Judet prosthesis group was evaluated with DASH scoring. Half of the patients were male and 1/3 had a work-related injury. Regarding the following criteria, there was no significant difference between the resection and the Swanson group: Morrey scoring 77/100, SOO scoring 7.4/11, mean flexion was 130 degrees, mean deficient extension was 18 degrees, mean pronation 60 degrees, mean supination 67 degrees, grasp reached 90% of the controlateral side. Arthritis was noted with the same frequency at the elbow (87%), and wrist levels (66%) in each group. Ninety four percent (94%) of patients in the resection group and 89% in the Swanson group were satisfied. Excellent and good results were reported more frequently in the Swanson group (37% resection group, 51% Swanson group). In the resection group the following complications were significantly more frequent: ulnar nerve irritation (2 x), ulnar head dislocation (2 x), ulnar head instability (3 x), para articular ossification (5 x), ulnar variance positive in all cases (mean value 3.20 mm). In the Swanson group only two implants were found to be destroyed at follow-up. In the prosthesis group function and satisfaction were higher than the two other groups but with a shorter follow up. Excellent and good results were reported in 62.5% cases with a DASH scoring between 0 and 16.7. Comminuted fractures of the radial head treated by resection or Swanson implant are both followed by fair results (same functional scoring). Only the level of complications differs between groups: at elbow level for Swanson group, at wrist level for the resection group. Patients in the Swanson group were most frequently satisfied. Metal radial head implant is an attractive solution yielding good functional outcomes in recent reported limited series and in our personal experience, but with a very short follow-up.

Accidents, Occupational↗

A floating prosthesis for radial-head fractures.

We report our experience over seven years with a floating radial-head prosthesis for acute fractures of the radial head and the complications which may result from such injury. The prosthesis has an integrated articulation which allows change of position during movement of the elbow. We present the results in 12 patients with a minimum follow-up of two years. Five prostheses had been implanted shortly after injury with an average follow-up of 49 months and seven for the treatment of sequelae with an average follow-up of 43 months. All prostheses have performed well with an improved functional score (modified from Broberg and Morrey 1986). We have not experienced any of the complications previously reported with silicone radial-head replacement. Our initial results suggest that the prosthesis may be suitable for the early or delayed treatment of Mason type-III fractures and more complex injuries involving the radial head.

Adolescent↗

An intramedullary aligned bone cutting jig for elbow replacement.

One of the main post operative complications of total elbow replacement has been dislocation and fracture of the olecranon. This has been associated with rotational malalignment of the bone cuts of the proximal ulna caused by the current free hand cutting method. This paper presents an intramedullary aligned bone cutting jig which gives rotational location on the ulna for the bone cuts, without notching the olecranon.

Biomedical Engineering↗

Locking mechanism failure in semiconstrained total elbow arthroplasty.

Between 1988 and 2000, we performed 170 total elbow replacements using a semiconstrained implant (custom or standard). We are aware of 9 implants (5.3%) that have had late failure of the locking mechanism, as evidenced by migration of the connecting axle on radiographic examination. The mean time to failure of the locking mechanism was 5.5 years. Eight elbows required surgery for revision of the polyethylene bushing and locking mechanism by use of a custom-made locking clip. A limited approach was devised that allowed sufficient exposure to change the bushing and axle. In 1 of the 8 cases, the revised locking mechanism failed at 5.5 years, necessitating a reoperation. Of the 8 patients, 7 have not had recurrence of the failure at a mean of 1.4 years (minimum follow-up of 1 year). These 7 patients have good or excellent results based on the Hospital for Special Surgery elbow scoring system, with a mean arc of motion of 130 degrees.

Adult↗

Long-term results with the Kudo type 3 total elbow arthroplasty.

From 1992 to 1993, 17 elbows were replaced with the unlinked Kudo type 3 total elbow arthroplasty in 16 patients with arthritic joint destruction. Of these elbows, 8 were available for clinical examination after a mean of 9.5 years. Of the 17 elbow implants, 5 were revised: 2 because of loosening of the ulnar component, 1 because of dislocation, 1 because of a periprosthetic fracture, and 1 because of a late deep infection. At 9.5 years' follow-up, 67.9% of the prostheses had survived, and the mean survival of the implant was 8.7 years (95% confidence interval, 7.5-10 years). In this study, we observed high rates of progressive valgus tilting of the ulnar component. This valgus tilting is of major concern because it provides more stress to a smaller area on the polyethylene. It may, therefore, cause an increased degree of polyethylene wear and, thereby, reduce the final implant survival rate.

Adult↗

Determination of the optimal elbow axis for evaluation of placement of prostheses.

