Abstract

Acute and chronic threshold measurements were compared for three different electrodes: the Medtronic Target Tip model 4011, the Medtronic Ring Tip Model 6971 and the Siemens-Elema Carbon Tip Model 412S. The results of this study show that acute thresholds are lower with the Target Tip lead as compared to the other two. This finding applies to chronic thresholds as well. The results of this study favor programming to lower output energies while maintaining adequate safety margins with this lead system.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

M Djordjevic, P Stojanov, D Velimirovic, D Kocovic. 1986. Target lead--low threshold electrode.. https://doi.org/10.1111/j.1540-8159.1986.tb06695.x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

The development of a multichannel electrode array for retinal prostheses.

The development of a multielectrode array is the key issue for retinal prostheses. We developed a 10 x 10 platinum electrode array that consists of an 8-microm polyimide layer sandwiched between 5-microm polymonochloro-para-xylylene (parylene-C) layers. Each electrode was formed as a 30-microm-high bump by Pt/Au double-layer electroplating. We estimated the charge delivery capability (CDC) of the electrode by measuring the CDCs of two-channel electrode arrays. The dimensions of each electrode of the two-channel array were the same as those of each electrode formed on the 10 x 10 array. The results suggest that for cathodic-first (CF) pulses, 80% of electrodes surpassed our development target of 318 microC/cm2, which corresponds to the charge density of pulses of 500 micros duration and 200 microA amplitude for a 200-microm-diameter planar electrode.

Electrodes, Implanted↗

Improved version of the printed circuit board (PCB) modular multi-channel microdrive for extracellular electrophysiological recordings.

The modular multi-channel PCB microdrive was described some years ago, since then several improvements were introduced while using these drives. Utilizing several years of experience with the original PCB microdrive we redesigned it to improve its stability and usability. The application of the printed circuit board technology and the extensive use of flexible fused silica capillaries for fabrication of the microdrive are described in detail. The improved design led to a low cost and light-weight multi-channel microdrive with outstanding modularity for extracellular field, single unit or multiunit tetrode recording up to 64/128 channels.

Electrodes, Implanted↗