PubMed · 6610037
Two algorithms for use with an orthogonal-view coded-aperture system.
Abstract
ohe goal of single photon emission computerized tomography is to map out a three-dimensional distribution of a radionuclide that is concentrated in a structure of interest. There are a number of imaging modalities that achieve this goal with varying degrees of success. In this study, computer simulation is used to explore a novel imaging modality, orthogonal-view coded-aperture imaging. Furthermore, a comparison is made between two reconstruction algorithms, one being an iterative back-projection algorithm and the other a Monte Carlo algorithm. Finally, evidence is presented suggesting that a deficiency in the projection data due to multiplexing is less disturbing than that due to limited angular range.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
R G Paxman, W E Smith, H H Barrett. 1984. Two algorithms for use with an orthogonal-view coded-aperture system.. https://searchpubmed.org/record/pubmed%3A6610037/two-algorithms-for-use-with-an-orthogonal-view-coded-aperture-system
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search PubMed by Indiciae: BibTeX · RIS.