PubMed · 914928
Quantitation with high-performance thin-layer chromatography and programmed multiple development with high-performance micro-thin-layer material for drug analyses in biological fluids.
Abstract
The programmed multiple development thin-layer chromatography (PMD-TLC) technique was compared with the conventional chromatographic technique on both normal and HPTLC plates. The potency of this high-performance TLC is illustrated and discussed by means of the data obtained with the determination of digitoxin in human serum. Improvement of the efficiency of TLC, resulting in a better resolution and sensitivity with PMD-TLC using using HPTLC plates, makes this high-performance TLC technique comparable to high-performance liquid chromatography. This TLC approach might lead to a sensitive, rapid, selective and simply assay for routine serial analyses and, because of its specificity and flexibility, it may facilitate drug interaction studies.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
D B Faber. 1977-11-11. Quantitation with high-performance thin-layer chromatography and programmed multiple development with high-performance micro-thin-layer material for drug analyses in biological fluids.. https://doi.org/10.1016/s0021-9673(01)92055-2. https://searchpubmed.org/record/pubmed%3A914928/quantitation-with-high-performance-thin-layer-chromatography-and-programmed-multiple-development-wit