[Development of a control method for sample distributors].

Abstract

The use in routine of distribution automations in Blood Bank laboratories should imply regular testing of the equipment by users. This work was carried out on Microlabs as well as on an IEMS. We used a coloured solution (Alizarine Red in pH9 buffer). We applied the Beer-Lambert law, knowing that any volume modification induces a change in optical density. After testing the linearity of the photometers used in the laboratory, we set up 4 distribution programmes so as to simulate classic distribution of diluted or undiluted serum: 1) for large volume dispensing, 2) for small volume distribution in a small amount of buffer, 3) for distribution of small volumes in large buffer volumes, 4) a specific IEMS programme. Results, expressed as coefficients of variation (CV) for each volume, varied from 9.8 to 0.7% according to the programme. These results show that large volume distribution does not pose any problem (CV < 5%). When dilution occurs, the error observed in serum distribution adds up to that of the thinner, and the smaller the two volumes, the greater is the error. It therefore appears necessary to find the best compromise between serum volume and thinner volume to obtain a CV < 5%. This error may appear negligible in semi-quantitative tests, but it may be important in quantitative tests. Lastly, this programme can easily be adapted in routine to monitor the dispensers efficiently.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

F Durand, B Annic, B Genetet. 1995. [Development of a control method for sample distributors].. https://doi.org/10.1016/s1246-7820(05)80005-9

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Therapeutic drug monitoring of tacrolimus with the dried blood spot method.

In a preliminary investigation an assay for tacrolimus based on fingerprick sampling and consecutive application as a blood spot on sampling paper has been developed. The dried blood spot was analysed by HPLC-tandem mass spectrometry. The validated range was 1-30 microg/l. Intra- and inter-assay variability for precision and accuracy was <7.5% and 15%, respectively. Tacrolimus concentrations of 24 stable out patients were compared after both blood spot sampling and conventional venous sampling. Method agreement was investigated with the methods of Passing and Bablok and Bland Altman and proved suitable for clinical use. The dried blood spot method for tacrolimus seems promising for patient monitoring.

Blood Specimen Collection↗