From Peaks to Power: Systematic Evaluation of Chromatographic Sampling Reveals Determinants of Quantification and Biological Discovery in DIA Proteomics.

Abstract

Modern DIA proteomics increasingly emphasizes throughput and depth for large-cohort studies, but methods are often optimized using proxy metrics that can mask losses in quantifiable signal and statistical power. Here, we evaluate how data points per peak and other chromatographic features jointly contribute to quantification and downstream biological discovery. Using a matrix-matched calibration curve dataset, we checked how the number of data points per peak (DPPP) affects the limits of detection and quantification (LOD/LOQ). Reduced DPPP minimally affected LOD but substantially degraded LOQ. Feature modeling and nonparametric association analyses identified precursor peak area as the strongest feature-level predictor of LOQ, whereas DPPP showed weaker and context-dependent effects. Simulations of chromatographic peak integration recapitulated these trends, showing that increased sampling primarily improves integration precision, while quantitative accuracy is strongly governed by peak height and peak shape. Finally, when comparing 20 cancer vs 20 control plasma samples processed with Seer Proteograph, the decrease in DPPP led to a loss of statistical significance for proteins with low-abundance precursors. These findings argue that DIA optimization should prioritize LOQ and statistical power metrics─not identifications alone─by balancing sampling density with chromatographic peak height and quality to maximize useful biological signal.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Lee S Cantrell, Seth Just, Alexey Stukalov, Omid C Farokhzad, Serafim Batzoglou. 2026-09-04. From Peaks to Power: Systematic Evaluation of Chromatographic Sampling Reveals Determinants of Quantification and Biological Discovery in DIA Proteomics.. https://doi.org/10.1021/acs.jproteome.6c00415

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Multi-omic analyses of the same sample using metabolomics, lipidomics, proteomics, phosphoproteomics, and glycoproteomics.

Mass spectrometry (MS)-based multi-omics offers powerful tools to comprehensively characterize proteins, post-translational modifications, metabolites, and lipids. However, these measurements are typically performed using separate sample preparation workflows and modality-specific liquid chromatography mass spectrometry (LC-MS) platforms, limiting integration and constraining applications to small amounts of sample materials, especially scarce clinical specimens. Here, we describe a unified nano-LC-MS framework that enables metabolomic, lipidomic, proteomic, phosphoproteomic, and glycoproteomic analyses from the same starting material using a single nano-LC-MS platform, with only the chromatographic conditions, acquisition methods, and enrichment procedures tailored to each omics. This integrated strategy reduces workflow complexity and sample consumption while improves analytical continuity across molecular layers. By enabling deep multi-omics characterization from the same sample, this platform provides a practical foundation for comprehensive analysis of precious clinical samples.

Proteomics

4D-DIA proteomics reveals distinct proteolytic landscapes induced by mechanical stress, Agrobacterium, and a viral capsid precursor.

Nicotiana benthamiana is a widely used platform for plant molecular farming, yet recombinant protein yields are frequently compromised by the host's innate defense mechanisms, particularly proteolytic degradation. While the general effects of Agroinfiltration are known, the distinct contributions of mechanical injury, bacterial perception, and product-specific stress remain poorly resolved. Here we utilized high-depth 4D-DIA proteomics to dissect the host response across three dimensions: physical stress (buffer infiltration), pathogen-associated stress (Agrobacterium), and product-associated stress (GFP vs. the FMDV capsid precursor P1_2A). We demonstrate that buffer infiltration is not a neutral event but an independent inducer of cell wall remodeling and oxidative stress. By filtering out these background effects, we defined a core Agrobacterium-responsive proteome characterized by a growth-defense trade-off. We also expanded the known protease repertoire of N. benthamiana to 1,505 enzymes through improved genomic annotation. We found that the expression of the FMDV capsid precursor P1_2A was associated with a distinct and more pronounced protease profile compared to soluble GFP, characterized by the upregulation of subtilases and cysteine proteases. These findings suggest that host proteolytic responses vary with the recombinant cargo, a factor worth considering when designing engineering strategies for the production of complex biopharmaceuticals in plants.

Proteomics

An Instrumental Optimization of a Label-Free Proteomic Method for Trace Protein Input.

Liquid chromatography-mass spectrometry (LC-MS)-based proteomics of trace-level samples, such as tens of cells or spatially resolved tissue regions, offers unique biological insights but is often constrained by the requirement for specialized, costly instrumentation. In this study, we developed a scalable workflow for the deep proteomic analysis of low- to ultralow-input samples by systematically optimizing a widely adopted Orbitrap and UHPLC platform to maximize sensitivity, precision, and throughput. This optimized workflow identified over 5600 proteins from 5 ng of peptides and 3400 proteins from 20 sorted cells, achieving a throughput of 30 analyses per day while maintaining deep proteome coverage and high quantitative reproducibility. Furthermore, by applying this method to spatially resolved proteomics, we identified over 6100 proteins from microscale regions of interest (ROIs) within a formalin-fixed, paraffin-embedded (FFPE) tissue. A data-driven normalization strategy was employed to correct for variable cellularity across tissue regions, effectively revealing intratumor heterogeneity and distinct molecular and functional signatures, including pathway activations not apparent in parallel spatial transcriptomic analysis. Ultimately, this accessible, high-performance method substantially lowers the instrumentation barrier for the deep proteomic profiling of trace-level biological samples.

Proteomics