PubMed · 9481863
Complement-binding anti-Jka not detectable by DiaMed gels.
Abstract
BACKGROUND: An anti-Jka was not detectable by the DiaMed-ID Micro Typing System but was strongly reactive when tested using other indirect antiglobulin test (IAT) methods. However, it was detectable in the DiaMed-ID system when the recommended diluent ID-Diluent 2, was replaced by other low ionic Strength saline (LISS) solutions. This led us to believe that the problem lay with the ID-Diluent 2. METHODS: The anti-Jka was investigated using various low ionic strength diluents in the IAT against both polyspecific and monospecific antiglobulin reagents. We tested the ID-Diluent 2 for its ability to inhibit complement activation, and bind calcium. RESULTS: The anti-Jka was shown to be a 'complement-only' binding antibody that reacted in all IATs performed except when tested with ID-Diluent 2. The ID-Diluent was shown to prevent haemolysis of ABO lytic antibodies, and further testing shows that it binds calcium. CONCLUSION: The use of a diluent which prevents haemolysis and binds calcium in an IAT system that contains anti-complement as well as anti-IgG must be questioned.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
P O'Brien, L Hopkins, D McCarthy, S Murphy. 1998. Complement-binding anti-Jka not detectable by DiaMed gels.. https://searchpubmed.org/record/pubmed%3A9481863/complement-binding-anti-jka-not-detectable-by-diamed-gels
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search PubMed by Indiciae: BibTeX · RIS.