An improved image algorithm for CT scanners.

Abstract

A common artifact in CT head-section images is a cupping or broad "whitening" effect near the skull which is due at least in part to the polychromaticity of the x-ray beam. In this paper, a general method is presented for removing this artifact empirically by a combination of two approaches. The gross cupping is removed by modifying the raw transmission data prior to reconstruction. The residual whitening near the bone is removed conveniently by modifying the reconstruction filter function. Examples of the modifications are shown using the AS&E CT scanner. The method convolves or deconvolves the CT image with an appropriate point spread function. Since the filter-function modifications are done conceptually in real space rather than in frequency space, the details of the modifications are more easily understood.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R C Chase, J A Stein. An improved image algorithm for CT scanners.. https://doi.org/10.1118/1.594486

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Sensitivity coefficients, trajectories and graphs of a human head-movement model.

With time as the criterion to be optimized, the divergences from optimal head movements show systematic differences in their control signal variables with respect to single behaviours. To clarify these relationships, this study applies manipulation and mathematical analysis of the 6th-order nonlinear head movement model, using the three-fold approach of sensitivity analysis. The sensitivity analysis of the improved and refined head movement model explains the different tasks of the control parameters and their relation to the plant. It shows that width more than height of the agonistic and antagonistic pulses dominate the important behaviours of the movement: acceleration, magnitude and duration.

Head