PubMed · 2715829
Evaluation method for krypton-81m reservoir administration systems.
Abstract
Large variations have been reported in counting rates during lung ventilation studies using different 81mKr administration systems and among different patients. A method was set up to determine the activity utilization efficiency (AUE) using various administration systems. For that purpose a simple lung simulator was developed for combination with reservoir administration systems to be tested. It was found that under normal breathing conditions the AUE is 50% using a reservoir system and only 18% in the absence of a reservoir in the administration system. The measured results were confirmed by a mathematic model. The suggested simulator is suitable for use in hospitals and also enables an indirect check on the 81Rb/81mKr generator performance.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
A G Janssen, F H van Weeren, J J de Goeij, G P Wijnhoven, T J Witsenboer. 1989. Evaluation method for krypton-81m reservoir administration systems.. https://searchpubmed.org/record/pubmed%3A2715829/evaluation-method-for-krypton-81m-reservoir-administration-systems
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search PubMed by Indiciae: BibTeX · RIS.