PubMed · 2918747
Single-channel current simulation and recording using a photodiode as current generator.
Abstract
A device which can generate rectangular currents in the picoampere range is described. The current generator is a photodiode connected to the head stage of a single-channel recording amplifier. The photodiode is activated by a light-emitting diode controlled by a computer or any other current source. The device can transmit signals corresponding to simulated single-channel behaviour. Since the kinetic parameters of the simulation are known, the user can test the data acquisition and analysis system under conditions similar to those prevailing during recording from a biological membrane. This current generator can also be used for the tuning of patch-clamp amplifiers; rectangular currents generated by the photodiode allow the frequency response of the amplifier to be properly adjusted.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
D Bertrand, C R Bader, C Distasi, I C Forster. 1989. Single-channel current simulation and recording using a photodiode as current generator.. https://doi.org/10.1016/0165-0270(89)90121-0. https://searchpubmed.org/record/pubmed%3A2918747/single-channel-current-simulation-and-recording-using-a-photodiode-as-current-generator