PubMed · 2338894

Modified gradients for motion suppression: variable echo time and variable bandwidth.

Abstract

A linear algebra based deprivation is presented to demonstrate that linearly time scaling an entire gradient waveform by a factor "R" exponentially increases its sensitivity to time derivatives of position by R(i + 1), where i refers to the i-th derivative of position (e.g., i = 1 is velocity). Thus, time scaling will preserve zero valued refocussing moments associated with artifact reduction techniques designed for motion occurring between excitation and detection. Typically, gradient waveforms for artifact reduction techniques are derived for use only at specific echo times. The time scaling described here allows for simple modification of refocussing gradient waveforms for use at variable echo times. Motion sensitivity associated with non-zero moment gradient waveforms can be easily predicted and modified using this technique, with consideration for field of view, resolution, and bandwidth. A clinical example is presented showing the predicted changes in sensitivity to nonrefocussed derivatives of position as the imaging gradients are time scaled. Further, trade-offs and alternatives in sensitivity to motion, slice thickness, image bandwidth, field of view and resolution will be discussed in conjunction with time scaling. This technique will have applicability in many situations involving MRI of moving tissue and a clinical example in cardiac imaging is presented.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J L Duerk, O P Simonetti, G C Hurst. 1990. Modified gradients for motion suppression: variable echo time and variable bandwidth.. https://doi.org/10.1016/0730-725x(90)90247-y

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Blood flow velocity in the fetal abdominal aorta and in the umbilical artery in uncomplicated pregnancies.

The blood flow velocity waveform (BFWV) in fetal vessels depends on the stroke volume and the frequency of the fetal heart, the compliance of the vessel and the peripheral resistance of the fetal vascular bed. The objective of the present study was to establish the change of the BFVW throughout gestation and whether the change of the resistance indices are related to the peak flow velocity and/or the end-diastolic flow velocity. The BFVW of the fetal abdominal aorta and of the umbilical artery at 27th-30th week and at 37th-40th week of gestation were analysed in fifteen patients with uncomplicated pregnancies and subsequent normal outcome. The measurement was performed with a pulsed duplex scanner (Kranzbühler). The Doppler beam had a fixed angle of 50 degrees to the fetal vessel in all cases. The peak flow velocity in the fetal aorta remained constant from the 27th-30th week to the 37th-40th week of gestation: 1418 +/- 248 Hz vs. 1448 +/- 269 Hz. The end-diastolic flow, however, showed a significant increase during the respective periods: from 270 +/- 59 Hz to 325 +/- 69 Hz. The peak flow velocity in the umbilical artery was about 25% below that of the fetal aorta: 1028 +/- 149 Hz (27th-30th week) and 1106 +/- 196 Hz (37th-40th week). The end-diastolic flow, however, increased by about 41%: 292 +/- 86 Hz vs. 412 +/- 83 Hz demonstrating a tremendous change of the compliance and the peripheral resistance in the umbilical vascular system. These alterations are also exhibited by the resistance indices. The resistance indices S/D, PI and RI of both vessels were related to the EDF of the abdominal aorta and the umbilical artery at the respective gestational age. It demonstrated that the EDF is of major influence on the calculated resistance indices. The blood flow in the aorta and the umbilical artery was 135(+/- 18) ml/kg/min and 143 (+/- 21) ml/kg/min, respectively. In conclusion, the BFVW for fetal surveillance should be measured under a constant angle in the fetal abdominal aorta. It gives a good information concerning the peripheral resistance and compliance in the respective vascular areas. The EDFV has its lowest value at 200 Hz in the fetal aorta and in the umbilical artery under physiological conditions.

Aorta, Abdominal