PubMed · 16636413
A high-precision voltage source for EIT.
Abstract
Electrical impedance tomography (EIT) utilizes electrodes placed on the surface of a body to determine the complex conductivity distribution within the body. EIT can be performed by applying currents through the electrodes and measuring the electrode voltages or by applying electrode voltages and measuring the currents. Techniques have also been developed for applying the desired currents using voltage sources. This paper describes a voltage source for use in applied-voltage EIT that includes the capability of measuring both the applied voltage and applied current. A calibration circuit and calibration algorithm are described which enables all voltage sources in an EIT system to be calibrated to a common standard. The calibration minimizes the impact of stray shunt impedance, passive component variability and active component non-ideality. Simulation data obtained using PSpice are used to demonstrate the effectiveness of the circuits and calibration algorithm.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Gary J Saulnier, Alexander S Ross, Ning Liu. 2006-04-24. A high-precision voltage source for EIT.. https://doi.org/10.1088/0967-3334%2F27%2F5%2Fs19
Cite the original work for its findings. Save a collection to share your selection of sources.