PubMed · 3352481
Errors in T1-determination using multislice technique and Gaussian slice profiles.
Abstract
Multislice SE- or IR-sequences with different TR- and T1-times, respectively, are often used, to determine T1- and T2-values for tissue characterisation. To investigate the perturbating influence of the sequential excitation in multislice technique, we measured the signal intensities and calculated the T1-values of phantoms as a function of gap widths between neighboured slices. Phantoms contained fluids of different T1. We found a strong dependence of signal intensities and therefore of calculated T1-values on the gap width, when a T1/TR-ratio of more than 0.3-0.4 was reached. T1-values are considerably overestimated in this case, whereas T2-values are not influenced by the sequential excitation. We conclude that reliable T1-measurements necessitate a large spacing of more than 1 slice thickness between adjacent slices.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
M Just, H P Higer, P Pfannenstiel. 1988. Errors in T1-determination using multislice technique and Gaussian slice profiles.. https://doi.org/10.1016/0730-725x(88)90523-1. https://searchpubmed.org/record/pubmed%3A3352481/errors-in-t1-determination-using-multislice-technique-and-gaussian-slice-profiles
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search PubMed by Indiciae: BibTeX · RIS.