PubMed · 16737217
Solid-support sample loading for DNA sequencing.
Abstract
We present a new method for simplified low-quantity DNA loading onto microelectrophoresis devices. The method is based on combined solid-phase extraction, purification, and transport of DNA reversibly bound on paramagnetic microspheres. DNA is adsorbed onto the microspheres, captured with a magnetized permalloy wire, and then directly injected as a highly focused sample plug into the separation channel. This method circumvents both the minimum volume requirement of pipettors (since only solid beads are transported) and the timing complications of double-T microfluidic injection. Injections from Sanger samples of <100 pg total suspended weight match the signal strength of our previous conventional injections at >10-times the starting DNA sample. Sequencing traces show a resolution that matches or exceeds double-T injections. A kinetic model reproduces the time-dependence of the injection signals and proves that total nonidealities in the method produce injection-broadened plugs of approximately 1-s duration. The method should be broadly extendable to DNA and protein separations in both microdevice and capillary electrophoresis.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Jörn Ueberfeld, Sameh A El-Difrawy, Korisha Ramdhanie, Daniel J Ehrlich. 2006-06-01. Solid-support sample loading for DNA sequencing.. https://doi.org/10.1021/ac052201x
Cite the original work for its findings. Save a collection to share your selection of sources.