PubMed · 15940207

Virtual micro MLC commissioning.

Abstract

The resolution of multileaf collimators (MLCs) is limited by their finite leaf width. A commercial package (HD-270) uses 3D couch translation and leaf adjustments to emulate smaller leaf widths. In this paper, we report on the commissioning of this feature using software testing, dial gauge measurements, and film dosimetry. We also identify a variety of limitations: software bugs and truncation artifacts, MLC leaf positioning uncertainties (random variations, systematic gantry dependence and backlash), and uncertainties in couch positioning. These reduce the capabilities of this implementation below that achievable theoretically.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Laura Drever, Peter Dickof. 2005-05-19. Virtual micro MLC commissioning.. https://doi.org/10.1120/jacmp.v6i2.2032

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

History of tomotherapy.

Tomotherapy is the delivery of intensity modulated radiation therapy using rotational delivery of a fan beam in the manner of a CT scanner. In helical tomotherapy the couch and gantry are in continuous motion akin to a helical CT scanner. Helical tomotherapy is inherently capable of acquiring CT images of the patient in treatment position and using this information for image guidance. This review documents technological advancements of the field concentrating on the conceptual beginnings through to its first clinical implementation. The history of helical tomotherapy is also a story of technology migration from academic research to a university-industrial partnership, and finally to commercialization and widespread clinical use.

Equipment Design↗

Modification of Gore suture passer instrument.

BACKGROUND: The Gore suture passer instrument is an ideal reusable device for various laparoscopic procedures albeit with a few drawbacks. The needle is not self-retracting inside the sheath and this leads to incomplete (manual) retraction of the needle, causing bending and breakage of the needle tip. The intracorporeal orientation of the suture grasper is difficult to maintain because the needle and suture grasper is fully rotational inside the handle. METHODS: We rectified these drawbacks by inserting a spring below the connecting nut in the handle component to make the needle self-retracting. The needle and suture grasper was stabilized by a screw in the main body of the handle to fit in a longitudinal slot cut on the brass connecting nut. RESULTS: Our modification made the needle and suture grasper self-retracting inside the sheath and stabilize the needle during the insertion. We have used this modified suture passer on 80 patients without breakage and bending of the needle. CONCLUSION: This modification has made this device user friendly, steady, and has reduced the operation time.

Equipment Design↗

Evaluation of individual monitoring in mixed neutron/photon fields: mid-term results from the EVIDOS project.

EVIDOS is an EC sponsored project that aims at an evaluation and improvement of radiation protection dosimetry in mixed neutron/photon fields. This is performed through spectrometric and dosimetric investigations during different measurement campaigns in representative workplaces of the nuclear industry. The performance of routine and, in particular, novel personal dosemeters and survey instruments is tested in selected workplace fields. Reference values for the dose equivalent quantities, H(*)(10) and H(p)(10) and the effective dose E, are determined using different spectrometers that provide the energy distribution of the neutron fluence and using newly developed devices that determine the energy and directional distribution of the neutron fluence. The EVIDOS project has passed the mid-term, and three measurement campaigns have been performed. This paper will give an overview and some new results from the third campaign that was held in Mol (Belgium), around the research reactor VENUS and in the MOX producing plant of Belgonucléaire.

Equipment Design↗