PubMed · 9682170
Automated sample preparation for drugs in plasma using a solid-phase extraction workstation.
Abstract
An automated solid-phase extraction workstation was used to develop, characterize and validate two separate HPLC methods for quantifying drugs in plasma. Method development was facilitated by workstation functions which allowed wash solvents of varying organic composition to be mixed and tested automatically. The precision estimates for the two methods were within 6.0 and 2.0% RSD across their respective calibration ranges. Accuracies for replicate determinations of quality controls were between -1.2 and +4.8 RE over ng ml-1 calibration ranges, respectively. Optimized recoveries were quantitative and were generally greater than 90% for the four analytes tested, and depended to a great extent, as expected, on the composition of the wash solvent. Sample throughput benchmarks for the two methods ranged from 3 to 10 min per sample, depending on the extent of air drying used. Because of parallel sample processing, 60 samples could be extracted in as little as 17 min.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
T D Parker, N Surendran, B H Stewart, D T Rossi. 1998. Automated sample preparation for drugs in plasma using a solid-phase extraction workstation.. https://doi.org/10.1016/s0731-7085(97)00278-1. https://searchpubmed.org/record/pubmed%3A9682170/automated-sample-preparation-for-drugs-in-plasma-using-a-solid-phase-extraction-workstation
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search PubMed by Indiciae: BibTeX · RIS.