Microarray scanner calibration curves: characteristics and implications.

Abstract

BACKGROUND: Microarray-based measurement of mRNA abundance assumes a linear relationship between the fluorescence intensity and the dye concentration. In reality, however, the calibration curve can be nonlinear. RESULTS: By scanning a microarray scanner calibration slide containing known concentrations of fluorescent dyes under 18 PMT gains, we were able to evaluate the differences in calibration characteristics of Cy5 and Cy3. First, the calibration curve for the same dye under the same PMT gain is nonlinear at both the high and low intensity ends. Second, the degree of nonlinearity of the calibration curve depends on the PMT gain. Third, the two PMTs (for Cy5 and Cy3) behave differently even under the same gain. Fourth, the background intensity for the Cy3 channel is higher than that for the Cy5 channel. The impact of such characteristics on the accuracy and reproducibility of measured mRNA abundance and the calculated ratios was demonstrated. Combined with simulation results, we provided explanations to the existence of ratio underestimation, intensity-dependence of ratio bias, and anti-correlation of ratios in dye-swap replicates. We further demonstrated that although Lowess normalization effectively eliminates the intensity-dependence of ratio bias, the systematic deviation from true ratios largely remained. A method of calculating ratios based on concentrations estimated from the calibration curves was proposed for correcting ratio bias. CONCLUSION: It is preferable to scan microarray slides at fixed, optimal gain settings under which the linearity between concentration and intensity is maximized. Although normalization methods improve reproducibility of microarray measurements, they appear less effective in improving accuracy.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Leming Shi, Weida Tong, Zhenqiang Su, Tao Han, Jing Han, Raj K Puri, Hong Fang, Felix W Frueh, Federico M Goodsaid, Lei Guo, William S Branham, James J Chen, Z Alex Xu, Stephen C Harris, Huixiao Hong, Qian Xie, Roger G Perkins, James C Fuscoe. 2005-07-15. Microarray scanner calibration curves: characteristics and implications.. https://doi.org/10.1186/1471-2105-6-s2-s11

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Chemical images: technical approaches and issues.

Chemical mapping techniques using Raman microscopy are introduced, and using an example of a pharmaceutical tablet, the practical aspects of data collection and processing to produce a chemical image of the sample are detailed. Issues related to data processing, instrument standards, chemical image reportable errors, and the interpretation of chemical images are presented to encourage debate, develop solutions, and promote use in other challenging scientific applications. applications.

Calibration↗

Nondestructive determination of lignans and lignan glycosides in sesame seeds by near infrared reflectance spectroscopy.

Sesame (Sesamum indicum L.) contains abundant lignans including lipid-soluble lignans (sesamin and sesamolin) and water-soluble lignan glycosides (sesaminol triglucoside and sesaminol diglucoside) related to antioxidative activity. In this study, near infrared reflectance spectroscopy (NIRS) was used to develop a rapid and nondestructive method for the determination of lignan contents on intact sesame seeds. Ninety-three intact seeds were scanned in the reflectance mode of a scanning monochromator. This scanning procedure did not require the pulverization of samples, allowing each analysis to be completed within minutes. Reference values for lignan contents were obtained by high-performance liquid chromatography analysis. Calibration equations for lignans (sesamin and sesamolin) and lignan glycosides (sesaminol triglucoside and sesaminol diglucoside) contents were developed using modified partial least squares regression with internal cross-validation (n = 63). The equations obtained had low standard errors of cross-validation and moderate R2 (coefficient of determination in calibration). The prediction of an external validation set (n = 30) showed significant correlation between reference values and NIRS predicted values based on the SEP (standard error of prediction), bias, and r2 (coefficient of determination in prediction). The models developed in this study had relatively higher values (more than 2.0) of SD/SEP(C) for all lignans and lignan glycosides except for sesaminol diglucoside, which had a minor amount, indicating good correlation between the reference and the NIRS estimate. The results showed that NIRS, a nondestructive screening method, could be used to rapidly determine lignan and lignan glycoside contents in the breeding programs for high quality sesame.

Calibration↗

Comparative uptake, metabolism, and utilization of menaquinone-4 and phylloquinone in human cultured cell lines.

It is generally accepted that the availability of vitamin K in vivo depends on its homologues, the biological activities of which would differ among organs. To test this hypothesis, we examined the uptake, metabolism, and utilization of menaquinone-4 (MK-4) and phylloquinone (PK) using 18O-labeled compounds in two cultured human cell lines (HepG2 and MG-63). Lipid extracts were prepared from the cells and media after 1, 3, and 6h of incubation. The detection of the vitamin K analogues (18O-, 16O-quinone, and epoxide forms) was carried out with LC-APCI-MS/MS as previously reported. The 18O of vitamin K was replaced with atmospheric 16O2 during the formation of vitamin K epoxide with a carboxylative catalytic reaction. As a result, a significant difference was observed between MK-4 and PK in the amounts taken up into the cells. The 18O-labeled MK-4 was rapidly and remarkably well absorbed into the cells and metabolized to the epoxide form via a hydroquinone form as compared to the 18O-labeled PK. The difference in uptake of MK-4 and PK was not affected by treatment with warfarin although the metabolism of both compounds was markedly inhibited. This methodology should be utilized to clarify some of the actions of vitamin K in target cells and facilitate the development of new vitamin K drugs.

Calibration↗