Which balloon embolectomy catheter?

Abstract

This study compared seven makes of balloon embolectomy catheter currently available, testing the most commonly used 4 Fr catheters. The volume of fluid and inflation pressure required to burst balloons of each type were measured. All balloons showed a similar shape of inflation curve, but compliance varied considerably, and the pressure required to burst different balloon types varied from 1060 to 1920 cmH2O. Two balloon types consistently fragmented when they burst in free space. On a specially designed mechanical rig balloon catheters were pulled through a 6 mm tube containing a fixed stenosis, with continuous measurement of traction force and intraballoon pressure. Traction force varied from 34 to 129 g (median 40 g) during passage along the tube, and from 254 to 463 g (median 350 g) during passage through the stenosis. The shear force imparted by the balloon to the tube wall was also calculated. This study provides comparative data which may help to guide surgeons in their choice of balloon embolectomy catheter.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

P G Niblett, J M Fleischl, W B Campbell. 1990. Which balloon embolectomy catheter?. https://doi.org/10.1002/bjs.1800770829

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Immune heparin-induced thrombocytopenia resulting from preceding exposure to heparin catheter flushes.

Immune heparin-induced thrombocytopenia (HIT) is a life-threatening adverse effect of heparin. It can result from any type of heparin exposure and by any route of administration; however only a few cases are reported after exposures to small quantities of heparin from catheter flushes. The major clinical problem associated with HIT is thrombosis. Early detection and institution of alternative, non-heparin anticoagulation are important. We report a patient with HIT associated with use of therapeutic-dose unfractionated heparin in whom immune sensitization to heparin was triggered by two 500-unit exposure to UFH associated with intravascular catheter flushing for antineoplastic chemotherapy in a patient with colon adenocarcinoma.

Catheterization↗