Radiometric procedure for linearising computer-interfaced display systems.

Abstract

Procedures for the adjustment of computer-interfaced display devices have been based on a subjective approach, generally relying on the visual examination of digital test patterns. The lack of an objective adjustment procedure means that the best image quality may not always be achieved. This paper describes how a simple home-made radiometer can be used to make screen luminance as well as density measurements on transparency film. The measurements are first used to determine the brightness and contrast settings for the display which enable the film to be used in the linear range of its optical densities. The display device characteristics, showing how luminance varies with grey level, are then determined. Using this information a simple display-linearising mapping can be produced to ensure that screen luminance (and hence film density) is a linear function of display grey-level value. Film-density measurements are also used to investigate the luminance and hence phosphor uniformity characteristics of the screen. The linearisation of the display device is an essential step in optimising the quality of recorded images. The measurements described are reproducible and sufficiently simple to provide the basis for a reliable quality-control procedure for the periodic assessment of the performance and settings of the display device.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

T Woldeselassie, P F Sharp. 1988. Radiometric procedure for linearising computer-interfaced display systems.. https://doi.org/10.1088/0031-9155%2F33%2F7%2F005

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Rapid and accurate measurement for phase-change optical recording bits.

Conducting atomic force microscopy (CAFM) and scanning surface potential microscopy (SSPM) have been used to image the phase-change optical recording bits. Commercially available digital versatile discs (DVD) + rewritable (RW) with initialization process were measured in experiments. Comparing the measurement results of both, the measurement resolution of CAFM is far superior to that of SSPM. With the DVD + RW disc rotating at a linear speed of 3.5 m/s, appropriate writing laser power range, may be precisely identified by CAFM as 10-15 mW. This is sufficient to verify the high-resolution recording bits research method. This new method may also be applied to the development of new types of phase-change recording materials.

Image Processing, Computer-Assisted↗

Image analysis by pulse coupled neural networks (PCNN)--a novel approach in granule size characterization.

A biologically inspired spiking neural network model, the pulse coupled neural network (PCNN), has been applied for the first time in bulk particle characterization, and specifically in the characterization of pharmaceutical granule size distributions. The PCNN was trained on surface images of pharmaceutical granule beds, and the adjustable parameters (radius neuron interconnection, r0, linking weight coefficient, beta, local threshold potential, VTheta, and number of iterations) were successfully optimized using design of experiments. As demonstrated with size fractions of granules, it was found that the PCNN produced granule size-dependent signals. In general, a first highest and relatively narrow peak located in the region of two to twelve iterations corresponded to smaller particle size, while larger particles resulted in wider peaks and in highest (not first) peak at a range between 13 and 25 iterations. Better predictions, i.e. lower RMSEP (root mean squared error of prediction) values, were obtained using high beta value, low r0 and VTheta values, while the number of iterations had to exceed 110 and the optimized model (RMSEP lower than 5) corresponded to PCNN variables: r0=1, beta=0.4, VTheta=2, and number of iterations=150. The coefficient of determination (R2) of the model was 0.94 and the predicted variation (Q2) was 0.91, while the Pearson correlation coefficient between the predicted and the measured mean particle size by sieving for eight test batches was 0.98. These findings could be characterized as promising and encouraging for the further use of image analysis by PCNNs in pharmaceutical bulk particle size and shape characterization.

Image Processing, Computer-Assisted↗

Production of siRNA- and cDNA-transfected cell arrays on noncoated chambered coverglass for high-content screening microscopy in living cells.

In this chapter, we provide a protocol for the production of transfected cell arrays in living mammalian cells on noncoated chambered coverglass for the systematic functional analyses of human genes by high-content screening microscopy. This method should facilitate drug target validation by small-interfering RNAs.

Image Processing, Computer-Assisted↗