Search PubMed⌕ Search

PubMed · 8829940

Five oxygen-nitrous oxide proportioning systems compared.

Abstract

The majority of contemporary gas delivery modules, on anaesthetic workstations, are equipped with oxygen-nitrous oxide proportioning systems which should prevent the delivery of hypoxic gas mixtures. We investigated five modules, of two different types, using fresh gas flows ranging from 50 mL min-1 to 20 L min-1 with minimally acceptable proportions of oxygen and maximally acceptable proportions of nitrous oxide set at the flow control valves. The oxygen concentrations of the resulting gas mixtures were measured at the fresh gas outlet of the gas delivery modules. All the modules prevented the delivery of hypoxic fresh gas flows both under normal and abnormal working conditions (unequal supply of gas pressures). All systems showed increased O2 concentrations at fresh gas flows below 1 L min-1. The systems can be used with low flow anaesthesia techniques with one exception (Dameca) also with minimal flow techniques.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

P Feigenwinter, T H Pohle, A M Zbinden. 1996. Five oxygen-nitrous oxide proportioning systems compared.. https://doi.org/10.1097/00003643-199601000-00013

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Perception of training needs and opportunities in advanced airway skills: a survey of British and Irish trainees.

BACKGROUND AND OBJECTIVE: We surveyed delegates at the Group of Anaesthetists in Training (UK) meeting to investigate evidence of a training-gap (number of fibreoptic intubations believed to bestow competence vs. number actually performed). METHODS: Questionnaires were distributed to and collected from delegates in person. Questions covered six areas, including experience of fibreoptic intubation and cricothyrotomy, fibreoptic intubation as a specialist skill and ethical issues. RESULTS: We received 221 replies (76%). All trainees believed competence to be achievable with 10 intubations (interquartile range (IQR) 10-20); the median number performed was 2 (IQR 0-4). This was statistically significant for the groups' senior house officers, 1st and 2nd year registrars and 3rd and 4th year registrars; P < 0.0001. Many final year trainees (12/20, 60%) also failed to achieve their competency target. Few trainees had seen or performed any cricothyrotomies (medians 0, IQRs 0-1 and 0-0). Most (195/208, 94%) believed that fibreoptic intubation was a core skill and 199/212 (94%) believed that all should be competent by completion of training. Ten percent (n = 208) felt it unethical to perform an awake training intubation with full consent and 10% believed it acceptable without explanation. Most (82.7%) would fibreoptically intubate an asleep patient (requiring intubation) without consent. CONCLUSION: Trainees reported a gap between their perception of competence and achievement in awake fibreoptic intubation. Simple and complex simulations and structured training programmes may help. Anaesthetists must address the ethics of clinical training in advanced airway management.

Anesthesiology↗