Construction and performance of a variable-frequency phase-modulation fluorometer.

Abstract

We describe the construction and performance of a variable-frequency phase-modulation fluorometer. This instrument, which provides modulation frequencies from 1 to 200 MHz, was constructed using commercially available components. To facilitate the introduction of these instruments into other laboratories we describe in detail the chosen components and the principles of operation. The present light source is a continuous-wave helium-cadmium laser, which provides convenient excitation wavelengths of 325 and 442 nm. Modulation of the incident light is provided by one of several electro-optic modulators. The extent of modulation ranges from 1.0 to 0.2 as the frequency increases from 1 to 200 MHz. Phase angles and demodulation factors are measured using the cross-correlation method. The closely spaced frequencies are provided by two direct frequency synthesizers. The phase and modulation measurements are accurate to 0.2 degrees and 0.002, respectively, from 1 to 200 MHz. This accuracy allows considerable resolution of complex decay laws. The usefulness of frequency-domain fluorometry for the resolution of multiexponential decays is illustrated by the analysis of several difficult mixtures. As examples, we resolved a two-component mixture of anthracene (4.1 ns) and 9,10-diphenylanthracene (6.3 ns), and confirmed that the intensity decay of NADH in aqueous buffer is at least a double exponential (0.2 and 0.86 ns). We also resolved an especially difficult mixture of anthracene (4.1 ns) and 9-methylanthracene (4.5 ns), and a three-component mixture with decay times of 1.3, 4.1 and 7.7 ns. Frequency-domain fluorometers appear to be particularly useful for determination of complex decays of fluorescence anisotropy. This capability is illustrated by the determination of rotational correlation times as short as 47 ps for p-bis[2-(5-phenyloxazolyl)]benzene (POPOP) in hexane at 40 degrees C, and by the resolution of the two correlation times of anisotropic rotators such as perylene and 9-aminoacridine. Resolution of two anisotropy decay times for 9-aminoacridine is a difficult test because these correlation times differ by less than 2-fold. The resolution of multiexponential decays of intensity and anisotropy possible with this instrument is at least equivalent to that obtained using state-of-the-art time-resolved instruments based on mode-locked laser sources. The ease and rapidity of frequency-domain measurements, the relative simplicity of the equipment, the accuracy of the measurements and the lack of significant systematic errors indicate that frequency-domain fluorometry will be widely useful in chemical and biochemical research.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J R Lakowicz, B P Maliwal. 1985. Construction and performance of a variable-frequency phase-modulation fluorometer.. https://doi.org/10.1016/0301-4622(85)85007-9

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Analysis of subcellular sized particles. Capillary electrophoresis with post-column laser-induced fluorescence detection versus flow cytometry.

Flow cytometry (FCM) and more recently capillary electrophoresis with post-column laser-induced fluorescence detection (CE-LIF) have both been used for subcellular particle analysis but their analytical performance has not been compared. In this work, we compare a commercial FCM with an in-house built CE-LIF instrument using fluorescently labeled microspheres and isolated mitochondria. As evidenced by the relative standard deviation (RSD) of the individual fluorescence intensities, FCM is two-fold better than CE-LIF for microspheres with > or =1.5 x 10(6) molecules of equivalent soluble fluorescein (MESF). However, FCM has a comparatively low signal-to-noise ratio (S/N) and high RSD for microspheres with <1.5 x 10(6) MESF. CE-LIF, on the other hand, produces S/N ratios that are >25 times higher than FCM for all the microspheres tested and a lower RSD for microspheres with <1.5 x 10(6) MESF. When 10-N-nonyl acridine orange (NAO)-labeled mitochondria are analyzed, the S/N ratios of both techniques are similar. This appears to result from photobleaching of NAO-labeled mitochondria as they are detected by the LIF detector of the CE-LIF instrument. Both techniques have a niche in subcellular analysis; FCM has the advantage of collecting data for thousands of particles quickly, whereas CE-LIF consumes less than a nanoliter of sample and provides the electrophoretic mobility for individual particles.

Aminoacridines↗

A real-time fluorescence assay for measuring N-dealkylation.

A real-time fluorescence assay system using a series of 9-N-(alkylamino)acridine derivatives (methyl, ethyl, n-propyl, n-butyl, n-pentyl, and benzyl) that are N-dealkylated to 9-aminoacridine (9AA) is described. The product, 9AA, is approximately 27-fold more fluorescent than the substrates using excitation and emission wavelengths of 405 and 455 nm, respectively. Tests using expressed CYP1A1, 1A2, 3A4, 3A5, 1B1, 2C9, 2C19, and 2D6 indicated that N-dealkylase activity is specific for CYP1A1 and CYP2D6. CYP2D6 N-dealkylated methyl, ethyl, n-propyl, and n-butyl substrates, whereas CYP1A1 N-dealkylated these plus the n-pentyl derivative. Activities using 5 microM 9-N-(alkylamino)acridine substrates ranged from 0.1 to 0.9 pmol 9AA/min/pmol P450. Kinetic constants for CYP1A1 N-dealkylation of the 9-N-(methylamino)acridine (MAA) and 9-N-(ethylamino)acridine (EAA) were K(m) 1.09 +/- 0.68 and 0.35 +/- 0.21 microM and the V(max) 61.9 +/- 48.5 and 113.8 +/- 8.4 pmol 9AA/min/pmol CYP1A1, respectively. Kinetic constants for CYP2D6 N-dealkylation of MAA and EAA were K(m) 7.9 +/- 5.4 and 3.2 +/- 1.6 microM, and V(max) 501 +/- 35.4 and 702.7 +/- 257 pmol 9AA/min/pmol CYP2D6, respectively. The experimental binding energies (DeltaG(bind)) were calculated for MAA with CYP1A1 and CYP2D6 to be -8.266 and -7.074 kcal/mol, respectively. The DeltaG(bind) values for EAA with CYP1A1 and CYP2D6 were -8.950 and -7.618 kcal/mol, respectively. The substrates were suitable for monitoring N-dealkylase activity in microsomal preparations (human, rat, and monkey hepatic preparations) and human hepatocellular carcinoma cell suspensions. Assays were conducted by monitoring reactions either in 96-well microtiter plates using a fluorescence plate reader or in cuvettes using a spectrofluorimeter.

Aminoacridines↗

Fast and sensitive DNA hybridization assays using microwave-accelerated metal-enhanced fluorescence.

A new, fast, and sensitive DNA hybridization assay platform based on microwave-accelerated metal-enhanced fluorescence (MAMEF) is presented. Thiolated oligonucleotide anchors were immobilized onto silver nanoparticles on a glass substrate. The hybridization of the complementary fluorescein-labeled DNA target with the surface-bound oligonucleotides was completed within 20 s upon heating with low-power microwaves. In addition, the signal is optically amplified, a consequence of close proximity of the fluorophore to the silvered substrate. In this proof-of-principle methodology, as low as 50 nM of a target DNA was detected, although we envisage far-lower detection limits. Control experiments, where the surface-bound oligonucleotide was omitted, were also performed to determine the extent of non-specific binding. In these studies a significantly reduced non-specific adsorption was found when using microwave heating near to silvered structures as compared to room temperature incubation. These findings suggest that MAMEF could be a most useful alternative to the DNA hybridization assays used today, especially with regard to substantially increasing both the assay rapidity and sensitivity.

Aminoacridines↗