Tissue compensation using dynamic collimation on a linear accelerator.

Abstract

PURPOSE: The availability of computer-controlled collimators on some accelerators has led to techniques for dynamic beam modification, mainly to simulate beam wedge filters. This work addresses the practical aspects of dynamic tissue compensation in one dimension using available treatment-planning software. METHODS AND MATERIALS: Data derived from the treatment-planning program is used with an iterative calculational routine to determine the monitor unit settings needed for the collimator-controlling computer. The method was first tested by simulating a 60 degrees physical wedge. Further studies were carried out on a specially fabricated plastic phantom that modeled the sagittal contour of the upper torso, neck, and lower head regions. RESULTS: Dynamic wedge point doses generated by the planning program agreed within 1% with the values directly measured in a polystyrene phantom. In the patient phantom, dynamic collimation achieved calculated dose uniformity within 0.5% in a reference plane near the phantom midline. A comparison of computer-generated and measured point doses in this case showed agreement within 3%. CONCLUSIONS: Dynamic collimation can provide effective compensation for contours that vary primarily along one direction. A conventional treatment-planning program can be used to plan dynamic collimation and deliver a prescribed dose with reliable accuracy.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

H E Gaballa, G Mitev, R D Zwicker, J Y Ting, O F Tercilla. 1995-06-15. Tissue compensation using dynamic collimation on a linear accelerator.. https://doi.org/10.1016/0360-3016(94)00593-a

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Correction of phase aberrations for sectored annular array ultrasound transducers.

Two methods for correction of unknown phase aberrations induced by inhomogeneous acoustic velocities in tissues are explored for the two dimensional geometry of a sectored annular array system. The methods employed are adaptations of a cross correlation technique and a speckle brightness maximization technique. The methods correct phase distortions via the introduction of phase shifts in the timing sequence at the beamformer stage of a sectored annular array transducer. The techniques are investigated employing software models and a computer controlled automated scanning system. A 65-element sectored annular array is modelled via a rotating 5 element transducer. Tissue equivalent materials were moulded into a double layer aberrating medium to simulate phase distortions encountered in the rectus abdominis muscle in vivo. A comparison of the effectiveness of the two correction methods is presented. Contrast of an anechoic region is increased from 0.34 +/- 0.08 to 0.48 +/- 0.06 for the cross correlation technique, and up to 0.62 +/- 0.05 for the speckle brightness maximization method. The performance of these correction techniques on target phantoms suggests that considerable improvements in image quality should be possible for clinical systems.

Models, Anatomic

Quality assurance control in the EORTC cooperative group of radiotherapy. 3. Intercomparison in an anatomical phantom.

Two papers concerning the quality control study organised by the EORTC (European Organisation for Research on Treatment of Cancer) Cooperative Group on Radiotherapy have been published. The medical profile (part 1) and the dosimetric intercomparison (part 2) of the participating institutions were presented. This part (paper 3) presents an integrated clinical and dosimetric investigation in an anatomical phantom. A tonsillar tumour and a homolateral subdigastric node were marked in an anatomical phantom. The institutions were asked to treat the phantom once like an ordinary patient. The phantom was loaded with dosimeters and irradiated. From the results obtained, it can be concluded that we did not find any major dosimetric problem related to absorbed dose calibration or calculation in the phantom. However, several major problems were related to non-optimal planning, treatment technique and dose distribution. The investigation shows the importance of a quality assurance programme for cooperative groups.

Models, Anatomic