Reprogramming of implanted pacemaker following external defibrillation.

Abstract

This report describes reprogramming of a pulse generator consequent to cardiac defibrillation. Analysis of the explanted pulse generator revealed normal function. We tested the electrical properties of several defibrillators and our studies suggest that erratic electrostatic discharges at the time of paddle application before defibrillation may create a signal sequence capable of reprogramming the particular pulse generator by activating either the reed switch or internal electronics.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

S S Barold, L S Ong, J Scovil, R A Heinle, T Wright. 1978. Reprogramming of implanted pacemaker following external defibrillation.. https://doi.org/10.1111/j.1540-8159.1978.tb03514.x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A biventricular ICD system with biventricular defibrillation.

We describe the case of a 59-year-old gentleman with severe dilated cardiomyopathy requiring implantation of a dual-chamber biventricular implantable cardioverter-defibrillator (ICD). High defibrillation thresholds (DFT) were encountered at implant with an inadequate defibrillation safety margin. Testing of all possible shock vectors/polarities with and without the SVC coil and optimization of the distal RV coil position all proved inadequate. A satisfactory defibrillation safety margin was achieved following placement of a second lead in the coronary sinus to enable biventricular defibrillation. This case highlights an additional strategy for combating high DFTs and is an option even in dual-chamber biventricular ICD systems.

Electric Countershock↗