OBJECTIVE: To present a method to determine the position and orientation of the mean optimal flexion axis of the elbow in vivo to be used in clinical research. DESIGN: Registering the movements of the forearm with respect to the upper arm during five cycles of flexion and extension of the elbow using a 6 degrees-of-freedom electromagnetic tracking device. BACKGROUND: Loosening of elbow endoprostheses could be caused by not placing the prostheses in a biomechanically optimal way. To evaluate the placement of endoprostheses with regard to loosening, a method to determine the elbow axis is needed. METHODS: The movements of the right forearm with respect to the upper arm during flexion and extension were registered with a 6 degrees-of-freedom electromagnetic tracking device. A mean optimal instantaneous helical axis of 10 elbows was calculated in a coordinate system related to the humerus. RESULTS: The average position of the flexion/extension axis was 0.81 cm (SD 0.66 cm) cranially and 1.86 cm (SD 0.72 cm) ventrally of the epicondylus lateralis. The average angle with the frontal plane was 15.3 degrees (SD 2 degrees). CONCLUSIONS: A useful estimation of the position and orientation of a mean optimal flexion axis can be obtained in vivo.

Adult↗

The effect of cement restrictors on the occlusion of the humeral canal: an in vitro comparative study of 2 devices.

An in vitro study investigated the efficacy of cement restrictor devices on cement containment and penetration within the humerus. Eight pairs of preserved humeri were prepared using advanced cementing technique followed by insertion of 1 of 2 cement restrictors. Low-viscosity cement was injected followed by sham humeral stem insertion. Each specimen was sectioned into transverse 10-mm slices. Slices were photographed and digitized to quantify the cement/stem and canal cross-sectional area. Cement penetration was determined from the ratio of cement area to canal area. A significant increase in cement penetration was observed among slices from distal to proximal for both restrictors (P = .02). There was no significant difference in cement penetration, leakage, or migration between restrictors. Five of the 8 specimens migrated, with means of 21.5 +/- 25.0 mm and 24.0 +/- 36.0 mm for the polyethylene and silicone restrictors. Leakage or migration resulted in a significant decrease in cement penetration with the polyethylene restrictors (P = .001). In the silicone restrictor group, migration resulted in decreased cement penetration (P = .04). When using advanced cementing techniques, intramedullary restrictors allow improved cement penetration; however, they do not ensure cement containment.

Arthroplasty, Replacement↗

Early failure of silicone radial head implants in the rheumatoid elbow. A complication of silicone radial head implant arthroplasty.

Three cases of early recurrent synovitis of the rheumatoid elbow following silicone radial head implant arthroplasty are presented. All three patients underwent synovectomy with silicone radial head implant for pain and loss of function due to rheumatoid synovitis. All three implants failed within 4-9 months after surgery. This failure was associated with radiographic medial joint space widening. At reexploration, recurrent synovitis with hemosiderin pigmentation was noted. Patterns of prosthetic failure included fracture, fraying, and compressive deformation or rotary wear ("motar and pestle") of the silicone radial head; a prosthetic stem fracture was also present. Histologic examination of the synovium revealed evidence of silicone-induced synovitis; particulate, refractile silicone debris was associated with inflammatory and foreign body giant cells. The presence of rheumatoid destruction of the joint surface may have accelerated this process and may be a relative contraindication to silicone radial head implant arthroplasty.

Adult↗

Capitello-Condylar total elbow arthroplasty for rheumatoid arthritis.

Twenty-one total elbow arthroplasties using the Capitello-Condylar design of Ewald were performed at the Denver Orthopedic Clinic from 1977 to 1986. Indications for operation included intractable pain unrelieved by medical therapy as well as late state III or IV rheumatoid arthritis. The average total elbow score at 48 months after operation was 89.5 points, including excellent pain and function ratings. Modest gains in motion occurred with lesser gains in patients with juvenile rheumatoid arthritis than those of the adult variety. Forty percent of the elbows suffered some type of complication although most are successfully manageable. There was one clinical failure due to deep infection in a renal deficient patient.

Adult↗

Nonconstrained total elbow arthroplasty. Development and results in patients with functional class IV rheumatoid arthritis.

Twenty-four nonconstrained total elbow arthroplasties (TEAs) in 20 patients with severe disabling functional class IV rheumatoid arthritis were reviewed retrospectively to determine the effects of their disease on the end result. The mean follow-up period was 6 years (range, 2-11 years). All planes of elbow motion improved after surgery (P less than .005), there was a highly significant reduction in pain and improvement in function (both P less than .005) and patient satisfaction was high. The severity of the disease does not appear to affect the end result when a nonconstrained TEA is used, and the results compare favorably with those of other series reported.

Adult↗